Image-Guided Surgical Device Failures in Functional Endoscopic Sinus Surgery: A MAUDE Analysis

被引:1
|
作者
Hassanin, Samir W. [1 ]
Kshirsagar, Rijul S. [2 ]
Eide, Jacob G. [2 ]
Chang, Jeremy [2 ]
Liang, Jonathan [3 ]
Palmer, James N. [2 ]
Adappa, Nithin D. [2 ]
机构
[1] Rutgers Robert Wood Johnson Med Sch, Dept Med Educ, Piscataway, NJ USA
[2] Univ Penn, Dept Otorhinolaryngol Head & Neck Surg, Perelman Sch Med, Philadelphia, PA 19104 USA
[3] Kaiser Permanente Oakland Med Ctr, Dept Head & Neck Surg, Oakland, CA USA
来源
LARYNGOSCOPE | 2023年 / 133卷 / 06期
关键词
failure; FESS; guidance; IGS; image; image-guided surgery; MAUDE; sinus; COMPLICATIONS;
D O I
10.1002/lary.30296
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Objective Image-guided surgery (IGS) devices have become widely used for anatomic localization during functional endoscopic sinus surgery (FESS). However, there are no studies that analyze the post-market complications associated with IGS device use during FESS. The objective of this study was to better characterize post-market complications associated with the use of IGS devices during sinus surgery. Methods The US Food and Drug Administration's Manufacturer and User Facility Device Experience database was queried for event reports associated with neurological stereotaxic devices utilized in IGS between the dates of January 1, 2016 and December 31, 2020. Medical device reports that were analyzed for this study pertained strictly to FESS. Results There were 1873 reports involving IGS devices for FESS included in this study. Fifty-five reports involved adverse events to patients (2.9%) and 1818 (97.1%) involved device malfunctions. Of the adverse events to patients, the most common included cerebrospinal fluid leakage (45.6%), tissue damage (12.7%), and nervous system injury (3.6%). The most commonly reported device malfunction was imprecision (21.1%). Conclusion IGS devices are widely utilized in FESS. Of the medical device reports between 2016 and 2020, less than 3% resulted in adverse events. Further studies of the infrequent post-market complications of IGS devices used in FESS can help guide surgeons on the risks of their clinical use. Level of Evidence 4-Retrospective database survey without controls Laryngoscope, 2022
引用
收藏
页码:1310 / 1314
页数:5
相关论文
共 50 条
  • [1] Image-guided functional endoscopic sinus surgery
    Olson, G
    Citardi, MJ
    [J]. OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2000, 123 (03) : 188 - 194
  • [2] Adverse Events Associated With Image-Guided Sinus Navigation in Endoscopic Sinus Surgery: A MAUDE Database Analysis
    Orozco, Francis Reyes
    Ulloa, Ruben
    Lin, Matthew
    Xepoleas, Meredith
    Paoletti, Marcus
    Liu, Xuan
    Hur, Kevin
    [J]. OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2023, 168 (03) : 501 - 505
  • [3] Comparative analysis of surgical processes for image-guided endoscopic sinus surgery
    Sugino, Takaaki
    Nakamura, Ryoichi
    Kuboki, Akihito
    Honda, Osamu
    Yamamoto, Masashi
    Ohtori, Nobuyoshi
    [J]. INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2019, 14 (01) : 93 - 104
  • [4] Comparative analysis of surgical processes for image-guided endoscopic sinus surgery
    Takaaki Sugino
    Ryoichi Nakamura
    Akihito Kuboki
    Osamu Honda
    Masashi Yamamoto
    Nobuyoshi Ohtori
    [J]. International Journal of Computer Assisted Radiology and Surgery, 2019, 14 : 93 - 104
  • [5] Image-guided sphenoidotomy in revision functional endoscopic sinus surgery
    Jiang, Rong-San
    Liang, Kai-Li
    [J]. ALLERGY & RHINOLOGY, 2014, 5 (03): : E116 - E119
  • [6] SAFETY AND COST ANALYSIS OF IMAGE-GUIDED FUNCTIONAL ENDOSCOPIC SINUS SURGERY IN CHINA
    Pang, P.
    Tan, B.
    Wang, Z.
    Liu, C.
    Lv, K.
    Xie, H.
    Ou, X.
    Zeng, X.
    Xuan, J.
    [J]. VALUE IN HEALTH, 2022, 25 (12) : S300 - S300
  • [7] Nationwide survey on the use of image-guided functional endoscopic sinus surgery
    Hepworth, Edward J.
    Bucknor, Matthew
    Patel, Ankit
    Vaughan, Winston C.
    [J]. OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2006, 135 (01) : 68 - 73
  • [8] Application of image-guided system to endoscopic sinus surgery
    Murata, H
    Tomoda, K
    [J]. CARS '99: COMPUTER ASSISTED RADIOLOGY AND SURGERY, 1999, 1191 : 651 - 655
  • [9] Endoscopic Video Images Analysis for Surgical Training and Image-Guided Surgery
    Sanchez-Gonzalez, P.
    Oropesa, I.
    Cano, A. M.
    Gaya, F.
    Lamata, P.
    Sanchez-Margallo, F. M.
    del Pozo, F.
    Gomez, E. J.
    [J]. WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 6, 2009, 25 : 251 - +
  • [10] Image-guided sinus surgery
    Gall, RM
    Witterick, IJ
    Hawke, M
    [J]. JOURNAL OF OTOLARYNGOLOGY, 2004, 33 (01): : 22 - 25