Clinical application of a new method for determination of the erythrocyte sedimentation rate using the BC-720 automated hematology analyzer

被引:0
|
作者
Shen, Yimin [1 ]
Liu, Dongmei [1 ,2 ]
Wang, Yong [1 ]
Cao, Jun [1 ]
Zhang, Shuaishuai [1 ]
Wen, Hui [1 ]
Dong, Qiuqiu [1 ]
Zheng, Dong [1 ]
Qiu, Jun [1 ]
机构
[1] Soochow Univ, Dept Lab Med, Affiliated Hosp 1, Suzhou, Jiangsu, Peoples R China
[2] Soochow Univ, Dept Lab Med, Affiliated Hosp 1, Suzhou 215000, Jiangsu, Peoples R China
关键词
BC-720; erythrocyte sedimentation rate; International Council for Standardization in Haematology; performance verification; Westergren alternate method; SYSTEM; RECOMMENDATIONS; AGGREGATION; MANAGEMENT; VISCOSITY; MECHANISM; BLOOD;
D O I
10.1111/ijlh.14044
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: The erythrocyte sedimentation rate (ESR) is a nonspecific inflammatory indicator and is widely used in clinical diagnosis. Westergren is the gold standard method recommended by the International Committee for Standardization of Hematology (ICSH), but it is time-consuming and inconvenient and has biosafety risks. A new alternate method for ESR (Easy-W ESR) measurement was designed and integrated into the Mindray BC-720 series automated hematology analyzer to meet the clinical needs of hematology laboratories for efficiency, safety, and automation. In this study, the performance of the new ESR method was evaluated based on the ICSH recommendations on modified and alternate ESR methods.Methods: Methodological comparisons using the BC-720 analyzer, TEST 1, and the Westergren method were performed to assess repeatability, carryover, sample stability, reference range validation, factors influencing the ESR, and clinical applicability in rheumatology and orthopedics.Results: The correlation between the BC-720 analyzer and the Westergren method was good (Y = 2.082 + 0.9869X, r = 0.9657, P > 0.0001, n = 342), carryover was < 1%, the repeatability standard deviation was <= 1 mm/h, and the coefficient of variation (CV) was <= 5%. The reference range meets the manufacturer's claim. For rheumatology patients, the BC-720 analyzer showed a good correlation with the Westergren method (Y = 1.021X-1.941, r = 0.9467, n = 149). For orthopedic patients, the BC-720 analyzer also showed a good correlation with the Westergren method (Y = 1.037X + 0.981, r = 0.978, n = 97).Conclusion: This study verified the clinical and analytical performance of the new ESR method, indicating that the results are very similar to those obtained using the Westergren method.
引用
收藏
页码:449 / 459
页数:11
相关论文
共 20 条
  • [1] Performance evaluation of the BC-720 auto hematology analyzer and establishment of the reference intervals of erythrocyte sedimentation rate in healthy adults
    Shu, Guofang
    Ding, Rui
    Ding, Rong
    He, Zhi
    Shen, Yimin
    Liu, Dongmei
    Duan, Zhaohui
    ANNALS OF TRANSLATIONAL MEDICINE, 2022, 10 (17)
  • [2] EVALUATION OF THE CLINICAL SUITABILITY OF MIND RAY BC-720 IN MEASURING ERYTHROCYTE SEDIMENTATION RATE
    Shu, Guofang
    He, Zhi
    Chen, Mindjie
    Li, Molong
    Ding, Rong
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2023, 45 : 105 - 105
  • [3] EVALUATION OF THE AUTOMATED HEMATOLOGY ANALYZER C-MEK-04X FOR DETERMINATION OF THE ERYTHROCYTE SEDIMENTATION RATE
    Maki, Ryosuke
    Yamamoto, Takamasa
    Takemura, Hiroyuki
    Nagai, Yutaka
    Esaki, Tatsuhiro
    Higuchi, Makoto
    Horii, Takashi
    Tabe, Yoko
    Ohsaka, Akimichi
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2018, 40 : 87 - 87
  • [4] EVALUATION OF BC-780: AN HEMATOLOGY ANALYZER WITH ERYTHROCYTE SEDIMENTATION RATE ANALYSIS MODULE INTEGRATED
    Seghezzi, MIchela
    Mangili, Ivonne Francesca
    Vaca, Maria del Carmen Baigorria
    Giuliani, Giorgia
    Cesani, Alessia
    Dominoni, Paola
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2024, 46 : 94 - 94
  • [5] A NOVEL ALTERNATE METHOD FOR ERYTHROCYTE SEDIMENTATION RATE MEASUREMENT INTEGRATED INTO A HEMATOLOGY ANALYZER
    Qiu, Jun
    Liu, Dongmei
    Shen, Yimin
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2023, 45 : 142 - 142
  • [6] Evaluation of an automated erythrocyte sedimentation rate analyzer as compared to the Westergren manual method in measurement of erythrocyte sedimentation rate
    Subramanian, Arulselvi
    Rangarajan, Kanchana
    Pandey, Ravindra Mohan
    Gandhi, Jatin S.
    Sharma, Vijay
    Bhoi, Sanjeev Kumar
    INDIAN JOURNAL OF PATHOLOGY AND MICROBIOLOGY, 2011, 54 (01) : 70 - 74
  • [7] Accuracy study of a novel alternate method measuring erythrocyte sedimentation rate for prototype hematology analyzer Celltac α
    Maki, Ryosuke
    Tabe, Yoko
    Yamamoto, Takamasa
    Takemura, Hiroyuki
    Sawada, Tomohiro
    Esaki, Tatsuhiro
    Higuchi, Makoto
    Misawa, Shigeki
    Miyake, Kazunori
    Miida, Takashi
    Nagai, Yutaka
    Ohsaka, Akimichi
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2021, 43 (04) : 588 - 596
  • [8] PERFORMANCE EVALUATION OF A NOVEL METHOD FOR THE MEASUREMENT OF ERYTHROCYTE SEDIMENTATION RATE (ESR) ON THE NEW HEMATOLOGY ANALYSER MIND RAY BC-780
    Tejedor Ganduxa, Xavier
    Rufo-Vicente, Elena
    Martinez-Irribaren, Alicia
    Leis-Sestaio, Alba
    Morales-Indiano, Cristian
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2023, 45 : 115 - 115
  • [9] Analytical validation of the modified Westergren method on the automated erythrocyte sedimentation rate analyzer CUBE 30 touch
    Lapic, Ivana
    Rade, Anamarija
    Kraljevic, Andela
    Milos, Marija
    Herak, Desiree Coen
    Daskijevic, Lucija
    Cerovac, Paula
    Rogic, Dunja
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2023, 61 (08) : 1463 - 1469
  • [10] The TEST 1 automated system - A new method for measuring the erythrocyte sedimentation rate
    Plebani, M
    De Toni, S
    Sanzari, MC
    Bernardi, D
    Stockreiter, E
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1998, 110 (03) : 334 - 340