Phantom study for 90Y liver radioembolization dosimetry with a long axial field-of-view PET/CT

被引:2
|
作者
Mercolli, Lorenzo [1 ]
Zeimpekis, Konstantinos [1 ]
Prenosil, George A. [1 ]
Sari, Hasan [2 ]
Rathke, Hendrik G. [1 ]
Rominger, Axel [1 ]
Shi, Kunagyu [1 ]
机构
[1] Bern Univ Hosp, Univ Bern, Dept Nucl Med, Inselspital, Freiburgstr 18, CH-3010 Bern, Switzerland
[2] Adv Clin Imaging Technol, Siemens Healthcare AG, Lausanne, Switzerland
关键词
Dosimetry; Yttrium-90; Long axial field-of-view PET/CT; Monte Carlo simulations;
D O I
10.1016/j.ejmp.2024.103296
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The physical properties of yttrium -90 (90Y) allow for imaging with positron emission tomography/computed tomography (PET/CT). The increased sensitivity of long axial field -of -view (LAFOV) PET/CT scanners possibly allows to overcome the small branching ratio for positron production from 90Y decays and to improve for the post -treatment dosimetry of 90Y of selective internal radiation therapy. Methods: For the challenging case of an image quality body phantom, we compare a full Monte Carlo (MC) dose calculation with the results from the two commercial software packages Simplicit90Y and Hermes. The voxel dosimetry module of Hermes relies on the 90Y images taken with a LAFOV PET/CT, while the MC and Simplicit90Y dose calculations are image independent. Results: The resulting doses from the MC calculation and Simplicit90Y agree well within the error margins. The image -based dose calculation with Hermes, however, consistently underestimates the dose. This is due to the mismatch of the activity distribution in the PET images and the size of the volume of interest. We found that only for the smallest phantom sphere there is a statistically significant dependence of the Hermes dose on the image reconstruction parameters and scan time. Conclusion: Our study shows that Simplicit90Y's local deposition model can provide a reliable dose estimate. On the other hand, the image based dose calculation suffers from the suboptimal reconstruction of the 90Y distribution in small structures.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Optimization of Y-90 Radioembolization Imaging for Post-Treatment Dosimetry on a Long Axial Field-of-View PET/CT Scanner
    Linder, Pia M.
    Lan, Wenhong
    Trautwein, Nils F.
    Brosch-Lenz, Julia
    von Beschwitz, Sebastian
    Kupferschlaeger, Juergen
    Reischl, Gerald
    Groezinger, Gerd
    Dittmann, Helmut
    la Fougere, Christian
    Schmidt, Fabian P.
    DIAGNOSTICS, 2023, 13 (22)
  • [2] Liver Dose Calculation for 90Y Radioembolization with PET/CT
    Yu, N.
    Xia, P.
    MEDICAL PHYSICS, 2012, 39 (06) : 3695 - 3695
  • [3] Evaluation of long axial field-of-view (LAFOV) PET/CT for post-treatment dosimetry in Yttrium-90 radioembolization of liver tumors: a comparative study with conventional SPECT imaging
    Zeimpekis, Konstantinos G.
    Sari, Hasan
    Goezluegoel, Nasir
    Achangwa, Ngwe Rawlings
    Shi, Kuangyu
    Schindewolf, Marc
    Afshar-Oromieh, Ali
    Rominger, Axel
    Seifert, Robert
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2025, 52 (04) : 1460 - 1471
  • [4] Phantom validation of quantitative Y-90 PET/CT-based dosimetry in liver radioembolization
    Marco D’Arienzo
    Maria Pimpinella
    Marco Capogni
    Vanessa De Coste
    Luca Filippi
    Emiliano Spezi
    Nick Patterson
    Francesca Mariotti
    Paolo Ferrari
    Paola Chiaramida
    Michael Tapner
    Alexander Fischer
    Timo Paulus
    Roberto Pani
    Giuseppe Iaccarino
    Marco D’Andrea
    Lidia Strigari
    Oreste Bagni
    EJNMMI Research, 7
  • [5] Phantom validation of quantitative Y-90 PET/CT-based dosimetry in liver radioembolization
    D'Arienzo, Marco
    Pimpinella, Maria
    Capogni, Marco
    De Coste, Vanessa
    Filippi, Luca
    Spezi, Emiliano
    Patterson, Nick
    Mariotti, Francesca
    Ferrari, Paolo
    Chiaramida, Paola
    Tapner, Michael
    Fischer, Alexander
    Paulus, Timo
    Pani, Roberto
    Iaccarino, Giuseppe
    D'Andrea, Marco
    Strigari, Lidia
    Bagni, Oreste
    EJNMMI RESEARCH, 2017, 7
  • [6] Radioembolization Dosimetry with Total-Body 90Y PET
    Costa, Gustavo
    Spencer, Benjamin
    Omidvari, Negar
    Foster, Cameron
    Rusnak, Michael
    Hunt, Heather
    Caudle, Denise T.
    Pillai, Rex T.
    Vu, Catherine Tram
    Roncali, Emilie
    JOURNAL OF NUCLEAR MEDICINE, 2022, 63 (07) : 1101 - 1107
  • [7] The Impact of Image Reconstruction Bias on PET/CT 90Y Dosimetry After Radioembolization
    Walrand, Stephan
    Hesse, Michel
    Renaud, Lhommel
    Jamar, Francois
    JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (03) : 494 - 495
  • [8] The Impact of Image Reconstruction Bias on PET/CT 90Y Dosimetry After Radioembolization
    Tapp, Katie N.
    Lea, William B.
    Johnson, Matthew S.
    Tann, Mark
    Fletcher, James W.
    Hutchins, Gary D.
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55 (09) : 1452 - 1458
  • [9] Y-90 PET/CTwith Long Axial Field-Of-View Digital Detectors
    Beijst, C.
    Lopez, B.
    de Jong, H.
    Kappadath, S.
    MEDICAL PHYSICS, 2019, 46 (06) : E194 - E195
  • [10] Impact of image reconstruction algorithm and acquisition time for 90Y NEMA phantom acquisitions for radioembolization treatment on a 24 cm-axial field of view PET/CT
    Macis, C.
    Bisello, F.
    Della Gala, G.
    Santoro, M.
    Zagni, F.
    Strolin, S.
    Civollani, S.
    Maccagnani, M.
    Fanti, S.
    Strigari, L.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2024, 51 : S710 - S711