Estimate of Errors in Measurements of Refractive Index by Modified Prism Methods

被引:0
|
作者
Yurin, A. I. [1 ,2 ]
Vishnyakov, G. N. [2 ,3 ]
Minaev, V. L. [1 ,2 ]
机构
[1] HSE Univ, Moscow, Russia
[2] All Russian Res Inst Opt & Phys Measurements, Moscow, Russia
[3] Bauman Moscow State Tech Univ, Moscow, Russia
关键词
refractive index; goniometer; prism method; measurement error; ACCURACY;
D O I
10.1007/s11018-023-02206-9
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Goniometric methods of measuring the refractive index of optically transparent materials based on the refraction of light by a triangular prism are studied. A modified minimum deviation method and 3 modified constant deviation methods are examined which make it possible to determine the refractive indices of triangular prisms with unknown refracting angles. According to the modified prism methods the deflection angles of the light by the prism are measured with a goniometer, while the refractive index of the material and the refraction angles of the prism are determined by solving systems of equations. Thus, there is no need for preliminary measurement of the prism angles, which would require special autocollimation goniometers. In addition, in the modified prism methods, light reflected from the faces of the prism is not used, which makes it possible to extend the spectral range of a measurement of the refractive index to the infrared and ultraviolet ranges. The errors in measurements of the refractive index by these methods are compared for the example of a prism with a refractive index of 1.5 and a refraction angle of 60 degrees. It is shown that the modified minimum deviation method has the smallest error among all the prism methods, so it can be recommended for high-precision measurements of the refractive index in those cases where the refractive angles of the prism are unknown or it is technically difficult to measure them. The modified methods examined here can be used for measuring the refractive index of triangular prisms made of optically transparent materials, as well as of liquids poured into hollow prisms with plane-parallel transparent windows. Practical implementation of methods of this type should be useful in the optical, chemical, and food industries for monitoring the composition and properties of optically transparent materials.
引用
收藏
页码:168 / 172
页数:5
相关论文
共 50 条
  • [21] Measurement of refractive index of powder by prism coupler
    Chen, Feng
    Xu, Qing
    Zhang, Pei
    Zhang, Zi-Bo
    Wong, Wing-Han
    Pun, Edwin Yue-Bun
    Zhang, De-Long
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2019, 90 (09):
  • [22] REFRACTIVE-INDEX MEASUREMENTS FOR THE CORRECTION OF PARTICLE SIZING METHODS
    ROTH, N
    ANDERS, K
    FROHN, A
    APPLIED OPTICS, 1991, 30 (33): : 4960 - 4965
  • [23] ON ERRORS OF REFLECTIVITY MEASUREMENTS + OF CALCULATIONS OF REFRACTIVE INDEX N + ABSORPTION COEFFICIENT K
    PILLER, H
    VONGEHLEN, K
    AMERICAN MINERALOGIST, 1964, 49 (7-8) : 867 - &
  • [24] DETERMINING THE REFRACTIVE-INDEX AND THICKNESS OF THIN-FILMS FROM PRISM COUPLER MEASUREMENTS
    KIRSCH, ST
    APPLIED OPTICS, 1981, 20 (12): : 2085 - 2089
  • [25] ERRORS OF REFRACTOMETRIC METHODS OF DETERMINING THE REFRACTIVE-INDEX PROFILE OF OPTICAL FIBERS
    MOROZOV, VN
    MOLOCHNIKOV, BI
    GORLOVA, ZV
    SOVIET JOURNAL OF OPTICAL TECHNOLOGY, 1990, 57 (06): : 350 - 352
  • [26] SAVART INTERFEROMETER MODIFIED FOR HIGH SENSITIVITY IN MEASUREMENTS OF REFRACTIVE INDEX GRADIENTS
    WALLIN, LE
    WALLIN, K
    OPTICA ACTA, 1970, 17 (05): : 381 - &
  • [28] HOLLOW PRISM FOR GONIOMETRIC CALIBRATION OF REFRACTIVE INDEX MEDIA
    COGSWELL, RF
    AMERICAN MINERALOGIST, 1945, 30 (7-8) : 541 - 543
  • [29] COMPARISON OF REFRACTIVE-INDEX MEASUREMENTS OF OPTICAL FIBERS BY 3 METHODS
    RAINE, KW
    BAINES, JG
    KING, RJ
    IEE PROCEEDINGS-J OPTOELECTRONICS, 1988, 135 (03): : 190 - 195
  • [30] ELIMINATION OF ERRORS DUE TO SAMPLE POLISHING IN REFRACTIVE-INDEX PROFILE MEASUREMENTS BY INTERFEROMETRY
    STONE, J