A calculable pistonphone for the absolute calibration of hydrophones in the frequency range from 0.5 Hz to 250 Hz

被引:1
|
作者
Malcher, Freya [1 ]
Ford, Ben [1 ]
Barham, Richard [2 ]
Robinson, Stephen [1 ]
Ward, Jake [1 ]
Wang, Lian [1 ]
Bridges, Angus [1 ]
Yacoot, Andrew [1 ]
Cheong, Sei-Him [1 ]
Rodrigues, Dominique [3 ]
Barrera-Figueroa, Salvador [4 ]
机构
[1] Natl Phys Lab, Teddington TW11 0LW, England
[2] St Andrews Church, Acoust Sensor Networks ASN, Cuckfield Lane, Warninglid RH17 5UB, England
[3] Lab Commun Metrol LNE CNAM LCM, 29 Ave Roger Hennequin, F-78197 Trappes, France
[4] Danish Fundamental Metrol DFM, Kogle Alle 5, DK-2970 Horsholm, Denmark
基金
欧盟地平线“2020”;
关键词
infrasound; calibration; calculable pistonphone; hydrophone; underwater acoustics; OCEAN; SOUND; PRESSURE;
D O I
10.1088/1681-7575/ad2d5a
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The demand for traceable hydrophone calibrations at low frequencies in support of ocean monitoring applications requires primary standard methods that are able to realise the acoustic pascal. In this paper, a new method for primary calibration of hydrophones is described based on the use of a calculable pistonphone to cover frequencies from 0.5 Hz to 250 Hz. The design consists of a pre-stressed piezoelectric stack driving a piston to create a varying pressure in an air-filled enclosed cavity, the displacement (and so the volume velocity) of the piston being measured by a laser interferometer. The dimensions of the front cavity were designed to allow the calibration of reference hydrophones, but it may also be used to calibrate microphones. Examples of calibration results for several sensors are presented alongside an uncertainty budget for hydrophone calibration with expanded uncertainties ranging from 0.45 dB at 0.5 Hz to 0.30 dB at 20 Hz, and to 0.35 at 250 Hz (expressed for a coverage factor of k = 2). The metrological performance is demonstrated by comparisons with results for other calibration methods and an independent implementation of primary calibration methods at other institutes.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] A laser pistonphone designed for absolute calibration of infrasound sensors from 10 mHz up to 20 Hz
    Rodrigues, D.
    Vincent, P.
    Barham, R.
    Larsonnier, F.
    Durand, S.
    METROLOGIA, 2023, 60 (01)
  • [2] ABSOLUTE CALIBRATION OF MEASURING HYDROPHONES IN AUDIO-FREQUENCY RANGE
    GOLENKOV, AN
    PAVLOV, LE
    MEASUREMENT TECHNIQUES-USSR, 1967, (05): : 579 - &
  • [3] Development and evaluation of primary calibration methods in underwater acoustics in a frequency range from 1 Hz to 100 Hz
    Biber, Alper
    Corakci, Ata C.
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2023, 153 (03):
  • [4] CCAUV.W-K2 final report - key comparison CCAUV.W-K2: calibration of hydrophones in the frequency range from 250 Hz to 500 kHz
    Robinson S.P.
    Harris P.M.
    Ford B.
    Ablitt J.
    Yi C.
    Biber A.
    Corakci A.C.
    Costa-Felix R.
    Crocker S.
    Isaev A.
    Tyalimpi V.
    Latha G.
    Metrologia, 2023, 59 (1A)
  • [5] Inductance calibration in the frequency range from 50 Hz to 1 MHz at PTB
    Hanke, R
    Kölling, A
    Melcher, J
    2002 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS, CONFERENCE DIGEST, 2002, : 186 - 187
  • [6] Neonatal frequency discrimination in 250-4000-Hz range:: Electrophysiological evidence
    Novitski, Nikolai
    Huotilainen, Minna
    Tervaniemi, Mari
    Naatanen, Risto
    Fellman, Vineta
    CLINICAL NEUROPHYSIOLOGY, 2007, 118 (02) : 412 - 419
  • [7] Primary calibration of membrane hydrophones in the frequency range 0.5 MHz to 60 MHz
    Preston, RC
    Robinson, SP
    Zeqiri, B
    Esward, TJ
    Gélat, PN
    Lee, ND
    METROLOGIA, 1999, 36 (04) : 331 - 343
  • [8] Primary calibration of membrane hydrophones in the frequency range 0.5 MHz to 60 MHz
    National Physical Laboratory, Teddington, Middlesex, TW11 0LW, United Kingdom
    Metrologia, 4 (331-343):
  • [9] Study on calibration of hydrophones from 25 Hz to 2 kHz based on coupler reciprocity method
    Yuan, Pengcheng
    Wang, Min
    Zhu, Haijiang
    PROCEEDINGS OF THE 2019 31ST CHINESE CONTROL AND DECISION CONFERENCE (CCDC 2019), 2019, : 3227 - 3232
  • [10] Measurement Uncertainty in Vibration Calibration in Frequency Range of 5 Hz to 10 kHz
    Garg, Naveen
    Chauhan, B. S.
    MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA, 2020, 35 (03): : 397 - 405