Portable Single-Exposure Dual-Energy X-ray Detector for Improved Point-of-Care Diagnostic Imaging

被引:3
|
作者
Karim, Karim S. [1 ,2 ]
Ii, Steven Tilley [2 ]
机构
[1] Univ Waterloo, Elect & Comp Engn, Waterloo, ON N2L 3G1, Canada
[2] KA Imaging, Waterloo, ON N2L 5Z4, Canada
关键词
DIGITAL RADIOGRAPHY; CHEST RADIOGRAPHY; SUBTRACTION;
D O I
10.1093/milmed/usad034
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Dual-energy subtraction (DES) imaging is well known to reduce anatomical noise and enable material classification. The current approaches to DES imaging have trade-offs, such as motion artifacts, low sensitivity because of losses in a mid-filter, and lack of portability. Recently, a portable triple-layer flat-panel detector (FPD) was proposed for use in single-shot DES imaging that can provide improved sensitivity and removal of motion artifacts in a point-of-care setting. The purpose of this study is to evaluate the feasibility of such a detector. Various image quality metrics and clinical images are provided. Materials and Methods: An FDA-cleared single-exposure DES FPD consisting of three stacked sensors was used for all measurements. This detector generates three images per exposure: A digital radiography (DR) image, i.e., as would be produced with a conventional detector, and two DES images, bone and soft tissue. To evaluate DR image quality, detective quantum efficiency (DQE) and modulation transfer function were measured for multiple radiation quality beams. Digital radiography and DES images obtained from this FPD were evaluated in previously reported fixed and portable clinical studies. Digital radiography and DES images from case studies are presented for qualitative assessment. Results: Modulation transfer function and DQE were measured across multiple radiation quality beams for the DR image. The DES images showed good tissue separation and uniformity with no visible motion or alignment artifacts. The DES images, when read in conjunction with the DR image, resulted in increased reader confidence and revealed abnormalities or details that were sometimes overlooked in the DR image. Conclusions: The proposed panel produces high-quality DR images as indicated by the DQE and modulation transfer function. The DES images have been shown to improve sensitivity in clinical applications and increase reader confidence. This detector can enable DES in portable or otherwise difficult applications, opening new doors for improved patient care.
引用
收藏
页码:84 / 91
页数:8
相关论文
共 50 条
  • [31] Quantitative single-exposure x-ray phase contrast imaging using a single attenuation grid
    Morgan, Kaye S.
    Paganin, David M.
    Siu, Karen K. W.
    OPTICS EXPRESS, 2011, 19 (20): : 19781 - 19789
  • [32] Development of a deep neural network for generating synthetic dual-energy chest x-ray images with single x-ray exposure
    Lee, Donghoon
    Kim, Hwiyoung
    Choi, Byungwook
    Kim, Hee-Joung
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (11):
  • [33] Material discrimination by high-energy x-ray dual-energy imaging
    Wang Xue-Wu
    Li Jian-Min
    Kang Ke-Jun
    Tang Chuan-Xiang
    Mang Li
    Chen Zhi-Qiang
    Li Yuan-Jing
    Zhong Hua-Qiang
    HIGH ENERGY PHYSICS AND NUCLEAR PHYSICS-CHINESE EDITION, 2007, 31 (11): : 1076 - 1081
  • [34] Investigation of dual-energy X-ray photon counting using a cadmium telluride detector with dual-energy selection electronics
    Sato, Eiichi
    Kosuge, Yoshiyuki
    Yamanome, Hayato
    Mikata, Aldko
    Miura, Tatsuya
    Oda, Yasuyuki
    Ishii, Tomotaka
    Hagiwara, Osahiko
    Matsukiyo, Hiroshi
    Watanabe, Manabu
    Kusachi, Shinya
    RADIATION PHYSICS AND CHEMISTRY, 2017, 130 : 385 - 390
  • [35] Robust Material Classification on Dual-Energy X-Ray Imaging Devices
    Cigla, Cevahir
    Kucukates, Busra
    Yalcin, Ozan
    Ak, Duygu Selin
    Isikoglu, Sukrucan Taylan
    ANOMALY DETECTION AND IMAGING WITH X-RAYS (ADIX) VII, 2022, 12104
  • [36] Evaluation of a dichromatic X-ray source for dual-energy imaging in mammography
    Tuffanelli, A
    Fabbri, S
    Sarnelli, A
    Taibi, A
    Gambaccini, M
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 489 (1-3): : 509 - 518
  • [37] Theoretical Feasibility of Dual-Energy X-Ray Imaging of Respiratory Disease
    Basharat, F.
    Tanguay, J.
    MEDICAL PHYSICS, 2020, 47 (06) : E558 - E558
  • [38] Stereoscopic dual-energy X-ray imaging for target materials identification
    Wang, TW
    Evans, JPO
    IEE PROCEEDINGS-VISION IMAGE AND SIGNAL PROCESSING, 2003, 150 (02): : 122 - 130
  • [39] Image segmentation for binocular stereoscopic dual-energy x-ray imaging
    Yong, YS
    Evans, JPO
    Chan, JW
    JOURNAL OF ELECTRONIC IMAGING, 2005, 14 (04)
  • [40] Dual-energy medical X-ray imaging using LaPlace transformation
    Tang, SJ
    Mou, XQ
    Yang, Y
    Luo, T
    Cai, YL
    METMBS '05: PROCEEDINGS OF THE 2005 INTERNATIONAL CONFERENCE ON MATHEMATICS AND ENGINEERING TECHNIQUES IN MEDICINE AND BIOLOGICAL SCIENCES, 2005, : 117 - 123