Accuracy of determination of free light chains (Kappa and Lambda) in plasma and serum by Swedish laboratories as monitored by external quality assessment

被引:4
|
作者
Rollborn, Niclas [1 ]
Jakobsson, Jenny [1 ]
Campbell, Andrew [2 ]
Nordin, Gunnar [3 ]
Karlsson, Mathias [1 ,3 ]
Larsson, Anders [1 ]
Kultima, Kim [1 ]
机构
[1] Uppsala Univ, Dept Med Sci, Clin Chem, Uppsala, Sweden
[2] NABAS, As, Norway
[3] Equalis AB, Uppsala, Sweden
关键词
External quality assessment scheme; Free light chains; Kappa; Lambda; MONOCLONAL GAMMOPATHIES; NORMAL INDIVIDUALS; DIAGNOSIS; ASSAYS; IMMUNOASSAY; PROTEINURIA; MYELOMA; DIMERS;
D O I
10.1016/j.clinbiochem.2022.10.003
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Free light chain (FLC) measurements are important in diagnosing monoclonal gammopathies. As FLC are heterogeneous, different reagents and instruments for measuring FLC concentrations may give diverging results that affect assessment of patients with monoclonal gammopathies. Here we investigated agreement be-tween different FLC methods using data from the Swedish external quality assurance (EQA) programme.Methods: The two main FLC assays, N Latex FLC (Siemens) and Serum Freelite (The Binding Site), using four nephelometric or turbidimetric instrument platforms, were compared. Results from 27 EQA rounds distributed to 11-16 Swedish hospital laboratories during 2015-2020 were investigated.Results: The kappa (Kappa) FLC measurements deviated significantly over time, but when only nephelometry was used, deviation from the mean was lower (median ranges:-5% to 13 %). The CV was significantly higher for the Freelite assay (mean CV = 8.7) than for the N latex assay (mean CV = 5.7) (p < 0.0001). The coefficient of determination between all combinations of reagents and instrument platforms used was generally good (r2 = 0.76-0.87), and the correlation slope acceptable (0.81-1.2). For lambda (lambda) FLC measurements, no concordance between combinations of instruments and reagents is apparent, deviating between-40 % to + 48 % from the mean. The CV was significantly higher for the combination with nephelometry and the Freelite assay (CV mean = 13.9 %) than nephelometry and the N latex assay (CV mean = 9.9 %) (p <0.001). The coefficient of deter-mination varied between combinations of reagents and instrument platforms (r2 = 0.59-0.89) and the slope ranged between 0.48 and 1.5. Significant differences between the two reagents used were sometimes noted.Conclusions: Imprecision in lambda FLC affects the Kappa FLC/lambda FLC ratio. This may be important in clinical assessment of patients, especially differentiating between monoclonal and polyclonal gammopathies.
引用
收藏
页码:47 / 53
页数:7
相关论文
共 27 条
  • [21] CAN A PERSISTENTLY POSITIVE SERUM OR URINARY IMMUNOFIXATION (SIFE/UIFE), WITH NORMALIZATION OF SERUM FREE LIGHT CHAINS (FLC) ASSAY AND KAPPA/LAMBDA RATIO AFTER TREATMENT, HAVE A PROGNOSTIC IMPACT IN SYSTEMIC LIGHT CHAIN AMYLOIDOSIS (AL)?
    De Novellis, D.
    Balsamo, F.
    Pugliese, N.
    Giordano, C.
    Sierchio, L.
    Vasco, A.
    Savoia, M.
    Cerciello, G.
    Pane, F.
    HAEMATOLOGICA, 2020, 105 : S106 - S106
  • [22] CLINICAL UTILITY OF THE SERUM FREE LIGHT CHAINS (FLCS) KAPPA AND LAMBDA ASSAY IN THE FOLLOW-UP OF PATIENTS WITH MULTIPLE MYELOMA (MM) UNDERGOING AUTOLOGOUS PERIPHERAL BLOOD STEM CELLS TRANSPLANTATION (PBSCT)
    Culurgioni, F.
    Musiu, C.
    Uda, S.
    Schirru, G.
    Piras, D.
    Derudas, D.
    Pilo, F.
    Cabras, M. G.
    HAEMATOLOGICA, 2017, 102 : 168 - 168
  • [23] A supervised analysis of gene-expression profiles of purified clonal plasma cells from patients with systemic light-chain amyloidosis (AL) who have high or low levels of serum free lambda light chains
    Comenzo, R. L.
    Zhou, P.
    Fleisher, M.
    Olshen, A. B.
    XIth International Symposium on Amyloidosis, 2008, : 237 - 239
  • [24] A supervised analysis of gene-expression profiles of purified clonal plasma cells from patients with systemic light-chain amyloidosis (AL) who have high or low levels of serum free lambda light chains
    Comenzo, R. L.
    Zhou, P.
    Fleisher, M.
    Olshen, A. B.
    AMYLOID-JOURNAL OF PROTEIN FOLDING DISORDERS, 2006, 13 : 23 - 24
  • [25] The Total Amount of Kappa Plus Lambda Serum Immunoglobulin Free Light Chains (sFLC κ plus λ) Is a Powerful Independent Predictor of Time to First Treatment In Chronic Lymphocytic Leukemia (CLL) and Allows Definition of a Novel Prognostic Scoring System: A Study of 449 Therapy-naive Patients
    Morabito, Fortunato
    De Filippi, Rosaria
    Laurenti, Luca
    Zirlik, Katja
    Recchia, Anna Grazia
    Gentile, Massimo
    Morelli, Emanuela
    Cavalcanti, Solluzzo
    Iaccarino, Giancarla
    Amoroso, Barbara
    Nert, Antonino
    Cutrona, Giovanna
    Ferrarini, Manlio
    Molica, Stefano
    Del Poeta, Giovanni
    Pinto, Antonio
    BLOOD, 2010, 116 (21) : 1011 - 1011
  • [26] Correlation of Image Analysis-Assisted Detection of Monoclonal Plasma Cells in Bone Marrow Tissue Sections Using In-Situ Hybridization (ISH) Versus Serum Free Kappa and Lambda Light Chain Ratio and Immunofixation Electrophoresis (IFE) Studies in Patients with Plasma Cell Myeloma (PCM) Following Bone Marrow Transplantation
    Jahnke, R. K. Walsh
    Tilzer, L.
    Woodroof, J.
    MODERN PATHOLOGY, 2014, 27 : 532A - 532A
  • [27] Correlation of Image Analysis-Assisted Detection of Monoclonal Plasma Cells in Bone Marrow Tissue Sections Using In-Situ Hybridization (ISH) Versus Serum Free Kappa and Lambda Light Chain Ratio and Immunofixation Electrophoresis (IFE) Studies in Patients with Plasma Cell Myeloma (PCM) Following Bone Marrow Transplantation
    Jahnke, R. K. Walsh
    Tilzer, L.
    Woodroof, J.
    LABORATORY INVESTIGATION, 2014, 94 : 532A - 532A