Measurement of serum anti-Mullerian hormone by revised Gen II or automated assay: Reproducibility under various blood/serum storage conditions

被引:0
|
作者
Lee, Joong Yeup [1 ]
Kim, Chung Hyon [1 ]
Choe, Seung-Ah [2 ]
Seo, Soyeon [3 ]
Kim, Seok Hyun [4 ]
机构
[1] Hamchoon Womens Clin, Seoul, South Korea
[2] Korea Univ, Coll Med, Dept Prevent Med, Seoul, South Korea
[3] Samkwang Med Labs, Seoul, South Korea
[4] Seoul Natl Univ, Coll Med, Dept Obstet & Gynecol, 101 Daehak Ro, Seoul 03080, South Korea
关键词
Anti-Mullerian Hormone; Automation; Biological assay; Blood; Reproducibility of results; Serum; MLLERIAN HORMONE; IMMUNOASSAYS; PERFORMANCE; AGREEMENT; STABILITY; QUALITY; BLOOD;
D O I
10.5653/cerm.2022.05687
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Objective: We investigated the agreement between anti-Mullerian hormone (AMH) levels measured with revised Gen II (rev-Gen II) and au-tomated AMH (Access) assays and evaluated the reproducibility of each method under various blood/serum storage conditions. Methods: AMH levels in blood samples from 74 volunteers were measured by rev-Gen II and Access assays under various conditions: imme-diate serum separation and AMH measurement (fresh control); serum stored at -20 degrees C and AMH measured after 48 hours, 1 week, and 2 years; serum stored at 0 to 4 degrees C and AMH measured after 48 hours and 1 week; and blood kept at room temperature and delayed serum sep-aration after 48 hours and 1 week, with immediate AMH measurement.Results: In fresh controls, all rev-Gen II-AMH values were higher than comparable Access-AMH values (difference, 8.3% to 19.7%). AMH levels measured with the two methods were strongly correlated for all sample conditions (r=0.977 to 0.995, all p<0.001). For sera stored at -20 degrees C or 0 to 4 degrees C for 48 hours, Access-AMH values were comparable to control measurements, but rev-Gen II-AMH values were significantly lower. AMH levels in sera stored at -20 degrees C or 0 to 4 degrees C for 1 week were significantly lower than in fresh controls, irrespective of method. Across meth-ods, long-term storage at -20 degrees C for 2 years yielded AMH measurements significantly higher than control values. When serum separation was delayed, rev-Gen II-AMH values were significantly lower than control measurements, but Access-AMH values varied.Conclusion: The rev-Gen II and Access-AMH assays showed varying reproducibility across blood/serum storage conditions, but automated Access yielded superior stability to rev-Gen II.
引用
收藏
页码:107 / 116
页数:10
相关论文
共 45 条
  • [41] Prediction of the lower serum anti-Mullerian hormone threshold for ovarian stimulation prior to in-vitro fertilization using the Elecsys® AMH assay: a prospective observational study
    Grynnerup, A. G.
    Lossl, K.
    Pilsgaard, F.
    Lunding, S. A.
    Storgaard, M.
    Bogstad, J. W.
    Praetorius, L.
    Zedeler, A.
    Bungum, L.
    Andersen, A. Nyboe
    Pinborg, A.
    [J]. REPRODUCTIVE BIOLOGY AND ENDOCRINOLOGY, 2019, 17
  • [42] The lower serum Anti-Mullerian Hormone (S-AMH) threshold measured by the Elecsys assay that predicts development of three mature follicles for In-Vitro Fertilization (IVF)
    Pilsgaard, F.
    Grynnerup, A. G. A.
    Lossl, K.
    Zedeler, A.
    Bogstad, J. W.
    Praetorius, L.
    Bungum, L.
    Andersen, A. Nyboe
    Pinborg, A.
    [J]. HUMAN REPRODUCTION, 2017, 32 : 467 - 468
  • [43] EVALUATION OF AT-HOME SERUM ANTI-MULLERIAN HORMONE TESTING: A HEAD-TO-HEAD COMPARISON OF TWO SELF-ADMINISTERED BLOOD COLLECTION DEVICES.
    Chung, Esther H.
    Silliman, Erin
    Fitzpatrick, Elizabeth
    Jolin, Julie A.
    Brown, Michelle
    Hotaling, James
    Styer, Aaron K.
    Karmon, Anatte E.
    [J]. FERTILITY AND STERILITY, 2022, 118 (05) : E31 - E31
  • [44] Interaction between serum levels of Anti-Mullerian Hormone and the degree of sperm DNA fragmentation measured by sperm chromatin structure assay can be a predictor for the outcome of standard in vitro fertilization
    Zaren, Peter
    Alson, Sara
    Henic, Emir
    Bungum, Mona
    Giwercman, Aleksander
    [J]. PLOS ONE, 2019, 14 (08):
  • [45] Detection of minimal levels of serum anti-Mullerian hormone during follow-up of patients with ovarian granulosa cell tumor by means of a highly sensitive enzyme-linked immunosorbent assay
    Long, WQ
    Ranchin, V
    Pautier, P
    Belville, C
    Denizot, P
    Cailla, H
    Lhommé, C
    Picard, JY
    Bidart, JM
    Rey, R
    [J]. JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 2000, 85 (02): : 540 - 544