Biomechanical evaluation of a novel anterior transpedicular screw-plate system for anterior cervical corpectomy and fusion (ACCF): a finite element analysis

被引:1
|
作者
Huang, Shengbin [1 ,2 ]
Ling, Qinjie [3 ]
Lin, Xinxin [2 ]
Qin, Hao [2 ]
Luo, Xiang [2 ]
Huang, Wenhua [1 ,4 ]
机构
[1] Guangxi Med Univ, Sch Basic Med Sci, Dept Human Anat, Nanning, Guangxi, Peoples R China
[2] Guangxi Med Univ, Dept Cardiol, Affiliated Hosp 8, Guigang, Peoples R China
[3] Guangzhou Med Univ, Dept Spinal Surg, Affiliated Hosp 1, Guangzhou, Guangdong, Peoples R China
[4] Southern Med Univ, Guangdong Engn Res Ctr Translat Med 3D Printing Ap, Guangdong Prov Key Lab Med Biomech, Sch Basic Med Sci,Natl Key Discipline Human Anat, Guangzhou, Guangdong, Peoples R China
基金
中国国家自然科学基金;
关键词
cervical spine; anterior pedicle screw; anterior surgery; corpectomy; reconstruction; biomechanics; finite element analysis; PULL-OUT STRENGTH; PEDICLE-SCREW; FIXATION; SPINE; RECONSTRUCTION; STABILITY; 4-LEVEL; INJURY; LOAD;
D O I
10.3389/fbioe.2023.1260204
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Background and objective: Cervical fusion with vertebral body screw (VBS)-plate systems frequently results in limited biomechanical stability. To address this issue, anterior transpedicular screw (ATPS) fixation has been developed and applied preliminarily to multilevel spinal fusion, osteoporosis, and three-column injury of the cervical spine. This study aimed to compare the biomechanical differences between unilateral ATPS (UATPS), bilateral ATPS (BATPS), and VBS fixation using finite element analysis.Materials and methods: A C6 corpectomy model was performed and a titanium mesh cage (TMC) and bone were implanted, followed by implantation of a novel ATPS-plate system into C5 and C7 to simulate internal fixation with UATPS, BATPS, and VBS. Internal fixation with UATPS comprises ipsilateral transpedicular screw-contralateral vertebral body screw (ITPS-CVBS) and cross transpedicular screw-vertebral body screw (CTPS-VBS) fixations. Mobility, the maximal von Mises stress on TMC, the stress distribution and maximal von Mises stress on the screws, and the maximum displacement of the screw were compared between the four groups.Results: Compared with the original model, each group had a reduced range of motion (ROM) under six loads. After ACCF, the stress was predominantly concentrated at two-thirds of the length from the tail of the screw, and it was higher on ATPS than on VBS. The stress of the ATPS from the cranial part was higher than that of the caudal part. The similar effect happened on VBS. The screw stress cloud maps did not show any red areas reflective of a concentration of the stress on VBS. Compared with VBS, ATPS can bear a greater stress from cervical spine movements, thus reducing the stress on TMC. The maximal von Mises stress was the lowest with bilateral transpedicular TMC and increased with cross ATPS and with ipsilateral ATPS. ITPS-CVBS, CTPS-VBS, and BATPS exhibited a reduction of 2.3%-22.1%, 11.9%-2.7%, and 37.9%-64.1% in the maximum displacement of screws, respectively, compared with that of VBS.Conclusion: In FEA, the comprehensive stability ranked highest for BATPS, followed by CTPS-VBS and ITPS-CVBS, with VBS demonstrating the lowest stability. Notably, utilizing ATPS for fixation has the potential to reduce the occurrence of internal fixation device loosening after ACCF when compared to VBS.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Biomechanical Comparison of Integrated Fixation Cage Versus Anterior Cervical Plate and Cage in Anterior Cervical Corpectomy and Fusion (ACCF): A Finite Element Analysis
    Ouyang, Pengrong
    Lu, Teng
    He, Xijing
    Gao, Zhongyang
    Cai, Xuan
    Jin, Zhongmin
    MEDICAL SCIENCE MONITOR, 2019, 25 : 1489 - 1498
  • [2] Biomechanical study of anterior transpedicular root screw intervertebral fusion system of lower cervical spine: a finite element analysis
    Ye, Senqi
    Ye, Jiachun
    Hou, Zhipeng
    You, Xinmao
    Shen, Shufeng
    Zhang, Jihui
    Yu, Liang
    Gu, Yongjie
    Wang, Wei
    Zhao, Liujun
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2024, 12
  • [3] Biomechanical analysis on of anterior transpedicular screw-fixation after two-level cervical corpectomy using finite element method
    Zhao, Liujun
    Chen, Jianqing
    Liu, Jiayong
    Elsamaloty, Lina
    Liu, Xiaochen
    Li, Jie
    Elgafy, Hossein
    Zhang, Jihui
    Wang, Leining
    CLINICAL BIOMECHANICS, 2018, 60 : 76 - 82
  • [4] Screw angulation affects bone-screw stresses and bone graft load sharing in anterior cervical corpectomy fusion with a rigid screw-plate construct: a finite element model study
    Hussain, Mozammil
    Natarajan, Raghu N.
