Influence of the Tube Angle on the Measurement Accuracy of Peri-Implant Bone Defects in Rectangular Intraoral X-ray Imaging

被引:0
|
作者
Rugani, Petra [1 ]
Weingartner, Katharina [1 ]
Jakse, Norbert [1 ]
机构
[1] Med Univ Graz, Dept Dent Med & Oral Hlth, Div Oral Surg & Orthodont, Billrothgasse 4, A-8010 Graz, Austria
关键词
peri-implant bone level; intra-oral radiography; implantology; RADIOGRAPHIC MEASUREMENT; PARALLELING TECHNIQUE; CONSENSUS REPORT; WORKSHOP; DISEASES; LEVEL; REPRODUCIBILITY; CLASSIFICATION; IMPACT; ERRORS;
D O I
10.3390/jcm13020391
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Intraoral radiography in the right-angle technique is the standard procedure to examine the peri-implant bone level in implant follow-up and implant-related studies. For the implementation of the right-angle or parallel technique, mostly ready-made image receptor holders are used. The aim of this experimental study is to analyze changes in the measurement of standardized peri-implant defects caused by a deviation in the position of the image receptor. Methods: Eleven Xive (R) implants (Dentsply Sirona, Bensheim, Germany) were placed in bovine bone, and peri-implant defects of varying depths were created. The preparations were fixed in a specially made test stand, and intraoral radiographs were taken using the right-angle technique with standard film holders at various horizontal and vertical projection angles. Defect measurement was carried out with the imaging software Sidexis 4 V 4.3 (Dentsply Sirona, Bensheim, Germany). Results: With increasing angular deviation, larger deviations between the measured and the real extent of the defect occurred. Vertical tilting caused significant distortion, while horizontal rotation showed less effect. Conclusion: Intraoral radiography only provides a valid representation of the peri-implant bone level for follow-up or as a tool in implant-related studies if a reproducible projection direction is assured.
引用
收藏
页数:14
相关论文
共 50 条
  • [31] Influence of structure defects of microchannel-plate-based X-ray optics on imaging performance
    Lian, Jiao
    Zhou, Zhenyu
    Mei, Lin
    Wu, Xianmin
    Wang, Sanzhao
    Liu, Chang
    Wang, Chen
    Feng, Yuechong
    Zhang, Xian
    Liu, Hui
    AOPC 2020: TELESCOPES, SPACE OPTICS, AND INSTRUMENTATION, 2020, 11570
  • [32] Imaging the Nanostructure of Bone and Dentin Through Small- and Wide-Angle X-Ray Scattering
    Pabisch, Silvia
    Wagermaier, Wolfgang
    Zander, Thomas
    Li, Chenghao
    Fratzl, Peter
    RESEARCH METHODS IN BIOMINERALIZATION SCIENCE, 2013, 532 : 391 - 413
  • [33] X-RAY TUBE FOCAL SPOTS - IMAGING PROPERTIES, MEASUREMENT OF FOCAL SPOT SIZES, STANDARDIZATION PROCEDURE
    ZIELER, E
    MULLER, CHF
    PHYSICS IN MEDICINE AND BIOLOGY, 1977, 22 (03): : 595 - 595
  • [34] Influence of measurement conditions on the accuracy of measured stress using X-ray 2D detector
    Dept. of Mech. Eng., Yokohama National Univ., Hodogaya-ku, Yokohama, 240-8603, Japan
    Zairyo, 7 (521-526):
  • [35] Measurement accuracy in X-ray computed tomography metrology: Toward a systematic analysis of interference effects in tomographic imaging
    Hiller, Jochen
    Hornberger, Peter
    PRECISION ENGINEERING-JOURNAL OF THE INTERNATIONAL SOCIETIES FOR PRECISION ENGINEERING AND NANOTECHNOLOGY, 2016, 45 : 18 - 32
  • [36] Techniques to Improve the Accuracy of Presampling MTF Measurement in Digital X-ray Imaging Based on Constrained Spline Regression
    Zhou, Zhongxing
    Zhu, Qingzhen
    Zhao, Huijuan
    Zhang, Lixin
    Ma, Wenjuan
    Gao, Feng
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2014, 61 (04) : 1339 - 1349
  • [37] High resolution X-ray imaging of bone-implant interface by large area flat-panel detector
    Kytyr, D.
    Jirousek, O.
    Dammer, J.
    JOURNAL OF INSTRUMENTATION, 2011, 6
  • [38] Influence of tube voltage and current on in-line phase contrast imaging using a microfocus x-ray source
    Department of Engineering Physics, Tsinghua University, Beijing 100084, China
    Chin. Phys., 2007, 8 (2319-2324):
  • [39] Influence of tube voltage and current on in-line phase contrast imaging using a microfocus x-ray source
    Li Chen
    Li Zheng
    Yu Ai-Min
    Li Cheng-Quan
    CHINESE PHYSICS, 2007, 16 (08): : 2319 - 2324
  • [40] Influence Mechanism of Texture on the Accuracy of X-ray Stress Measurement for 6061-T6 Aluminum Alloy
    Wang X.
    Li X.
    Wu Q.
    Xu Z.
    1600, Cailiao Daobaoshe/ Materials Review (34): : 20081 - 20085