Reliability of Visual Field Testing in a Telehealth Setting Using a Head-Mounted Device: A Pilot Study

被引:2
|
作者
McLaughlin, Danielle E. [1 ]
Savatovsky, Eleonore J. [1 ,2 ]
O'Brien, Robert C. [1 ]
Vanner, Elizabeth A. [1 ]
Munshi, Hounsh K. [1 ]
Pham, Anh H. [1 ]
Grajewski, Alana L. [1 ]
机构
[1] Univ Miami, Dept Bascom Palmer Eye Inst, Miller Sch Med, Miami, FL USA
[2] Bascom Palmer Eye Inst, 900 NW 17th St, Miami, FL 33136 USA
关键词
telehealth; glaucoma; perimetry; visual fields; virtual reality; STANDARD AUTOMATED PERIMETRY; GLAUCOMA; EXPERIENCE; VARIABILITY; PROGRESSION; FREQUENCY; ACCURACY; DEFECTS; TIME;
D O I
10.1097/IJG.0000000000002290
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Precis:Monitoring visual fields (VFs) through virtual reality devices proved to have good inter-test and test-retest reliability, as well as easy usability, when self-administered by individuals with and without VF defects in a remote setting.Purpose:To assess the reliability of remote, self-administered VF monitoring using a virtual reality VF (VRVF) device in individuals without ocular disease and with stable VF defects.Materials and Methods:Individuals without ocular disease and with stable defects were recruited. All participants had a baseline standard automated perimetry (SAP) test. Participants tested remotely on a VRVF device for 4 weeks (examinations V1, V2, V3, and V4), with the last 3 unassisted. The mean sensitivities of VRVF results were compared with each other and to SAP results for reliability.Results:A total of 42 eyes from 21 participants were tested on the VRVF device. Participants tested consistently although external factors impacted outcomes. VRVF results were in reasonable agreement with the baseline SAP. Examinations performed by the cohort with stable defects evinced better agreement with SAP examinations (V2, P = 0.79; V3, P = 0.39; V4, P = 0.35) than those reported by the cohort without ocular disease (V2, P = 0.02; V3, P = 0.15; V4, P = 0.22), where the null hypothesis is that the instruments agree. Fixation losses were high and variable in VRVF examinations compared with those of SAP, particularly in certain test takers. Participants considered the device comfortable and easy to use.Conclusions:Self-administered, remote VF tests on a VRVF device showed satisfactory test-retest reliability, good inter-test agreement with SAP, and acceptability by its users. External factors may impact at-home testing and age and visual impairment may hinder fixation. Future studies to expand the sample size and understand inconsistencies in fixation losses are recommended.
引用
收藏
页码:15 / 23
页数:9
相关论文
共 50 条
  • [1] Response to Letter to the Editor: Reliability of Visual Field Testing in a Telehealth Setting Using a Head-Mounted Device: A Pilot Study
    McLaughlin, Danielle E.
    Savatovsky, Eleonore J.
    O'Brien, Robert C.
    Vanner, Elizabeth A.
    Munshi, Hounsh K.
    Pham, Anh H.
    Grajewski, Alana L.
    [J]. JOURNAL OF GLAUCOMA, 2024, 33 (08) : e76 - e77
  • [2] Visual Field Testing in a Telehealth Setting: Remote Perimetry Using a Head-Mounted Device in Normal Eyes
    McLaughlin, Danielle
    Munshi, Hounsh
    Savatovsky, Eleonore
    Vanner, Elizabeth
    Chang, Ta Chen
    Grajewski, Alana L.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2022, 63 (07)
  • [3] Visual Field Testing with Head-Mounted Perimeter 'imo'
    Matsumoto, Chota
    Yamao, Sayaka
    Nomoto, Hiroki
    Takada, Sonoko
    Okuyama, Sachiko
    Kimura, Shinji
    Yamanaka, Kenzo
    Aihara, Makoto
    Shimomura, Yoshikazu
    [J]. PLOS ONE, 2016, 11 (08):
  • [4] Effects of head tilt on visual field testing with a head-mounted perimeter imo
    Yamao, Sayaka
    Matsumoto, Chota
    Nomoto, Hiroki
    Numata, Takuya
    Eura, Mariko
    Yamashita, Marika
    Hashimoto, Shigeki
    Okuyama, Sachiko
    Kimura, Shinji
    Yamanaka, Kenzo
    Chiba, Yasutaka
    Aihara, Makoto
    Shimomura, Yoshikazu
    [J]. PLOS ONE, 2017, 12 (09):
  • [5] Validation of a Head-mounted Virtual Reality Visual Field Screening Device
    Mees, Lukas
    Upadhyaya, Swati
    Kumar, Pavan
    Kotawala, Sandal
    Haran, Shankar
    Rajasekar, Shruthi
    Friedman, David S.
    Venkatesh, Rengaraj
    [J]. JOURNAL OF GLAUCOMA, 2020, 29 (02) : 86 - 91
  • [6] The Effect of a Head-mounted Low Vision Device on Visual Function
    Wittich, Walter
    Lorenzini, Marie-Celine
    Markowitz, Samuel N.
    Tolentino, Michael
    Gartner, Scott A.
    Goldstein, Judith E.
    Dagnelie, Gislin
    [J]. OPTOMETRY AND VISION SCIENCE, 2018, 95 (09) : 774 - 784
  • [7] Visual Evoked Potential Using Head-Mounted Display Versus Cathode Ray Tube: A Pilot Study
    Choi, Hyo Seon
    Im, Sang Hee
    Kim, Yong Kyun
    Lee, Sang Chul
    [J]. ANNALS OF REHABILITATION MEDICINE-ARM, 2016, 40 (02): : 334 - 340
  • [8] Enhancing Presence in Head-mounted Display Environments by Visual Body Feedback Using Head-mounted Cameras
    Bruder, Gerd
    Steinicke, Frank
    Rothaus, Kai
    Hinrichs, Klaus
    [J]. 2009 INTERNATIONAL CONFERENCE ON CYBERWORLDS, 2009, : 43 - 50
  • [9] ENHANCING THE PERCEPTION OF A HAZY VISUAL WORLD USING A SEE-THROUGH HEAD-MOUNTED DEVICE
    Lin, Kai-En
    Shih, Kuang-Tsu
    Chen, Homer
    [J]. 2017 24TH IEEE INTERNATIONAL CONFERENCE ON IMAGE PROCESSING (ICIP), 2017, : 4397 - 4401
  • [10] Light-field head-mounted displays reduce the visual effort: A user study
    Panzirsch, Michael
    Weber, Bernhard
    Bechtel, Nicolai
    Grabner, Nicole
    Lingenauber, Martin
    [J]. JOURNAL OF THE SOCIETY FOR INFORMATION DISPLAY, 2022, 30 (04) : 319 - 334