Impact of the automatic tube current modulation (ATCM) system on virtual monoenergetic image quality for dual-source CT: A phantom study

被引:2
|
作者
Dabli, D. [1 ,2 ,3 ]
Durand, Quentin [1 ,2 ]
Frandon, J. [1 ]
de Oliveira, Fabien [1 ]
Pastor, Maxime [1 ]
Beregi, J. P. [1 ]
Greffier, J. [1 ,2 ]
机构
[1] Montpellier Univ, Nimes Univ Hosp, Dept Med Imaging, IMAGINE UR UM 103, F-30029 Nimes, France
[2] Nimes Univ Hosp, Dept Med Phys, F-30029 Nimes 9, France
[3] Montpellier Univ, Nimes Univ Hosp, Dept Med Imaging, CHU Nimes,IMAGINE UR UM 103, Bd Prof Robert Debre, F-30029 Nimes 9, France
关键词
Dual-source CT scan; Dual-energy CT; Virtual monochromatic images; Automatic tube current modulation; Detectability; ITERATIVE RECONSTRUCTION ALGORITHM; ENERGY COMPUTED-TOMOGRAPHY; LIVER-LESIONS; IODINE QUANTIFICATION; MAGNETIC-RESONANCE; EXPOSURE CONTROL; DOSE REDUCTION; OPTIMIZATION; PERFORMANCE; ACCURACY;
D O I
10.1016/j.ejmp.2023.102574
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To assess the impact of the automatic tube current modulation (ATCM) on virtual monoenergetic images (VMIs) quality in dual-source CT(DSCT).Materials and methods: Acquisitions were performed on DSCT using the Mercury phantom. The acquisition pa-rameters for an abdomen-pelvic examination with single-energy CT(SECT) and dual-energy CT(DECT) imaging were used. Acquisitions were performed for each imaging mode using fixed mAs and ATCM. The mAs value was set to obtain a volume CT dose index of 11 mGy in fixed mAs acquisitions. This value was used as the reference mAs in ATCM acquisitions. The noise power spectrum and task-based transfer function at 40,50,60 and 70 keV levels were computed on VMIs and SECT images. The detectability index (d ') was calculated for a lesion with an iodine concentration of 10 mg/mL.Results: The noise magnitude on VMIs was higher with the ATCM system than with fixed mAs for all energy levels and section diameters of 21,26 and 31 cm. The noise texture and spatial resolution were similar between the fixed mAs and ATCM acquisitions for both imaging modes. The d' values were lower for all energy levels with ATCM than with fixed mAs acquisitions for 21 and 26 cm diameters by-39.82 +/- 9.32%, similar at 31 cm diameter-4.13 +/- 0.24% and higher at 36 cm diameter 10.40 +/- 6.69%. It was higher on VMIs at all energy levels compared to SECT images.Conclusions: The ATCM system could be used with DECT imaging to optimize patient exposure without changing the noise texture and spatial resolution of VMIs compared to fixed mAs and SECT.
引用
收藏
页数:8
相关论文
共 50 条
  • [41] Effect of High-Pitch Dual-Source CT to Compensate Motion Artifacts: A Phantom Study
    Farshad-Amacker, Nadia A.
    Alkadhi, Hatem
    Leschka, Sebastian
    Frauenfelder, Thomas
    ACADEMIC RADIOLOGY, 2013, 20 (10) : 1234 - 1239
  • [42] Dual-source CT coronary angiography: factors determining image quality and diagnostic accuracy
    Yerramasu, A.
    Venuraju, S.
    Atwal, S.
    Goodman, D.
    Lipkin, D.
    Lahiri, A.
    EUROPEAN HEART JOURNAL SUPPLEMENTS, 2011, 13 (0A) : A71 - A71
  • [43] Comparison of image quality and radiation dose between combined automatic tube current modulation and fixed tube current technique in CT of abdomen and pelvis
    Lee, Sanghee
    Yoon, Sang-Wook
    Yoo, Seung-Min
    Ji, Young Geon
    Kim, Kyoung Ah
    Kim, Sang Heum
    Lee, Jong Tae
    ACTA RADIOLOGICA, 2011, 52 (10) : 1101 - 1106
  • [44] Image quality optimization and evaluation of linearly mixed images in dual-source, dual-energy CT
    Yu, Lifeng
    Primak, Andrew N.
    Liu, Xin
    McCollough, Cynthia H.
    MEDICAL PHYSICS, 2009, 36 (03) : 1019 - 1024
  • [45] Left Ventricular Function Can Reliably be Assessed From Dual-Source CT Using ECG-Gated Tube Current Modulation
    Mahnken, Andreas H.
    Bruners, Philipp
    Schmidt, Bernhard
    Bornikoel, Christoph
    Flohr, Thomas
    Guenther, Rolf W.
    INVESTIGATIVE RADIOLOGY, 2009, 44 (07) : 384 - 389
  • [46] Dual-Source Versus Single-Source Cardiac CT Angiography: Comparison of Diagnostic Image Quality
    Donnino, Robert
    Jacobs, Jill E.
    Doshi, Jay V.
    Hecht, Elizabeth M.
    Kim, Danny C.
    Babb, James S.
    Srichai, Monvadi B.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 192 (04) : 1051 - 1056
  • [47] A study of CT dose distribution in an elliptical phantom and the influence of automatic tube current modulation in the x-y plane
    Sookpeng, S.
    Martin, C. J.
    Gentle, D. J.
    JOURNAL OF RADIOLOGICAL PROTECTION, 2013, 33 (02) : 461 - 483
  • [48] Optimization of keV-settings in abdominal and lower extremity dual-source dual-energy CT angiography determined with virtual monoenergetic imaging
    Sudarski, Sonja
    Apfaltrer, Paul
    Nance, John W., Jr.
    Schneider, David
    Meyer, Mathias
    Schoenberg, Stefan O.
    Fink, Christian
    Henzler, Thomas
    EUROPEAN JOURNAL OF RADIOLOGY, 2013, 82 (10) : E574 - E581
  • [49] Influence of Contrast Agent Injection Scheme Customized by Dual-Source CT Based on Automatic Tube Voltage Technology on Image Quality and Radiation Dose of Coronary Artery Imaging
    He, Weiling
    Chen, Xin
    Hu, Rui
    Sun, Wenjie
    Tan, Weili
    FRONTIERS IN SURGERY, 2022, 9
  • [50] Radiation dose and physical image quality in 128-section dual-source computed tomographic coronary angiography: a phantom study
    Matsubara, Kosuke
    Koshida, Haruka
    Sakuta, Keita
    Takata, Tadanori
    Horii, Junsei
    Iida, Hiroji
    Koshida, Kichiro
    Ichikawa, Katsuhiro
    Matsui, Osamu
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (05): : 252 - 261