High Spatial-Resolution Skull Base Imaging With Photon-Counting Computed Tomography and Energy-Integrating Computed Tomography: A Comparative Phantom Study

被引:4
|
作者
Rajagopal, Jayasai R. [1 ,2 ]
Schwartz, Fides R. [3 ]
Solomon, Justin B. [1 ]
Enterline, David S. [3 ]
Samei, Ehsan [1 ,4 ]
机构
[1] Duke Univ Med Ctr, Dept Radiol, Med Phys Grad Program, Carl E Ravin Adv Imaging Labs, Durham, NC USA
[2] NIH, Clin Ctr, Radiol & Imaging Sci, Bethesda, MD USA
[3] Duke Univ Med Ctr, Duke Univ Hlth Syst, Dept Radiol, Durham, NC USA
[4] Carl E Ravin Adv Imaging Labs, Hock Plaza,Suite 302,2424 Erwin Rd, Durham, NC 27710 USA
基金
美国国家卫生研究院;
关键词
computed tomography; high-resolution; photon-counting; skull base; CT; SYSTEM; RECONSTRUCTION; HEIGHT; NOISE;
D O I
10.1097/RCT.0000000000001464
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Photon-counting computed tomography (PCCT) offers better high-resolution and noise performance than energy integrating detector (EID) CT. In this work, we compared both technologies for imaging of the temporal bone and skull base. A clinical PCCT system and 3 clinical EID CT scanners were used to image the American College of Radiology image quality phantom using a clinical imaging protocol with matched CTDIvol (CT dose index-volume) of 25 mGy. Images were used to characterize the image quality of each system across a series of high-resolution reconstruction options. Noise was calculated from the noise power spectrum, whereas resolution was quantified with a bone insert by calculating a task transfer function. Images of an anthropomorphic skull phantom and 2 patient cases were examined for visualization of small anatomical structures. Across measured conditions, PCCT had a comparable or smaller average noise magnitude (120 Hounsfield units [HU]) to the EID systems (144-326 HU). Photon-counting CT also had comparable resolution (task transfer function f(25): 1.60 mm(-1)) to the EID systems (1.34-1.77 mm(-1)). Imaging results supported quantitative findings as PCCT more clearly showed the 12-lp/cm bars from the fourth section of the American College of Radiology phantom and better represented the vestibular aqueduct and oval and round windows when compared with the EID scanners. A clinical PCCT system was able to image the temporal bone and skull base with improved spatial resolution and lower noise than clinical EID CT systems at matched dose.
引用
收藏
页码:613 / 620
页数:8
相关论文
共 50 条
  • [21] Photon-Counting Computed Tomography: A Potentially Groundbreaking Imaging Technology With Challenges to be Faced
    Leguizamo-Isaza, Juan
    Perilla, Laura
    Valderrama, Juan
    Valderrama, Bibiana
    CANADIAN ASSOCIATION OF RADIOLOGISTS JOURNAL-JOURNAL DE L ASSOCIATION CANADIENNE DES RADIOLOGISTES, 2024, 75 (02): : 431 - 431
  • [22] Photon counting computed tomography of in-stent-stenosis in a phantom: Optimal virtual monoenergetic imaging in ultra high resolution
    Michael, Arwed Elias
    Schoenbeck, Denise
    Woeltjen, Matthias Michael
    Boriesosdick, Jan
    Niehoff, Julius Henning
    Surov, Alexey
    Borggrefe, Jan
    Schmidt, Bernhard
    Panknin, Christoph
    Hickethier, Tilman
    Maintz, David
    Bunck, Alexander Christian
    Gertz, Roman Johannes
    Kroeger, Jan Robert
    HELIYON, 2024, 10 (06)
  • [23] A High-Rate Energy-Resolving Photon-Counting ASIC for Spectral Computed Tomography
    Gustavsson, Mikael
    Ul Amin, Farooq
    Bjorklid, Anders
    Ehliar, Andreas
    Xu, Cheng
    Svensson, Christer
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2012, 59 (01) : 30 - 39
  • [24] Energy-Sensitive, Photon-Counting Computed Tomography: Opportunities and Technological Challenges
    Roessl, E.
    Brendel, B.
    Martens, G.
    Proksa, R.
    Schmidt, F.
    Thran, A.
    Schlomka, J.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [25] Optimization of beam quality for photon-counting spectral computed tomography in head imaging: simulation study
    Chen, Han
    Xu, Cheng
    Persson, Mats
    Danielsson, Mats
    JOURNAL OF MEDICAL IMAGING, 2015, 2 (04)
  • [26] A simulation study of high-resolution x-ray computed tomography imaging using irregular sampling with a photon-counting detector
    Lee, Seungwan
    Choi, Yu-Na
    Kim, Hee-Joung
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 726 : 175 - 180
  • [27] Utility of photon-counting detectors for MV-kV dual-energy computed tomography imaging
    Jadick, Giavanna
    Ventura, Maya
    La Riviere, Patrick J.
    JOURNAL OF MEDICAL IMAGING, 2024, 11
  • [28] Ultra-high-resolution photon-counting detector computed tomography of the lungs: Phantom and clinical assessment of radiation dose and image quality
    Chamberlin, Jordan H.
    Smith, Carter D.
    Maisuria, Dhruw
    Parrish, Joe
    van Swol, Elizabeth
    Mah, Eugene
    Emrich, Tilman
    Schoepf, U. Joseph
    Varga-Szemes, Akos
    O'Doherty, Jim
    Munden, Reginald F.
    Tipnis, Sameer, V
    Baruah, Dhiraj
    Kabakus, Ismail M.
    CLINICAL IMAGING, 2023, 104
  • [29] Optimizing Quantum Iterative Reconstruction for Ultra-high-resolution Photon-counting Computed Tomography of the Lung
    Toth, Adrienn
    Chamberlin, Jordan H.
    Puthoff, Gregory
    Baruah, Dhiraj
    O'Doherty, Jim
    Maisuria, Dhruw
    Mcguire, Aaron M.
    Schoepf, U. Joseph
    Munden, Reginald F.
    Kabakus, Ismail M.
    JOURNAL OF THORACIC IMAGING, 2025, 40 (02)
  • [30] Timing resolution in double-sided silicon photon-counting computed tomography detectors
    Sundberg, Christel
    Persson, Mats
    Wikner, J. Jacob
    Danielsson, Mats
    JOURNAL OF MEDICAL IMAGING, 2023, 10 (02)