Validation of a novel automated system, Fluispotter®, for serial sampling of dried blood spots

被引:1
|
作者
Krogh, Jesper [1 ,2 ]
Plomgaard, Peter [2 ,3 ]
Frikke-Schmidt, Ruth [2 ,3 ]
Velschow, Sten [4 ]
Johannesen, Jesper [2 ,5 ]
Hilsted, Linda Maria [3 ]
Schroder, Malene [4 ]
Feldt-Rasmussen, Ulla [1 ,2 ]
机构
[1] Copenhagen Univ Hosp, Dept Endocrinol & Metab, Rigshosp, Copenhagen, Denmark
[2] Univ Copenhagen, Dept Clin Med, Copenhagen, Denmark
[3] Copenhagen Univ Hosp, Dept Clin Biochem, Rigshosp, Copenhagen, Denmark
[4] Fluisense ApS, Lillerod, Denmark
[5] Copenhagen Univ Hosp Herlev & Gentofte, Dept Pediat, Copenhagen, Denmark
关键词
dried blood spots; cortisol; diurnal; automated sampling; CORTISOL;
D O I
10.1530/EC-23-0087
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Repeated blood sampling is required in certain clinical and research settings, which is currently performed by drawing blood from venous catheters requiring manual handling of each sample at the time of collection. A novel body-worn device for repeated serial samples, Fluispotter & REG;, with automated extraction, collection, and storage of up to 20 venous dried blood spot samples over the course of 20 h may overcome problems with current methods for serial sampling. The purpose of this study was to assess the performance and safety of Fluispotter for the first time in healthy subjects. Fluispotter consists of a cartridge with tubing, a reservoir for flushing solution, pumps and filterpaper, and a multi-lumen catheter placed in the brachial vein. We recruited healthy subjects for testing in an in-hospital setting. Fluispotter was attached by an anesthesiologist to 22 healthy subjects of which 9/22 (40.9%) participants had all 20 samples taken, which was lower than the goal of complete sampling in 80% of the subjects (P = 0.02). The main reason for sample failure was clogging of blood flow which was observed in 11/22 (50%) of the participants. No serious adverse events occurred, and the participants rated the pain from the insertion and the removal of catheter as very low. A cortisol profile showed nadir values at midnight and highest values at 05:00 h. Although full sampling was not successful in all participants, the Fluispotter technology proved safe and highly acceptable to the participants producing the expected cortisol profile without the requirement of staff during sample collection.
引用
收藏
页数:9
相关论文
共 50 条
  • [41] AN AUTOMATED IMMUNORADIOMETRIC ASSAY OF THYROTROPIN (TSH) IN DRIED BLOOD FILTER-PAPER SPOTS
    JOHN, R
    WOODHEAD, JS
    CLINICA CHIMICA ACTA, 1982, 125 (03) : 329 - 340
  • [42] Utilization of dried blood spots within drug discovery: modification of a standard DiLab® AccuSampler® to facilitate automatic dried blood spot sampling
    Clark, G. T.
    Giddens, G.
    Burrows, L.
    Strand, C.
    LABORATORY ANIMALS, 2011, 45 (02) : 124 - 126
  • [43] Methylmalonic acid concentration in dried blood spots - validation of a novel method for vitamin B12 status assessment
    Schroder, Theresa H.
    Quay, Teo A. W.
    Lamers, Yvonne
    FASEB JOURNAL, 2013, 27
  • [44] Validation of a Ligand Binding Assay Using Dried Blood Spot Sampling
    Burns, Daniel
    Brunner, Laura
    Rajendran, Surendran
    Johnson, Beth
    Ma, Mark
    Wang, Jin
    AAPS JOURNAL, 2013, 15 (01): : 123 - 131
  • [45] Validation of a Ligand Binding Assay Using Dried Blood Spot Sampling
    Daniel Burns
    Laura Brunner
    Surendran Rajendran
    Beth Johnson
    Mark Ma
    Jin Wang
    The AAPS Journal, 2013, 15 : 123 - 131
  • [46] Validation of dried blood spot sampling in therapeutic drug monitoring of cyclosporin
    Wilhelm, A. J.
    Klijn, A.
    Burger, J. C. G.
    Veldkamp, A., I
    Janssen, J. J. W. M.
    Swart, N.
    THERAPEUTIC DRUG MONITORING, 2011, 33 (04) : 545 - 545
  • [47] Retinol concentrations in capillary dried blood spots from healthy volunteers: method validation
    Craft, NE
    Bulux, J
    Valdez, C
    Li, YK
    Solomons, NW
    AMERICAN JOURNAL OF CLINICAL NUTRITION, 2000, 72 (02): : 450 - 454
  • [48] Concentration Determination of >200 Proteins in Dried Blood Spots for Biomarker Discovery and Validation
    Eshghi, Azad
    Pistawka, Adam J.
    Liu, Jun
    Chen, Michael
    Sinclair, Nicholas J. T.
    Hardie, Darryl B.
    Elliott, Monica
    Chen, Lei
    Newman, Rachael
    Mohammed, Yassene
    Borchers, Christoph H.
    MOLECULAR & CELLULAR PROTEOMICS, 2020, 19 (03) : 540 - 553
  • [49] Validation of dried blood spots for capturing hepatitis C virus diversity for genomic surveillance
    Tully, Damien C.
    Power, Karen A.
    Sarette, Jacklyn
    Stopka, Thomas J.
    Friedmann, Peter D.
    Korthuis, P. Todd
    Cooper, Hannah
    Young, April M.
    Seal, David W.
    Westergaard, Ryan P.
    Allen, Todd M.
    JOURNAL OF VIRAL HEPATITIS, 2024, 31 (05) : 266 - 270
  • [50] A SYSTEM FOR SERIAL BLOOD-SAMPLING IN CONSCIOUS RATS
    FORSLING, ML
    WELLS, T
    JOURNAL OF PHYSIOLOGY-LONDON, 1987, 390 : P2 - P2