The EURADOS CR-39 Quality task for the optimisation and harmonisation of personal neutron dosimetry with CR-39 (PADC)

被引:2
|
作者
Bolzonella, Matteo [1 ]
Caresana, Marco [1 ]
Gilvin, Philip J. [2 ]
Yukihara, Eduardo G. [3 ]
Chevallier, Marie-Anne [4 ]
机构
[1] Politecn Milan, DOE, Via Lambruschini 4, I-20156 Milan, Italy
[2] UK Hlth Secur Agcy, Personal Dosimetry Serv, Didcot OX11 0RQ, England
[3] Paul Scherrer Inst, Dept Radiat Safety & Secur, Forsch str 111, CH-5232 Villigen, Switzerland
[4] Inst Radioprotect & Surete Nucleaire IRSN, Dosimetry Res Dept, BP 17,31 Ave Div Leclerc, F-92260 Fontenay Aux Roses, France
关键词
D O I
10.1093/rpd/ncad068
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
CR-39 (PADC) nuclear track detectors are among the most widespread devices used for personal neutron dosimetry; however, some issues related to the variable material quality of the CR-39 polymer hinder the performance of CR-39-based dosemeters. For this reason, the Working Group 2 (WG2) of the European Radiation Dosimetry Group (EURADOS) has recently launched the CR-39 Quality task, a project aimed at improving and harmonising personal neutron dosimetry with CR-39 in Europe. Whitin this task, a close collaboration among researchers, individual monitoring services and dosemeter grade CR-39 manufacturers is achieved, thus facilitating the direct dialog between producer and consumer to reach an optimised material for personal neutron dosimetry applications.
引用
收藏
页码:1867 / 1871
页数:5
相关论文
共 50 条
  • [31] DOSIMETRY WITH CR-39 IN A PHANTOM IRRADIATED BY A P(43)-BE NEUTRON BEAM
    BLUE, TE
    REYES, JJ
    WEHRING, BW
    BLUE, JW
    ROBERTS, WK
    MEDICAL PHYSICS, 1983, 10 (01) : 132 - 132
  • [33] Neutron field mapping and dosimetry by CR-39 for radiography and other applications
    Pálfalvi, JK
    Sajó-Bohus, L
    Balaskó, M
    Balásházy, I
    RADIATION MEASUREMENTS, 2001, 34 (1-6) : 471 - 475
  • [34] NEUTRON DOSIMETRY WITH BORON-DOPED CR-39 PLASTIC PLATE
    TSURUTA, T
    KATAOKA, S
    MORISHIMA, H
    KAWAI, H
    KATSURAYAMA, K
    MATSUMOTO, S
    HEALTH PHYSICS, 1984, 46 (06): : 1324 - 1324
  • [35] A LONG-TERM STUDY OF CR-39 DETECTORS FOR NEUTRON DOSIMETRY
    TOMMASINO, L
    TORRI, G
    CAVAIOLI, M
    RICCARDI, M
    VANGIAP, T
    NUCLEAR TRACKS AND RADIATION MEASUREMENTS, 1993, 22 (1-4): : 719 - 722
  • [36] Performance of CR-39® with addition of DOP (dioctylphthalate) for fast neutron dosimetry
    Fantuzzi, E
    Marlotti, F
    Falangi, G
    RADIATION MEASUREMENTS, 2003, 36 (1-6) : 475 - 478
  • [37] DOSIMETRY IN FAST-NEUTRON THERAPY BY AUTOMATIC READOUT OF CR-39
    BLUE, TE
    WEHRING, BW
    BRIGGS, SJ
    WEPPRECHT, CS
    DURHAM, JS
    BLUE, JW
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1986, 14 (4-6): : 589 - 595
  • [38] NEUTRON PERSONNEL DOSIMETRY USING A CR-39 CARBONATE PLASTIC DETECTOR
    GOMAA, MA
    KASIM, SA
    NUCLEAR INSTRUMENTS & METHODS, 1980, 176 (03): : 579 - 582
  • [39] A CR-39 based thermal neutron dosimeter
    Ahmad, N
    Matiullah
    Kenawy, MA
    RADIATION MEASUREMENTS, 1997, 28 (1-6) : 413 - 414
  • [40] CR-39 based thermal neutron dosimeter
    Cent for Nuclear Studies, Islamabad, Pakistan
    Radiat Meas, 1-6 (413-414):