Validation of a novel automated system, Fluispotter®, for serial sampling of dried blood spots

被引:1
|
作者
Krogh, Jesper [1 ,2 ]
Plomgaard, Peter [2 ,3 ]
Frikke-Schmidt, Ruth [2 ,3 ]
Velschow, Sten [4 ]
Johannesen, Jesper [2 ,5 ]
Hilsted, Linda Maria [3 ]
Schroder, Malene [4 ]
Feldt-Rasmussen, Ulla [1 ,2 ]
机构
[1] Copenhagen Univ Hosp, Dept Endocrinol & Metab, Rigshosp, Copenhagen, Denmark
[2] Univ Copenhagen, Dept Clin Med, Copenhagen, Denmark
[3] Copenhagen Univ Hosp, Dept Clin Biochem, Rigshosp, Copenhagen, Denmark
[4] Fluisense ApS, Lillerod, Denmark
[5] Copenhagen Univ Hosp Herlev & Gentofte, Dept Pediat, Copenhagen, Denmark
关键词
dried blood spots; cortisol; diurnal; automated sampling; CORTISOL;
D O I
10.1530/EC-23-0087
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Repeated blood sampling is required in certain clinical and research settings, which is currently performed by drawing blood from venous catheters requiring manual handling of each sample at the time of collection. A novel body-worn device for repeated serial samples, Fluispotter & REG;, with automated extraction, collection, and storage of up to 20 venous dried blood spot samples over the course of 20 h may overcome problems with current methods for serial sampling. The purpose of this study was to assess the performance and safety of Fluispotter for the first time in healthy subjects. Fluispotter consists of a cartridge with tubing, a reservoir for flushing solution, pumps and filterpaper, and a multi-lumen catheter placed in the brachial vein. We recruited healthy subjects for testing in an in-hospital setting. Fluispotter was attached by an anesthesiologist to 22 healthy subjects of which 9/22 (40.9%) participants had all 20 samples taken, which was lower than the goal of complete sampling in 80% of the subjects (P = 0.02). The main reason for sample failure was clogging of blood flow which was observed in 11/22 (50%) of the participants. No serious adverse events occurred, and the participants rated the pain from the insertion and the removal of catheter as very low. A cortisol profile showed nadir values at midnight and highest values at 05:00 h. Although full sampling was not successful in all participants, the Fluispotter technology proved safe and highly acceptable to the participants producing the expected cortisol profile without the requirement of staff during sample collection.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Fluispotter, a novel automated and wearable device for accurate volume serial dried blood spot sampling
    Adhikari, Khem Bahadur
    Rohde, Morten
    Velschow, Sten
    Feldt-Rasmussen, Ulla
    Johannesen, Jesper
    Johnsen, Anders Holten
    BIOANALYSIS, 2020, 12 (10) : 665 - 681
  • [2] Validation of dried blood spots in monitoring toxoplasmosis
    Hegazy, Mamdouh M.
    Hegazy, Mona K.
    Azab, Manar S.
    Nabih, Nairmen
    PATHOGENS AND GLOBAL HEALTH, 2020, 114 (05) : 242 - 250
  • [3] Evaluation of Automated Blood Sampling for Low Volume Serial Sampling and Dried Blood Spot Pharmacokinetic Applications in Mouse and Rat
    Uyeda, C.
    Fide, S. N.
    Mangels, M.
    Phadnis, R.
    Pham, R.
    Xu, G.
    Henne, K.
    JOURNAL OF THE AMERICAN ASSOCIATION FOR LABORATORY ANIMAL SCIENCE, 2011, 50 (05): : 792 - 793
  • [4] Validation of an assay for quantifying ganciclovir in dried blood spots
    Rower, Joseph E.
    Nielson, Chris
    Shi, Kevin
    Park, Albert H.
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2020, 184
  • [5] VALIDATION OF DRIED BLOOD SPOTS FOR CARDIOVASCULAR DISEASE BIOMARKERS
    Samuelsson, Laura B.
    Buxton, Orfeu M.
    Hall, Martica
    McLean, Shakir
    Porter, James H.
    Berkman, Lisa
    Marino, Miguel
    Sembajwe, Grace
    McDade, Thomas
    PSYCHOSOMATIC MEDICINE, 2014, 76 (03): : A42 - A43
  • [6] Validation of a biomarker of bone resorption in dried blood spots
    Eick, G.
    Devlin, M. J.
    Cepon-Robins, T. J.
    Kowal, P.
    Sugiyama, L. S.
    Snodgrass, J. J.
    AMERICAN JOURNAL OF HUMAN BIOLOGY, 2018, 30 (02)
  • [7] Validation of whole genome sequencing from dried blood spots
    Pooja Agrawal
    Shanmukh Katragadda
    Arun K. Hariharan
    Vijayashree Gauribidanur Raghavendrachar
    Arunika Agarwal
    Rashmi Dayalu
    Disha Awasthy
    Sanjay C. Sharma
    Yasodha Kannan Sivasamy
    P. Lakshmana
    Ashwini Shanmugam
    Vamsi Veeramachaneni
    Vaijayanti Gupta
    B. P. Vani
    Lekha Subaiya
    T. S. Syamala
    Ramesh Hariharan
    Vijay Chandru
    David E. Bloom
    BMC Medical Genomics, 14
  • [8] Analytical Validation of Familial Hypercholesterolemia Biomarkers in Dried Blood Spots
    Held, Patrice K.
    Campbell, Kristin
    Wiberley-Bradford, Amy E.
    Lasarev, Michael
    Horner, Vanessa
    Peterson, Amy
    INTERNATIONAL JOURNAL OF NEONATAL SCREENING, 2022, 8 (01)
  • [9] Validation of whole genome sequencing from dried blood spots
    Agrawal, Pooja
    Katragadda, Shanmukh
    Hariharan, Arun K.
    Raghavendrachar, Vijayashree Gauribidanur
    Agarwal, Arunika
    Dayalu, Rashmi
    Awasthy, Disha
    Sharma, Sanjay C.
    Sivasamy, Yasodha Kannan
    Lakshmana, P.
    Shanmugam, Ashwini
    Veeramachaneni, Vamsi
    Gupta, Vaijayanti
    Vani, B. P.
    Subaiya, Lekha
    Syamala, T. S.
    Hariharan, Ramesh
    Chandru, Vijay
    Bloom, David E.
    BMC MEDICAL GENOMICS, 2021, 14 (01)
  • [10] Validation and clinical application of a method to quantify nevirapine in dried blood spots and dried breast-milk spots
    Olagunju, Adeniyi
    Amara, Alieu
    Waitt, Catriona
    Else, Laura
    Penchala, Sujan D.
    Bolaji, Oluseye
    Soyinka, Julius
    Siccardi, Marco
    Back, David
    Owen, Andrew
    Khoo, Saye
    JOURNAL OF ANTIMICROBIAL CHEMOTHERAPY, 2015, 70 (10) : 2816 - 2822