Motion Detectors as Additional Monitoring Devices in the Intensive Care Unit-A Proof-of-Concept Study

被引:0
|
作者
Gueder, Guelmisal [1 ,2 ]
von Rein, Eva [1 ]
Flohr, Thomas [1 ]
Weismann, Dirk [1 ]
Schmitt, Dominik [1 ]
Stoerk, Stefan [1 ,2 ]
Frantz, Stefan [1 ,2 ]
Kratzer, Vincent [3 ]
Kendi, Christian [3 ]
机构
[1] Univ Hosp Wurzburg, Dept Internal Med 1, Cardiol Div, D-97080 Wurzburg, Germany
[2] Univ Hosp Wurzburg, Comprehens Heart Failure Ctr, Dept Clin Res & Epidemiol, D-97078 Wurzburg, Germany
[3] IRON Software GmbH, D-82031 Grunwald, Germany
来源
APPLIED SCIENCES-BASEL | 2023年 / 13卷 / 16期
关键词
motion detector; noncontact monitoring; Internet of Things devices; ALARM FATIGUE; DELIRIUM;
D O I
10.3390/app13169319
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Background: Monitoring the vital signs of delirious patients in an intensive care unit (ICU) is challenging, as they might (un-)intentionally remove devices attached to their bodies. In mock-up scenarios, we systematically assessed whether a motion detector (MD) attached to the bed may help in identifying emergencies. Methods: We recruited 15 employees of the ICU and equipped an ICU bed with an MD (IRON Software GmbH, Grunwald, Germany). Participants were asked to replay 22 mock-up scenes of one-minute duration each: 12 scenes with movements and 10 without movements, of which 5 were emergency scenes ("lying dead-still, with no or very shallow breathing"). Blinded recordings were presented to an evaluation panel consisting of an experienced ICU nurse and a physician, who was asked to assess and rate the presence of motions. Results: Fifteen participants (nine women; 173 +/- 7.0 cm; 78 +/- 19 kg) joined the study. In total, 286 out of 330 scenes (86.7%) were rated correctly. Ratings were false negative (FN: "no movements detected, but recorded") in 7 out of 180 motion scenes (3.9%). Ratings were false positive (FP: "movements detected, but not recorded") in 37 out of 150 scenes (24.7%), more often in men than women (26 out of 60 vs. 11 out of 90, respectively; p < 0.001). Of note, in 16 of these 37 FP-rated scenes, a vibrating mobile phone was identified as a potential confounder. The emergency scenes were correctly rated in 64 of the 75 runs (85.3%); 10 of the 11 FP-rated scenes occurred in male subjects. Conclusions: The MD allowed for identifying motions of test subjects with high sensitivity (96%) and acceptable specificity (75%). Accuracy might increase further if activities are recorded continuously under real-world conditions.
引用
收藏
页数:15
相关论文
共 50 条
  • [21] Parents' Views of Family-Centered Care at a Pediatric Intensive Care Unit-A Qualitative Study
    Terp, Karina
    Weis, Janne
    Lundqvist, Pia
    FRONTIERS IN PEDIATRICS, 2021, 9
  • [22] Bioimpedance method for monitoring venous ulcers: Clinical proof-of-concept study
    Kekonen, Atte
    Bergelin, Mikael
    Eriksson, Jan-Erik
    Vaalasti, Annikki
    Ylanen, Heimo
    Kielosto, Sami
    Viik, Jari
    BIOSENSORS & BIOELECTRONICS, 2021, 178
  • [23] Use of Instagram filter in monitoring essential tremor: A proof-of-concept study
    Ismail, Ismail
    Kamel, Walaa
    Al-Hashel, Jasem
    JOURNAL OF THE NEUROLOGICAL SCIENCES, 2023, 455
  • [24] Tattoo-Based Noninvasive Glucose Monitoring: A Proof-of-Concept Study
    Bandodkar, Amay J.
    Jia, Wenzhao
    Yardimci, Ceren
    Wang, Xuan
    Ramirez, Julian
    Wang, Joseph
    ANALYTICAL CHEMISTRY, 2015, 87 (01) : 394 - 398
  • [25] Cough as Noninvasive Biomarker for Monitoring Tuberculosis Treatment: A Proof-of-Concept Study
    Janssen, Saskia
    Upton, Caryn M.
    De Jager, Veronique R.
    Faraj, Alan
    Pahar, Madhurananda
    Miranda, Igor D. S.
    Diacon, Andreas H.
    Simonsson, Ulrika S. H.
    Niesler, Thomas R.
    ANNALS OF THE AMERICAN THORACIC SOCIETY, 2023, 20 (12) : 1822 - 1825
  • [26] Sublingual microcirculation does not reflect red blood cell transfusion thresholds in the intensive care unit-a prospective observational study in the intensive care unit
    Scheuzger, Jonas
    Zehnder, Anna
    Meier, Vera
    Yeginsoy, Desiree
    Flukiger, Julian
    Siegemund, Martin
    CRITICAL CARE, 2020, 24 (01):
  • [27] Using Passive BCI for Personalization of Assistive Wearable Devices: A Proof-of-Concept Study
    Mahmoudi, Asghar
    Khosrotabar, Morteza
    Gramann, Klaus
    Rinderknecht, Stephan
    Sharbafi, Maziar A.
    IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2025, 33 : 476 - 487
  • [28] Blockchain for Increased Trust in Virtual Health Care: Proof-of-Concept Study
    Hasselgren, Anton
    Rensaa, Jens-Andreas Hanssen
    Kralevska, Katina
    Gligoroski, Danilo
    Faxvaag, Arild
    JOURNAL OF MEDICAL INTERNET RESEARCH, 2021, 23 (07)
  • [29] Sublingual microcirculation in prehospital critical care medicine: A proof-of-concept study
    Bruno, Raphael Romano
    Reed, Markus
    Bimpong-Buta, Nana-Yaw
    Muessig, Johanna M.
    Masyuk, Maryna
    Binneboessel, Stephan
    Franz, Marcus
    Kelm, Malte
    Jung, Christian
    MICROCIRCULATION, 2020, 27 (05)
  • [30] A tumor motion-compensating system with tracking and prediction - a proof-of-concept study
    Guo, Kaiming
    Teo, Peng
    Kawalec, Philip
    Pistorius, Stephen
    MEDICAL PHYSICS, 2016, 43 (08) : 4948 - 4949