Application research on neutron-gamma discrimination based on BC501A liquid scintillator

被引:1
|
作者
Luo, J. [1 ]
Hou, S. [1 ]
机构
[1] Xijing Univ, Shaanxi Engn Res Ctr Controllable Neutron Source, Sch Sci, Xian 710123, Peoples R China
关键词
Neutron-gamma discrimination; Geant4; simulation; Liquid scintillator; Fast neutron multiplicity counter;
D O I
10.1515/kern-2021-0014
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Liquid organic scintillators are widely used in non-destructive analysis, which plays an important role in nuclear disarmament verification. This paper focused on studying the neutron-gamma discrimination technology in the fast neutron multiplicity measuring counter based on BC501A liquid scintillation detector. First, the charge comparison method, the zero-crossing time method and the rise time method were compared via the Geant4 and Matlab algorithm, and the result shows that charge comparison has the highest Figure of Merit. Then, a neutron-gamma discrimination system based on the six-probe fast neutron multiplicity counter was built and tested with a conclusion that the mean value of Figure of merit is 1.08, which verify the satisfactory neutron-gamma discriminating capability of the system. Finally, for the uranium samples, the mass are detected by fast neutron multiplicity counter, and the enrichment are measured by the characteristic gamma-ray signals using the system. The experimental results are in good agreement with the actual data.
引用
收藏
页码:437 / 444
页数:8
相关论文
共 50 条
  • [21] NEUTRON-GAMMA DISCRIMINATION USING ZERO-CROSSING METHOD FOR BC454 BORON LOADED PLASTIC SCINTILLATOR
    Wang Dong
    He Bin
    Zhang Quanhu
    PROCEEDINGS OF THE 18TH INTERNATIONAL CONFERENCE ON NUCLEAR ENGINEERING 2010, VOL 3, 2011, : 89 - +
  • [22] Application of a BC501A Liquid Scintillation Detector with a Gain Stabilization System on the EAST Tokamak
    彭星宇
    陈忠靖
    杜腾飞
    胡志猛
    葛理健
    陈金象
    李湘庆
    樊铁栓
    Plasma Science and Technology, 2016, 18 (01) : 23 - 29
  • [23] Neutron-gamma discrimination with UGAB scintillator using zero-crossing method
    Divani-Vais, N.
    Bayat, E.
    Firoozabadi, M. M.
    Ghal-Eh, N.
    RADIATION PROTECTION DOSIMETRY, 2013, 154 (03) : 381 - 384
  • [24] Evaluating four neutron-gamma discrimination methods with EJ-301 scintillator
    Van Chuan Phan
    Xuan Hai Nguyen
    Analog Integrated Circuits and Signal Processing, 2019, 98 : 75 - 84
  • [25] Evaluating four neutron-gamma discrimination methods with EJ-301 scintillator
    Phan Van Chuan
    Nguyen Xuan Hai
    ANALOG INTEGRATED CIRCUITS AND SIGNAL PROCESSING, 2019, 98 (01) : 75 - 84
  • [26] Unfolding the fast neutron spectra of a BC501A liquid scintillation detector using GRAVEL method
    CHEN YongHao
    CHEN XiMeng
    LEI JiaRong
    AN Li
    ZHANG XiaoDong
    SHAO JianXiong
    ZHENG Pu
    WANG XinHua
    Science China(Physics,Mechanics & Astronomy), 2014, (10) : 1885 - 1890
  • [27] Application of a BC501A Liquid Scintillation Detector with a Gain Stabilization System on the EAST Tokamak
    彭星宇
    陈忠靖
    杜腾飞
    胡志猛
    葛理健
    陈金象
    李湘庆
    樊铁栓
    Plasma Science and Technology, 2016, (01) : 23 - 29
  • [28] Application of a BC501A Liquid Scintillation Detector with a Gain Stabilization System on the EAST Tokamak
    Peng Xingyu
    Chen Zhongjing
    Du Tengfei
    Hu Zhimeng
    Ge Lijian
    Chen Jinxiang
    Li Xiangqing
    Fan Tieshuan
    PLASMA SCIENCE & TECHNOLOGY, 2016, 18 (01) : 23 - 29
  • [29] Unfolding the fast neutron spectra of a BC501A liquid scintillation detector using GRAVEL method
    YongHao Chen
    XiMeng Chen
    JiaRong Lei
    Li An
    XiaoDong Zhang
    JianXiong Shao
    Pu Zheng
    XinHua Wang
    Science China Physics, Mechanics & Astronomy, 2014, 57 : 1885 - 1890
  • [30] Neutron spectrometry in mixed fields: NE213/BC501A liquid scintillation spectrometers