Electromagnetic Immunity of Pacemakers and Implantable Defibrillators to Wireless Power Transfer Systems for Automotive: A Provocative Study

被引:2
|
作者
Vivarelli, Cecilia [1 ,2 ]
Censi, Federica [1 ]
Calcagnini, Giovanni [1 ]
Freschi, Fabio [3 ]
Giaccone, Luca [3 ,4 ]
Canova, Aldo [3 ]
Mattei, Eugenio [1 ]
机构
[1] Italian Natl Inst Hlth ISS, Dept Cardiovasc, Endocrine Metab Dis & Aging, I-00161 Rome, Italy
[2] Univ Roma Tor Vergata, Dept Ingn Civile & Ingn Informat DICII, I-00173 Rome, Italy
[3] Politecn Torino, Dept Energy Galileo Ferraris, I-00173 Turin, Italy
[4] Univ Queensland, Sch Informat Technol & Elect Engn, Brisbane, Qld 4072, Australia
关键词
Coils; Lead; Phantoms; Electromagnetic interference; Standards; Torso; Sensors; Electromagnetic compatibility (EMC) and interference; immunity testing; implantable defibrillator; pacemakers (PMs); wireless power transmission; INTERFERENCE;
D O I
10.1109/TEMC.2023.3338453
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Wireless power transfer (WPT) in automotive is an emerging technology that utilizes coupled coils to transfer power through time-varying magnetic fields, enabling the charging of electric vehicles. Like any electromagnetic field source, WPT systems can potentially interfere with other electronic devices, impacting their proper functioning. This risk requires evaluation, especially when the affected device is a cardiac implantable electronic device (CIED), as malfunctioning can pose serious hazards to the patient's safety. Currently, WPT technology is in its early stages and has not been explicitly considered during the drafting of current CIED regulations. Consequently, predicting the behavior of a CIED when in proximity to a WPT system is not straightforward, and new experimental data are necessary. The objective of this article is to conduct in-vitro testing on a representative sample of pacemakers (PMs) and implantable cardioverter/defibrillators exposed to a magnetic field at the frequency of WTP systems, with field levels gradually increasing up to ICNIRP occupational reference levels. Compared to previous studies, the main novelty is to assess, on 11 commercial CIEDs, a safety margin for the exposure to the magnetic field dispersed by a WPT system, under worst-case exposure condition. Over a total of 51 conducted tests 24 interfering events were observed: only 1 occurred with a magnetic field at 50 mu T while all the others were observed with higher field levels, at 75 and 100 mu T. Thus, the results of this research demonstrate that the risk of interference is very low. Electromagnetic interference phenomena were observed only under worst-case conditions (e.g., maximum sensitivity of the device, maximum lead loop area, pulsed modulation, and unipolar configuration for PMs) and at exposure levels above the ICNIRP general population reference levels.
引用
收藏
页码:97 / 107
页数:11
相关论文
共 50 条
  • [31] Evaluation of Electromagnetic Exposure in Wireless Power Transfer Systems for Electric Vehicles
    Bang, Kyeung-Won
    Park, Sang-Wook
    Bae, Hong-Guk
    Lee, Byeong-Yoon
    Oh, Hui-Myoung
    Park, Chang-Un
    Baek, Seung-Hyuk
    [J]. JOURNAL OF ELECTROMAGNETIC ENGINEERING AND SCIENCE, 2024, 24 (01): : 34 - 41
  • [32] Non-thoracic MRI does not cause problems with pacemakers or implantable defibrillators, study finds
    Mayor, Susan
    [J]. BMJ-BRITISH MEDICAL JOURNAL, 2017, 356
  • [33] SAFETY OF OMOM CAPSULE ENDOSCOPY IN PATIENTS WITH PACEMAKERS AND IMPLANTABLE CARDIAC DEFIBRILLATORS: AN "IN VIVO" ELECTROPHYSIOLOGICAL STUDY
    Moneghini, D.
    Lipari, A.
    Missale, G.
    Bozzi, R.
    Minelli, L.
    Cengia, G.
    Bontempi, L.
    Curnis, A.
    Cestari, R.
    [J]. DIGESTIVE AND LIVER DISEASE, 2014, 46 : S33 - S34
  • [34] Immunity of Automotive Power Line Communication Systems
    Zeichner, Alexander
    Frei, Stephan
    [J]. IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2016, 58 (04) : 1289 - 1296
  • [35] A Pilot Study of a Mindfulness Based Stress Reduction Program in Adolescents with Implantable Cardioverter Defibrillators or Pacemakers
    Vicki A. Freedenberg
    Sue A. Thomas
    Erika Friedmann
    [J]. Pediatric Cardiology, 2015, 36 : 786 - 795
  • [36] A Pilot Study of a Mindfulness Based Stress Reduction Program in Adolescents with Implantable Cardioverter Defibrillators or Pacemakers
    Freedenberg, Vicki A.
    Thomas, Sue A.
    Friedmann, Erika
    [J]. PEDIATRIC CARDIOLOGY, 2015, 36 (04) : 786 - 795
  • [37] Electromagnetic interference from wireless video-capsule endoscopy on implantable cardioverter-defibrillators
    Dubner, Sergio
    Dubner, Yael
    Rubio, Horacio
    Goldin, Eran
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2007, 30 (04): : 472 - 475
  • [38] Design of High Efficiency Planar Spiral Coil for Implantable Wireless Power Transfer Systems
    Zhao, Weicheng
    Mai, Songping
    [J]. 2022 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS (ISCAS 22), 2022, : 147 - 151
  • [39] Study of Electromagnetic Immunity of Motors used in Automotive Applications
    Gliga, Marian
    Munteanu, Calin
    Andreica, Sergiu
    Pacurar, Claudia
    Constantinescu, Claudia
    Giurgiuman, Adina
    [J]. 2019 INTERNATIONAL CONFERENCE ON ELECTROMECHANICAL AND ENERGY SYSTEMS (SIELMEN), 2019,
  • [40] Electromagnetic Compatibility Evaluation of Wireless Charging Systems for Public Spaces: Wireless Power Transfer for Taxis
    David, Amelie
    Tiemann, Myrel
    Haussmann, Norman
    Stroka, Steven
    Clemens, Markus
    Schmuelling, Benedikt
    [J]. IEEE INDUSTRY APPLICATIONS MAGAZINE, 2024, 30 (01) : 59 - 67