Non-Enzymatic Electrochemical Sensing of Glucose with a Carbon Black/Polyaniline/Silver Nanoparticle Composite

被引:2
|
作者
Pinon-Balderrama, Claudia Ivone [1 ]
Hernandez-Escobar, Claudia Alejandra [1 ]
Reyes-Lopez, Simon Yobanni [2 ]
Conejo-Davila, Alain Salvador [3 ]
Estrada-Monje, Anayansi [3 ]
Zaragoza-Contreras, Erasto Armando [1 ]
机构
[1] SC Adv Mat Res Ctr, Ctr Invest Mat Avanzados, Miguel Cervantes Saavedra 120,Complejo Ind Chihuah, Chihuahua 31136, Chihuahua, Mexico
[2] Autonomous Univ Juarez City, Univ Autonoma Ciudad Juarez, Inst Biomed Sci, Envolvente PRONAF & Estocolmo S N, Juarez 32310, Chihuahua, Mexico
[3] Appl Innovat Ctr Adv Technol, Omega 201 Colonia Ind Delta, Leon 37545, Guanajuato, Mexico
关键词
carbon black; polyaniline; electrochemical glucose sensor; silver nanoparticles; GRAPHENE-OXIDE; MODIFIED ELECTRODES; BLACK; NANOTUBES; POLYANILINE; BIOSENSOR; POLYMER; NANOCOMPOSITES; PERFORMANCE; OXIDATION;
D O I
10.3390/chemosensors12020026
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The present work describes the synthesis of an electroactive nanocomposite consisting of carbon black (CB) and polyaniline (PANI) obtained by in situ oxidative polymerization. Monomer P1 was used as a polyaniline precursor. P1 has surfactant properties that allow obtaining core-shell structures dispersed in an aqueous medium. The nanocomposite, together with silver nanoparticles (AgNPs) as an electrocatalytic element, was used to modify the surface of a glassy carbon electrode (GCE) for glucose detection. Electroactive areas were calculated using the Randles-Sevick equation. The results showed that the CB-PANI.1-1/AgNP nanocomposite exhibited a larger electroactive surface area (0.3451 cm2) compared to AgNP alone (0.0973 cm2) or the CB-PANI.1-1 composite (0.2989 cm2). Characterization of CB-PANI.1-1/AgNP, by cyclic voltammetry in the presence of glucose, showed a new oxidation peak with a maximum current close to 0.7 V due to the oxidation of glucose to gluconolactone. The amperometry test at 0.7 V showed a linear response with R2 of 0.999 as a function of the analyte concentration. The glucose sensor presented a linear detection range of 1 to 10 mM, a sensitivity of 41 mu A mM-1 cm-2, and a limit of detection (LOD) of 520 mu M.
引用
收藏
页数:16
相关论文
共 50 条
  • [31] Ordered assemblies of silver nanoparticles on carbon nitride sheets and their application in the non-enzymatic sensing of hydrogen peroxide and glucose
    Kundu, Manas Kumar
    Sadhukhan, Mriganka
    Barman, Sudip
    JOURNAL OF MATERIALS CHEMISTRY B, 2015, 3 (07) : 1289 - 1300
  • [32] Incorporation of nickel particles into a polyaniline thin film for non-enzymatic glucose sensing in alkaline medium
    Belgherbi, Ouafia
    Messoudi, Meriem
    Bezi, Hamza
    Seid, Lamria
    Chouder, Dalila
    Lamiri, Leila
    Tounsi, Assia
    Akhtar, M. Saeed
    Saeed, M. A.
    JOURNAL OF APPLIED ELECTROCHEMISTRY, 2024, 54 (04) : 851 - 863
  • [33] Electrodeposition of hydroxyapatite on nickel foam and further modification with conductive polyaniline for non-enzymatic glucose sensing
    Jia, Linan
    Wei, Xubing
    Lv, Liulin
    Zhang, Xiaoxue
    Duan, Xinxin
    Xu, Yan
    Liu, Kaili
    Wang, Jun
    ELECTROCHIMICA ACTA, 2018, 280 : 315 - 322
  • [34] Lithium Niobate Perovskite as the Support for Silver Nanoparticles for Non-Enzymatic Electrochemical Detection of Glucose
    Pinon-Balderrama, Claudia Ivone
    Manriquez-Tristan, Atenea
    Maldonado-Orozco, Maria Cristina
    Hernandez-Escobar, Claudia Alejandra
    Reyes-Lopez, Simon Yobanny
    Espinosa-Cristobal, Leon Francisco
    Zaragoza-Contreras, Erasto Armando
    CHEMOSENSORS, 2024, 12 (10)
  • [35] A Review on Non-Enzymatic Electrochemical Biosensors of Glucose Using Carbon Nanofiber Nanocomposites
    Mohammadpour-Haratbar, Ali
    Mohammadpour-Haratbar, Saeid
    Zare, Yasser
    Rhee, Kyong Yop
    Park, Soo-Jin
    BIOSENSORS-BASEL, 2022, 12 (11):
  • [36] Electrochemical carbon-based sensors for non-enzymatic uric acid sensing
    Joshi, Anju
    Slaughter, Gymama
    MICROCHEMICAL JOURNAL, 2025, 208
  • [37] Electrochemical Characterization of Modified Glassy Carbon Electrodes for Non-Enzymatic Glucose Sensors
    Mazurkow, Julia Maria
    Kusior, Anna
    Radecka, Marta
    SENSORS, 2021, 21 (23)
  • [38] Conductive Polyaniline Nanosheets (CPANINS) for a non-enzymatic glucose sensor
    Kailasa, Saraswathi
    Geeta, B.
    Jayarambabu, N.
    Reddy, R. Kiran Kumar
    Sharma, Shubham
    Rao, K. Venkateswara
    MATERIALS LETTERS, 2019, 245 : 118 - 121
  • [39] A silver nanowire aerogel for non-enzymatic glucose detection
    Ma, Shichao
    Yuan, Xichen
    Yin, Xu
    Yang, Yang
    Ren, Li
    MICROCHEMICAL JOURNAL, 2023, 195
  • [40] Development of an electrochemical nanoplatform for non-enzymatic glucose sensing based on Cu/ZnO nanocomposite
    El Golli, Asma
    Echabaane, Mosaab
    Dridi, Cherif
    MATERIALS CHEMISTRY AND PHYSICS, 2022, 280