Interferometric thermometry of ocular tissues for retinal laser therapy

被引:2
|
作者
Veysset, David [1 ,2 ]
Zhuo, Yueming [1 ,3 ]
Hattori, Junya [1 ,4 ]
Buckhory, Mohajeet [1 ,2 ]
Palanker, Daniel [1 ,2 ]
机构
[1] Stanford Univ, Hansen Expt Phys Lab, Stanford, CA 94305 USA
[2] Stanford Univ, Dept Ophthalmol, Stanford, CA 94305 USA
[3] Stanford Univ, Dept Elect Engn, Stanford, CA 94305 USA
[4] Univ Tokyo, Dept Mech Engn, Tokyo, Japan
基金
美国国家卫生研究院;
关键词
MICRO-DISPLACEMENT MEASUREMENT; OPTICAL COHERENCE TOMOGRAPHY; TRANSFER-MATRIX METHOD; PHOTOTHERMAL THERAPY; PHOTOCOAGULATION; CONSOLIDATION; TEMPERATURE; TRANSMISSION; COAGULATION;
D O I
10.1364/BOE.475705
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Controlling the tissue temperature rise during retinal laser therapy is highly desirable for predictable and reproducible outcomes of the procedure, especially with non-damaging settings. In this work, we demonstrate a method for determining the optical absorption, the thermal conductivity, and the thermal expansion coefficients of RPE and choroid using phase-resolved optical coherence tomography (pOCT). These parameters are extracted from the measured changes in the optical path length (Delta OPL) using an axisymmetric thermo-mechanical model. This allows the calculation of the temperature rise during hyperthermia, which was further validated by imaging the temperature-sensitive fluorescence at the same location. We demonstrate that, with a temperature uncertainty of +/- 0.9 degrees C and a peak heating of about 17 degrees C following a laser pulse of 20 ms, this methodology is expected to be safe and sufficiently precise for calibration of the non-damaging retinal laser therapy. The method is directly translatable to in-vivo studies, where we expect a similar precision. (c) 2022 Optica Publishing Group under the terms of the Optica Open Access Publishing Agreement
引用
收藏
页码:37 / 53
页数:17
相关论文
共 50 条
  • [1] Laser interferometric thermometry of semiconductors and dielectrics
    Magunov, A.N.
    Prikladnaya Matematika i Mekhanika, 1998, (02): : 6 - 18
  • [2] Laser interferometric thermometry of semiconductors and dielectrics
    Magunov, AN
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1998, 41 (03) : 297 - 308
  • [3] Interferometric imaging of thermal expansion for temperature control in retinal laser therapy
    Veysset, David
    Ling, Tong
    Zhuo, Yueming
    Pandiyan, Vimal Prabhu
    Sabesan, Ramkumar
    Palanker, Daniel
    BIOMEDICAL OPTICS EXPRESS, 2022, 13 (02) : 728 - 743
  • [4] LASER INTERFEROMETRIC THERMOMETRY FOR SUBSTRATE-TEMPERATURE MEASUREMENT
    SAENGER, KL
    GUPTA, J
    APPLIED OPTICS, 1991, 30 (10): : 1221 - 1226
  • [5] Retinal absorption measurements for laser therapy through interferometric imaging of the thermal expansion
    Veysset, David
    Zhuo, Yueming
    Hattori, Junya
    Bhuckory, Mohajeet Balveer
    Pandiyan, Vimal Prabhu
    Sabesan, Ramkumar
    Palanker, Daniel V.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2022, 63 (07)
  • [6] Laser interferometric thermometry of Si monocrystals with two rough surfaces
    Gasilov, A.Yu.
    Magunov, A.N.
    Geologiya Rudnykh Mestorozhdenij, 38 (06): : 32 - 35
  • [7] LASER DAMAGE TO OCULAR TISSUES
    MARSHALL, J
    PROCEEDINGS OF THE ROYAL SOCIETY OF MEDICINE-LONDON, 1973, 66 (09): : 842 - 844
  • [8] LASER COAGULATION OF OCULAR TISSUES
    FLOCKS, M
    ZWENG, HC
    ARCHIVES OF OPHTHALMOLOGY, 1964, 72 (05) : 604 - +
  • [9] LASER DAMAGE TO OCULAR TISSUES
    HONEY, RC
    AMERICAN INDUSTRIAL HYGIENE ASSOCIATION JOURNAL, 1968, 29 (02): : 91 - &
  • [10] Novel interferometric thermometry
    Huazhong Univ of Science and, Technology, Wuhan, China
    Guangdian Gongcheng, 1 (46-50):