Improving ultrasound-based brachytherapy needle conspicuity by applying an echogenic coating

被引:5
|
作者
Brost, Eric E. [1 ]
Stish, Bradley J. [1 ]
Lee, Christine U. [2 ]
Urban, Matthew W. [1 ]
Deufel, Christopher L. [1 ]
机构
[1] Mayo Clin Rochester, Dept Radiat Oncol, Rochester, MN 55902 USA
[2] Mayo Clin Rochester, Dept Radiol, Rochester, MN USA
关键词
brachytherapy applicators; prostate brachytherapy; ultrasound-guided brachytherapy; ULTRASONOGRAPHIC ECHOGENICITY; HDR-BRACHYTHERAPY; PROSTATE; IDENTIFICATION; VISIBILITY; STANDARD;
D O I
10.1002/mp.16138
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundApplicator conspicuity in ultrasound-guided brachytherapy procedures is commonly impaired by imaging artifacts or non-ideal imaging geometry, which can slow down applicator position digitization and increase the geometric uncertainty of the delivered dose distribution. PurposeThe purpose of this study was to improve the conspicuity of high-dose rate (HDR) brachytherapy needles under B-mode ultrasound imaging by applying an echogenic surface coating. Our hypothesis was that an echogenic coating would reduce artifacts and improve needle visualization within regions of signal degradation. MethodsIn this study, 17-gauge, 25-cm long titanium HDR brachytherapy needles were coated with acoustically reflective microspheres over a 2.5 cm region starting from the needle tip. Three coating thicknesses (27 mu m, 40 mu m, 64 mu m) were compared against an uncoated control needle. The coated and uncoated needles were imaged using B-mode ultrasound in a tissue-equivalent prostate phantom and in a cadaverous male pelvis using a transrectal probe. Needle conspicuity was assessed under multiple conditions: a single needle implant, an implant with multiple needles between the probe and the needle of interest, and an angled needle implant. All images were assessed qualitatively for needle conspicuity and the presence of artifacts and quantitatively using grey-scale image intensity values. ResultsThe 64 mu m echogenic coating reduced the magnitude of reverberation artifacts by 31 +/- 14% and comet tail artifacts by 40%-70%. The echogenic coating also improved needle contrast, measured by the relative differences in signal intensity compared with the adjacent environment, when needles were angled up to 30 degrees with respect to the transducer probe in the cadaver. The improvements in conspicuity and artifact reduction increased with increasing coating thickness. The performance of the needles coated with the 64 mu m thickness was qualitatively superior and yielded high-contrast, well-circumscribed signals in the cadaverous male pelvis, even under situations where a needle was acoustically shadowed by multiple other needles. ConclusionsAn echogenic surface coating reduced imaging artifacts and improved needle conspicuity under realistic clinical conditions for ultrasound-based prostate or gynecological brachytherapy. The improved conspicuity has the potential to improve the efficiency of needle placement and the accuracy of needle position digitization during brachytherapy procedures.
引用
收藏
页码:1418 / 1427
页数:10
相关论文
共 50 条
  • [1] Improving brachytherapy needle ultrasound conspicuity with an echogenic coating
    Brost, E.
    Stish, B.
    Deufel, C.
    RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S88 - S89
  • [2] A power Doppler ultrasound-based intraoperative needle tip identification method for interstitial brachytherapy
    Orlando, Nathan
    Snir, Jonatan
    Barker, Kevin
    Hoover, Doug
    Fenster, Aaron
    MEDICAL PHYSICS, 2019, 46 (11) : 5398 - 5398
  • [3] Ultrasound-based conformal planning for gynaecological brachytherapy
    van Dyk, S.
    Bernshaw, D.
    JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2008, 52 (01): : 77 - 84
  • [4] Echogenic Surface Enhancements for Improving Needle Visualization in Ultrasound A PRISMA Systematic Review
    Hovgesen, Caroline Harder
    Wilhjelm, Jens E.
    Vilmann, Peter
    Kalaitzakis, Evangelos
    JOURNAL OF ULTRASOUND IN MEDICINE, 2022, 41 (02) : 311 - 325
  • [5] Echogenic polymer coating: Does, it improve needle visualization in ultrasound-guided biopsies?
    Bergin, DC
    Pappas, JN
    Hwang, JJ
    Sheafor, DH
    Paulson, EK
    RADIOLOGY, 2001, 221 : 260 - 260
  • [6] Improving ultrasound-based prostate volume estimation
    Saro Aprikian
    Murilo Luz
    Fadi Brimo
    Eleonora Scarlata
    Lucie Hamel
    Fabio L. Cury
    Simon Tanguay
    Armen G. Aprikian
    Wassim Kassouf
    Simone Chevalier
    BMC Urology, 19
  • [7] Improving the accuracy of ultrasound-based localisation systems
    Hubert Piontek
    Matthias Seyffer
    Jörg Kaiser
    Personal and Ubiquitous Computing, 2007, 11 : 439 - 449
  • [8] Improving ultrasound-based prostate volume estimation
    Aprikian, Saro
    Luz, Murilo
    Brimo, Fadi
    Scarlata, Eleonora
    Hamel, Lucie
    Cury, Fabio L.
    Tanguay, Simon
    Aprikian, Armen G.
    Kassouf, Wassim
    Chevalier, Simone
    BMC UROLOGY, 2019, 19 (1) : 68
  • [9] Improving the accuracy of ultrasound-based localisation systems
    Piontek, Hubert
    Seyffer, Matthias
    Kaiser, Joerg
    PERSONAL AND UBIQUITOUS COMPUTING, 2007, 11 (06) : 439 - 449
  • [10] Improving the accuracy of ultrasound-based localisation systems
    Piontek, H
    Seyffer, M
    Kaiser, J
    LOCATION- AND CONTEXT-AWARENESS, PROCEEDINGS, 2005, 3479 : 132 - 143