Pulsed Magnetic Field Generator for Enhancing Cell Membrane Permeability In Vitro

被引:3
|
作者
Ma, Chi [1 ]
Li, Zhengmin [1 ]
Zhu, Yakui [1 ]
Zhang, Mengnan [1 ]
Wang, Jianli [1 ]
Mi, Yan [1 ]
机构
[1] Chongqing Univ, Sch Elect Engn, State Key Lab Power Transmiss Equipment Technol, Chongqing 400044, Peoples R China
基金
中国国家自然科学基金;
关键词
Induced electric field; magnetic field gradient; pulsed current generator; pulsed magnetic field (PMF); ELECTROPORATION; ELECTROCHEMOTHERAPY; COMBINATION; DELIVERY;
D O I
10.1109/TPS.2023.3333335
中图分类号
O35 [流体力学]; O53 [等离子体物理学];
学科分类号
070204 ; 080103 ; 080704 ;
摘要
The noncontact pulsed magnetic field (PMF) treatment can enhance cell membrane permeability, but its efficiency is relatively low. To enhance its efficiency, this article has developed a more optimized PMF generator: this generator can produce a maximum magnetic induction intensity of 1.9 T; and a maximum-induced electric field of approximately 110 V/cm with a uniformity of up to 84%. Importantly, the generator developed in this article can also generate a higher spatial gradient of the magnetic field, approximately 1000 T/m, a capability that was previously lacking in research. Subsequently, by customizing the coil structure and the heat dissipation system, we ensured both temperature control of the treatment plane and a sufficiently large effective treatment volume. Finally, the effectiveness of the PMF generator was confirmed through in vitro cell experiments.
引用
收藏
页码:3647 / 3654
页数:8
相关论文
共 50 条
  • [21] A multi-module energy-saving pulsed magnetic field generator
    Liu, Jinzhen
    Wang, Yuling
    Chen, Yimei
    Xiong, Hui
    Sun, Wanpeng
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2019, 90 (01):
  • [22] A 7 T Pulsed Magnetic Field Generator for Magnetized Laser Plasma Experiments
    胡广月
    梁亦寒
    宋法伦
    袁鹏
    王雨林
    赵斌
    郑坚
    Plasma Science and Technology, 2015, (02) : 134 - 140
  • [23] A 7 T Pulsed Magnetic Field Generator for Magnetized Laser Plasma Experiments
    Hu Guangyue
    Liang Yihan
    Song Falun
    Yuan Peng
    Wang Yulin
    Zhao Bin
    Zheng Jian
    PLASMA SCIENCE & TECHNOLOGY, 2015, 17 (02) : 134 - 140
  • [24] A 7 T Pulsed Magnetic Field Generator for Magnetized Laser Plasma Experiments
    胡广月
    梁亦寒
    宋法伦
    袁鹏
    王雨林
    赵斌
    郑坚
    Plasma Science and Technology, 2015, 17 (02) : 134 - 140
  • [25] Can Lucifer Yellow Indicate Correct Permeability of Biological Cell Membrane under An Electric and Magnetic Field?
    Firoozabadi, Tahereh Pourmirjafari
    Shankayi, Zeinab
    Izadi, Azam
    Firoozabadi, Seyed Mohammad Pourmirjafari
    CELL JOURNAL, 2015, 16 (04) : 560 - 563
  • [26] An experimental study on the magnetic field permeability of an inductive-pulsed plasma thruster
    Che, Bi-Xuan
    Wu, Jian-Jun
    Cheng, Mou-Sen
    Li, Xiao-Kang
    Zhang, Yu
    Guo, Da-wei
    CHINESE PHYSICS B, 2025, 34 (02)
  • [27] An experimental study on the magnetic field permeability of an inductive-pulsed plasma thruster
    车碧轩
    吴建军
    程谋森
    李小康
    张宇
    郭大伟
    Chinese Physics B, 2025, 34 (02) : 345 - 354
  • [28] Effect of high intensity pulsed electric field on the membrane permeability and DNA of E coli
    Zhang, Ying
    Zeng, Xin'an
    Wen, Qibiao
    Feng, Jie
    Gaodianya Jishu/High Voltage Engineering, 2007, 33 (02): : 90 - 93
  • [29] Effect of Pulsed Electric Field on Electroporation Area of Cell Membrane
    Wu X.
    Mi Y.
    Zheng W.
    Ma C.
    Zhang M.
    Diangong Jishu Xuebao/Transactions of China Electrotechnical Society, 2023, 38 (14): : 3779 - 3788
  • [30] Enhancing cell membrane permeability of peptide nucleic acids and intracellular inhibition of telomerase
    Simmons, CG
    Mayfield, LD
    Shay, JW
    Corey, DR
    FASEB JOURNAL, 1997, 11 (09): : A1154 - A1154