摘要
Cantilever vibration in an optical lever AFM is theoretically analyzed considering the effect of surface force and ambient fluid. The surface force between the tip and the sample is modeled using a cantilever with a spring and damping added to the tip, and the effect of ambient fluid on the cantilever vibration is analyzed using the bead model. The amplitude ratios of the forced vibration were derived by solving the equation of motion. There are discrepancies between the amplitude ratio derived from bending vibration theory and that obtained through the optical lever method. The amplitude ratio from bending vibration theory reveals that the effect of surface force dominates primarily below the second resonant frequency. The tip-added spring affects the resonant frequency shift and amplitude ratio, while the tip-added damping mainly affects the amplitude ratio at the resonant frequency. In contrast, the amplitude ratio considering the optical lever method reveals that anti-resonance occurs when the tip-added spring constant is positive, which is not observed in bending vibration theory. The amplitude ratio of the cantilever is modified by introducing a correction factor. The correction factor decreases with excitation frequency except for the frequency where the anti-resonance occurs. The correction factor becomes small when the ambient fluid has high density and viscosity.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] On AFM Measurements of the Interaction Force Vector by Means of Interferometry, Optical Lever, and the Piezoresistive Method
    Ankudinov, A., V
    JOURNAL OF SURFACE INVESTIGATION, 2022, 16 (02): : 247 - 253
  • [2] On AFM Measurements of the Interaction Force Vector by Means of Interferometry, Optical Lever, and the Piezoresistive Method
    A. V. Ankudinov
    Journal of Surface Investigation: X-ray, Synchrotron and Neutron Techniques, 2022, 16 : 247 - 253
  • [3] SELF-CORRECTION METHOD FOR DYNAMIC MEASUREMENT ERROR OF FORCE SENSORS
    Fujii, Y.
    Maru, K.
    EXPERIMENTAL TECHNIQUES, 2011, 35 (03) : 15 - 20
  • [4] Self-correction method for dynamic measurement error of force sensors
    Y. Fujii
    K. Maru
    Experimental Techniques, 2011, 35 : 15 - 20
  • [5] Measurement of Chladni Mode Shapes with an Optical Lever Method
    Feng, Rong
    Luo, Yan
    Dong, Yixue
    Ma, Mengke
    Wang, Yuqi
    Zhang, Jing
    Ma, Wenjiang
    Liu, Donghuan
    JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2020, (160): : 1 - 9
  • [6] An Improved Method of Adhesion Force Measurement by Atomic Force Microscopy (AFM)
    Jin, Hong
    Li, Xiongyao
    Wen, Yuanyun
    Tang, Hong
    Zeng, Xiaojia
    ATOMIC SPECTROSCOPY, 2022, 43 (01) : 13 - 18
  • [7] Optical lever detection in higher eigenmode dynamic atomic force microscopy
    Stark, RW
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (11): : 5053 - 5055
  • [8] Direct Measurement of Optical Force Induced by Near-Field Plasmonic Cavity Using Dynamic Mode AFM
    Guan, Dongshi
    Hang, Zhi Hong
    Marcet, Zsolt
    Liu, Hui
    Kravchenko, I. I.
    Chan, C. T.
    Chan, H. B.
    Tong, Penger
    SCIENTIFIC REPORTS, 2015, 5
  • [9] Direct Measurement of Optical Force Induced by Near-Field Plasmonic Cavity Using Dynamic Mode AFM
    Dongshi Guan
    Zhi Hong Hang
    Zsolt Marcet
    Hui Liu
    I. I. Kravchenko
    C. T. Chan
    H. B. Chan
    Penger Tong
    Scientific Reports, 5
  • [10] Optical method for accurate force measurement: Dynamic response evaluation of an impact hammer
    Fujii, Y
    OPTOMECHATRONIC SENSORS, ACTUATORS, AND CONTROL, 2004, 5602 : 83 - 90