Incomplete Screw Thread Engagement of Proximal Fragment: A Possible Failure Risk After Internal Fixation for Femoral Neck Fractures

被引:0
|
作者
Tanaka, Atsuki [1 ,2 ]
Hiranaka, Takafumi [1 ,2 ]
Fujishiro, Takaaki [1 ,2 ]
Koide, Motoki [1 ,2 ]
Suda, Yoshihito [1 ,2 ]
Saito, Akira [1 ,2 ]
Arimoto, Akihiko [1 ,2 ]
Okamoto, Koji [1 ,2 ]
机构
[1] Takatsuki Gen Hosp, Dept Orthopaed Surg, Takatsuki, Japan
[2] Takatsuki Gen Hosp, Joint Surg Ctr, Takatsuki, Japan
关键词
incomplete thread engagement; fixation failure; sliding mechanism; thread length; internal fixation; femoral neck fracture; OUTCOMES;
D O I
10.7759/cureus.41349
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background For successful internal fixation for femoral neck fracture, the sliding mechanism of the screw is important because it can induce inter-fragmental compression. The thread should penetrate the fracture line and be located within the proximal fragment. If screw thread engagement is incomplete and a part of the thread remains within the distal fragment, the screw sliding can be disturbed, potentially leading to fixation failure. We hypothesized that screw thread in the fracture is a risk of fixation failure. Methods We studied 133 hips that underwent internal fixation for femoral neck fracture using dual sliding and compression screws (DSCS) with 20 mm threads. The existence of incomplete thread engagement and fixation failure (cut out, perforation, pseudoarthrosis, or femoral neck shortening) were evaluated on anteroposterior hip radiography postoperatively. The distances from the thread end to the fracture line, screw head to the femoral head cortex, and femoral head diameter were measured to analyze their relationships with any incomplete thread engagement and fixation failure. Differences in evaluation data were assessed using Fisher's exact test, Student's t -test, and receiver operating characteristic (ROC) analysis. Results Forty-six cases had at least one screw with incomplete thread engagement, and the other 87 hips had a complete engagement. The failure rate in the group of hips with incomplete thread engagement was significantly higher (7/46, 15.2%) than that in the group of hips with complete thread engagement (3/87, 3.4%) (P = 0.032). Incomplete thread engagement was found in 59 out of 266 screws (22.2%), and a femoral head <= 43.9 mm in diameter was associated with an increased risk of incomplete thread engagement. Most incomplete thread engagement screws (81.4%) had < 5 mm thread length within the distal fragment. Conclusion A partially threaded screw is a significant risk of fixation failure after internal fixation for a femoral neck fracture. The smaller femoral head diameter increases the possibility of incomplete thread engagement. Shortening the thread length by 5 mm may help to avoid incomplete thread engagement.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] Radiographic fracture features predicting failure of internal fixation of displaced femoral neck fractures
    N. L. Weil
    D. van Embden
    J. M. Hoogendoorn
    European Journal of Trauma and Emergency Surgery, 2015, 41 : 501 - 507
  • [22] Catastrophic Failure After Open Reduction Internal Fixation of Femoral Neck Fractures With a Novel Locking Plate Implant
    Berkes, Marschall B.
    Little, Milton T. M.
    Lazaro, Lionel E.
    Cymerman, Rachel M.
    Helfet, David L.
    Lorich, Dean G.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2012, 26 (10) : E170 - E176
  • [23] Risk factors for cut-out of double lag screw fixation in proximal femoral fractures
    Buyukdogan, Kadir
    Caglar, Omur
    Isik, Samet
    Tokgozoglu, Mazhar
    Atilla, Bulent
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2017, 48 (02): : 414 - 418
  • [24] Radiographic fracture features predicting failure of internal fixation of displaced femoral neck fractures
    Weil, N. L.
    van Embden, D.
    Hoogendoorn, J. M.
    EUROPEAN JOURNAL OF TRAUMA AND EMERGENCY SURGERY, 2015, 41 (05) : 501 - 507
  • [25] FAILURE OF INTERNAL-FIXATION OF DISPLACED FEMORAL-NECK FRACTURES IN RHEUMATOID PATIENTS
    BOGOCH, E
    OUELLETTE, G
    HASTINGS, D
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1991, 73 (01): : 7 - 10
  • [26] Fixation of intracapsular fractures of the femoral neck in young patients RISK FACTORS FOR FAILURE
    Duckworth, A. D.
    Bennet, S. J.
    Aderinto, J.
    Keating, J. F.
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (06): : 811 - 816
  • [27] Inadequate proximal screw fixation increases risk of failure following plate fixation of diaphyseal humerus fractures
    Manisha R. Mistry
    Jimmy Tat
    Rafi Husain
    Ujash Sheth
    Robin R. Richards
    Diane Nam
    Journal of Orthopaedic Surgery and Research, 18
  • [28] Inadequate proximal screw fixation increases risk of failure following plate fixation of diaphyseal humerus fractures
    Mistry, Manisha R. R.
    Tat, Jimmy
    Husain, Rafi
    Sheth, Ujash
    Richards, Robin R. R.
    Nam, Diane
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2023, 18 (01)
  • [29] The Targon Femoral Neck hip screw versus cannulated screws for internal fixation of intracapsular fractures of the hip
    Griffin, X. L.
    Parsons, N.
    Achten, J.
    Costa, M. L.
    BONE & JOINT JOURNAL, 2014, 96B (05): : 652 - 657
  • [30] SEASONAL PERIODICITY OF SECONDARY HIP REPLACEMENT AFTER FEMORAL NECK FRACTURES WITH REDUCTION INTERNAL SCREW FIXATION AGED OVER 60
    Sebestyen, A.
    Gajdacsi, J.
    Patzai, B.
    Molics, B.
    Varga, S.
    Sandor, J.
    Boncz, I
    VALUE IN HEALTH, 2012, 15 (07) : A403 - A403