    Fayyazi, Amir H.
    Braaksma, Brian R.
    Andersson, Gunnar B. J.
    An, Howard S.
    SPINE JOURNAL, 2009, 9 (12): : 1016 - 1023
  • [5] Effect of corpectomy and discectomy fusion procedures on the stability of multi-level cervical construct with anterior rigid screw-plate fixation - A finite element model study
    Hussain, Mozammil
    Nassr, Ahmad N.
    Natarajan, Raghu N.
    Andersson, Gunnar B. J.
    An, Howard S.
    PROCEEDING OF THE ASME SUMMER BIOENGINEERING CONFERENCE - 2007, 2007, : 435 - 436
  • [6] Biomechanical Comparison of Anterior Cervical Corpectomy Decompression and Fusion, Anterior Cervical Discectomy and Fusion, and Anterior Controllable Antedisplacement and Fusion in the Surgical Treatment of Multilevel Cervical Spondylotic Myelopathy: A Finite Element Analysis
    Kong, Qingjie
    Li, Fudong
    Yan, Chen
    Sun, Jingchuan
    Sun, Peidong
    Ou-Yang, Jun
    Zhong, Shizhen
    Wang, Yuan
    Shi, Jiangang
    ORTHOPAEDIC SURGERY, 2024, 16 (03) : 687 - 699
  • [7] Anterior cervical transpedicular screw fixation system in subaxial cervical spine: A finite element comparative study
    Li, Jie
    Gan, Kaifeng
    Chen, Binhui
    Chen, Yilei
    Hong, Jinjiong
    Bei, Dikai
    Fan, Tengdi
    Zheng, Minzhe
    Zhao, Liujun
    Zhao, Fengdong
    MEDICINE, 2022, 101 (29) : E29316
  • [8] Biomechanical Effects of a Novel Anatomic Titanium Mesh Cage for Single-Level Anterior Cervical Corpectomy and Fusion: A Finite Element Analysis
    Zhang, Ke-rui
    Yang, Yi
    Ma, Li-tai
    Qiu, Yue
    Wang, Bei-yu
    Ding, Chen
    Meng, Yang
    Rong, Xin
    Hong, Ying
    Liu, Hao
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2022, 10
  • [9] Effects of New Cage Profiles on the Improvement in Biomechanical Performance of Multilevel Anterior Cervical Corpectomy and Fusion: A Finite Element Analysis
    Liu, Ning
    Lu, Teng
    Wang, Yibin
    Sun, Zhongwei
    Li, Jialiang
    He, Xijing
    WORLD NEUROSURGERY, 2019, 129 : E87 - E96
  • [10] A Novel Anterior Transpedicular Screw Artificial Vertebral Body System for Lower Cervical Spine Fixation: A Finite Element Study
    Wu, Weidong
    Chen, Chun
    Ning, Jinpei
    Sun, Peidong
    Zhang, Jinyuan
    Wu, Changfu
    Bi, Zhenyu
    Fan, Jihong
    Lai, Xianliang
    Ouyang, Jun
    JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2017, 139 (06):