Influence of the Tube Angle on the Measurement Accuracy of Peri-Implant Bone Defects in Rectangular Intraoral X-ray Imaging

被引:0
|
作者
Rugani, Petra [1 ]
Weingartner, Katharina [1 ]
Jakse, Norbert [1 ]
机构
[1] Med Univ Graz, Dept Dent Med & Oral Hlth, Div Oral Surg & Orthodont, Billrothgasse 4, A-8010 Graz, Austria
关键词
peri-implant bone level; intra-oral radiography; implantology; RADIOGRAPHIC MEASUREMENT; PARALLELING TECHNIQUE; CONSENSUS REPORT; WORKSHOP; DISEASES; LEVEL; REPRODUCIBILITY; CLASSIFICATION; IMPACT; ERRORS;
D O I
10.3390/jcm13020391
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Intraoral radiography in the right-angle technique is the standard procedure to examine the peri-implant bone level in implant follow-up and implant-related studies. For the implementation of the right-angle or parallel technique, mostly ready-made image receptor holders are used. The aim of this experimental study is to analyze changes in the measurement of standardized peri-implant defects caused by a deviation in the position of the image receptor. Methods: Eleven Xive (R) implants (Dentsply Sirona, Bensheim, Germany) were placed in bovine bone, and peri-implant defects of varying depths were created. The preparations were fixed in a specially made test stand, and intraoral radiographs were taken using the right-angle technique with standard film holders at various horizontal and vertical projection angles. Defect measurement was carried out with the imaging software Sidexis 4 V 4.3 (Dentsply Sirona, Bensheim, Germany). Results: With increasing angular deviation, larger deviations between the measured and the real extent of the defect occurred. Vertical tilting caused significant distortion, while horizontal rotation showed less effect. Conclusion: Intraoral radiography only provides a valid representation of the peri-implant bone level for follow-up or as a tool in implant-related studies if a reproducible projection direction is assured.
引用
收藏
页数:14
相关论文
共 50 条
  • [1] Diagnostic accuracy of CBCT versus intraoral imaging for assessment of peri-implant bone defects
    Song, Dandan
    Shujaat, Sohaib
    Vasconcelos, Karla de Faria
    Huang, Yan
    Politis, Constantinus
    Lambrichts, Ivo
    Jacobs, Reinhilde
    BMC MEDICAL IMAGING, 2021, 21 (01)
  • [2] Accuracy of intraoral digital radiography in detecting peri-implant bone defects.
    Matsuda, Y
    Hanazawa, T
    Seki, K
    Sano, T
    Okano, T
    JOURNAL OF DENTAL RESEARCH, 1997, 76 : 2954 - 2954
  • [3] Diagnostic accuracy of CBCT versus intraoral imaging for assessment of peri‐implant bone defects
    Dandan Song
    Sohaib Shujaat
    Karla de Faria Vasconcelos
    Yan Huang
    Constantinus Politis
    Ivo Lambrichts
    Reinhilde Jacobs
    BMC Medical Imaging, 21
  • [4] Assessment of peripheral dual-energy X-ray absorptiometry measurements in peri-implant bone defects in dogs
    Sánchez, AR
    Sheridan, PJ
    Lohse, C
    Weaver, A
    JOURNAL OF PERIODONTOLOGY, 2004, 75 (05) : 658 - 662
  • [5] Comparative analysis of imaging techniques for diagnostic accuracy of peri-implant bone defects: a meta-analysis
    Lima Bohner, Lauren Oliveira
    Mukai, Eduardo
    Oderich, Elisa
    Porporatti, Andre Luis
    Pacheco-Pereira, Camila
    Tortamano, Pedro
    Canto, Graziela De Luca
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY, 2017, 124 (04): : 432 - +
  • [6] Intraoral radiography lacks accuracy for the assessment of peri-implant bone level - a controlled clinical study
    Christiaens, Veronique
    Jacobs, Reinhilde
    Dierens, Melissa
    Vervaeke, Stijn
    De Bruyn, Hugo
    Koole, Sebastiaan
    Cosyn, Jan
    EUROPEAN JOURNAL OF ORAL IMPLANTOLOGY, 2017, 10 (04) : 435 - 441
  • [7] Quantitative measurement of peri-implant bone defects using optical coherence tomography
    Kim, Sulhee
    Kang, Se-Ryong
    Park, Hee-Jung
    Kim, Bome
    Kim, Tae-Il
    Yi, Won-Jin
    JOURNAL OF PERIODONTAL AND IMPLANT SCIENCE, 2018, 48 (02): : 84 - 91
  • [8] ACCURACY OF HIGH-FREQUENCY ULTRASOUND SCANNER IN DETECTING PERI-IMPLANT BONE DEFECTS
    Bohner, Lauren
    Habor, Daniel
    Gremse, Felix
    Tortamano, Pedro
    Wolfart, Stefan
    Marotti, Juliana
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2019, 45 (03): : 650 - 659
  • [9] Rotational panoramic versus Intraoral rectangular radiographs for evaluation of peri-implant bone loss in the anterior atrophic mandible
    Zechner, W
    Watzak, G
    Gahleitner, A
    Busenlechner, D
    Tepper, G
    Watzek, G
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2003, 18 (06) : 873 - 878
  • [10] Dual X-ray absorptiometry for alveolar bone: Precision of peri-implant mineral measurements ex vivo
    Denissen, H
    Verhey, H
    deBlieck, J
    Corten, F
    Klein, C
    vanLingen, A