Optimization of VO2 and VCO2 outputs for the calculation of resting metabolic rate using a portable indirect calorimeter

被引:1
|
作者
Leal-Martin, Javier [1 ,2 ,3 ]
Manas, Asier [1 ,2 ,3 ,4 ,5 ]
Alfaro-Acha, Ana [2 ,3 ,6 ]
Garcia-Garcia, Francisco Jose [2 ,3 ,6 ]
Ara, Ignacio [1 ,2 ,3 ]
机构
[1] Univ Castilla La Mancha, Fac Sports Sci, GENUD Toledo Res Grp, Toledo, Spain
[2] Inst Salud Carlos III, CIBER Fragil & Envejecimiento Saludable CIBERFES, Madrid, Spain
[3] Inst Invest Sanitaria Castilla La Mancha IDISCAM, Junta Comun Castilla La Mancha JCCM, Castilla La Mancha, Spain
[4] Ctr UCM ISCIII Human Evolut & Behav, Madrid, Spain
[5] Univ Complutense Madrid, Fac Educ, Madrid, Spain
[6] Hosp Virgen Del Valle, Complejo Hosp Univ Toledo, Geriatr Dept, Toledo, Spain
关键词
fitting equation; indirect calorimetry; metabolic device; oxygen uptake; resting energy expenditure; ENERGY-EXPENDITURE; GAS-EXCHANGE; EQUIVALENT; STABILITY; EXERCISE; VALIDITY; HEALTHY;
D O I
10.1111/sms.14425
中图分类号
G8 [体育];
学科分类号
04 ; 0403 ;
摘要
This study aimed to compare the Cosmed K5 portable indirect calorimeter, using the mixing chamber mode and face mask, with a stationary metabolic cart when measuring the resting metabolic rate (RMR) and to derive fitting equations if discrepancies are observed. Forty--three adults (18--84 years) were assessed for their RMR for two 30--min consecutive and counterbalanced periods using a Cosmed K5 and an Oxycon Pro. Differences among devices were tested using paired sample Student's t--tests, and correlation and agreement were assessed using Pearson's correlation coefficients, intraclass correlation coefficient and Bland--Altman plots. Forward stepwise multiple linear regression models were performed to develop fitting equations for estimating differences among devices when assessing oxygen uptake (VO(2)diff, mL center dot min-1) and carbon dioxide production (VCO2diff, mL center dot min-1). Furthermore, the Oxycon Pro was tested before being confirmed as a reference device. Significant differences between devices were found in most metabolic and ventilatory parameters, including the primary outcomes of VO2 and VCO2. These differences showed an overestimation of the Cosmed K-5 in all metabolic outcomes, except for Fat, when compared to the Oxycon Pro. When derived fitting equations were applied (VO2diff - 139.210 + 0.786 [weight, kg] + 1.761 [height, cm] --0.941 [Cosmed K-5 VO2, mL center dot min(-1)]; VCO(2)diff - 86.569 + 0.548 [weight, kg] + 0.915 [height, cm] --0.728 [Cosmed K-5 VCO2, mL center dot min(-1)]), differences were minimized, and agreement was maximized. This study provides fitting equations which allow the use of the Cosmed K5 for reasonably optimal RMR determinations.
引用
收藏
页码:1648 / 1660
页数:13
相关论文
共 50 条
  • [1] Validity and reproducibility of a whole-room indirect calorimeter for the estimation of VO2, VCO2, and resting metabolic rate
    Henriksen, Hege Berg
    Henriksen, Christine
    de Sousa, Ana Rita Sequeira
    Alavi, Dena Treider
    Augestad, Elin Maria Sandstad
    Rising, Russell
    Doerner, Rebecca
    Haegele, Franziska Anna
    Bosy-Westphal, Anja
    Blomhoff, Rune
    Ulven, Stine Marie
    Olsen, Thomas
    PHYSIOLOGICAL REPORTS, 2023, 11 (07):
  • [2] VO2 and VCO2 variabilities through indirect calorimetry instrumentation
    Cadena-Mendez, Miguel
    Escalante-Ramirez, Boris
    Azpiroz-Leehan, Joaquin
    Infante-Vazquez, Oscar
    SPRINGERPLUS, 2013, 2
  • [3] THE VALIDITY OF 24-H VO2 AND VCO2 MEASUREMENTS WITH A WHOLE-BODY INDIRECT CALORIMETER
    BOUCHARD, G
    WHITE, MD
    BUEMANN, B
    ALMERAS, N
    DESPRES, JP
    BOUCHARD, C
    TREMBLAY, A
    FASEB JOURNAL, 1995, 9 (03): : A461 - A461
  • [4] TRENDS - MEASURING VO2/VCO2
    不详
    JOURNAL OF EQUINE VETERINARY SCIENCE, 1991, 11 (03) : 181 - 182
  • [5] EFFECT OF DIET AND METABOLIC-RATE ON OPEN CIRCUIT CALCULATIONS OF VO2 AND VCO2
    MUSCH, TI
    BROOKS, GA
    RESEARCH QUARTERLY, 1976, 47 (04): : 731 - 740
  • [6] EFFECT OF DIET AND METABOLIC-RATE ON OPEN CIRCUIT CALCULATIONS OF VO2 AND VCO2
    MUSCH, TI
    BROOKS, GA
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 1974, 6 (01): : 79 - 79
  • [7] 12 hour fast significantly alters resting VO2, VCO2 and RER as measured by indirect calorimetry
    Zwemer, CF
    O'Malley, MJ
    Powell, NM
    FASEB JOURNAL, 2005, 19 (04): : A675 - A675
  • [8] Steady state condition in the measurement of VO2 and VCO2 by Indirect Calorimetry
    Cadena, M.
    Sacristan, E.
    Infante, O.
    Escalante, B.
    Rodriguez, F.
    2005 27TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2005, : 7773 - 7776
  • [9] VO2 AND VCO2 FOLLOWING TOURNIQUET DEFLATION
    HOKA, S
    YOSHITAKE, J
    ARAKAWA, S
    OHTA, K
    YAMAOKA, A
    GOTO, T
    ANAESTHESIA, 1992, 47 (01) : 65 - 68
  • [10] Influence of different ventilator modes on VO2 and VCO2 measurements using a compact metabolic monitor
    Briassoulis, George
    Michaeloudi, Evi
    Fitrolaki, Diana-Michaela
    Spanaki, Anna-Maria
    Briassouli, Efrossini
    NUTRITION, 2009, 25 (11-12) : 1106 - 1114