Biomechanical Performance of Biodegradable Screw Fixation in Mandibular Distraction Osteogenesis

被引:0
|
作者
Firriolo, Joseph Mark [1 ,4 ]
Xue, Yunfeng [1 ]
Weyerbacher, Jonathan S. [1 ]
Saldana, Golddy M. [1 ]
Garfinkel, Alec [2 ]
Garcia, Tanya C. [3 ]
Arzi, Boaz [3 ]
Wong, Granger B. [1 ]
机构
[1] Univ Calif Davis, Div Plast & Reconstruct Surg, Sacramento, CA USA
[2] Calif Northstate Univ, Coll Med, Elk Grove, CA USA
[3] Univ Calif Davis, Sch Vet Med, Dept Surg & Radiol Sci, Davis, CA USA
[4] Univ Calif Davis, Div Plast & Reconstruct Surg, 2335 Stockton Blvd,6th Floor, Sacramento, CA 95817 USA
关键词
Biomechanics; mandible; micrognathism; osteogenesis; distraction; BONE;
D O I
10.1097/SCS.0000000000009691
中图分类号
R61 [外科手术学];
学科分类号
摘要
Internal distraction devices are commonly used in congenital micrognathia. The eventual need for device and screw removal can be challenging, requiring extensive dissection and disturbance of bone regenerate. Bioabsorbable poly-L-lactide (PLLA) screws, compared to traditional titanium screws, simplify device removal. Previous in vivo studies have found that the maximal compressive force generated by mandibular distraction is 69.4N. We hypothesized that PLLA screws could support these compressive/distraction forces. Ten mandibles were obtained from 5 canine cadavers. Paired mandibles from the same cadaver were each fixated to a mandibular distractor with eight screws (either titanium or PLLA). Devices were each set to 15 and 30 mm of distraction distance. Compression force of 80 N was then generated parallel to the axis of the distraction device. Distractor displacement was measured to detect any mechanical failure during this pre-set load. Finally, if no failure was observed at 80 N, a load-to-failure compression test was done in the PLLA group to determine the mechanical failure point. All distractors in both the titanium and PLLA screw groups withstood 80 N of compression without failure. When the load-to-failure test was performed in the PLLA group, the average device failure point was 172.8 N (range 148-196 N). Review of high-frame-rate video demonstrated that all failures occurred due to the PLLA screws breaking or falling out. Bioabsorbable PLLA screws can withstand compressive forces more than double that of the maximal in vivo forces needed during mandibular distraction. These screws may be an acceptable alternative for the fixation of internal mandibular distractors.
引用
收藏
页码:2228 / 2231
页数:4
相关论文
共 50 条
  • [41] A mouse model of mandibular distraction osteogenesis
    Fang, TD
    Nacamuli, RP
    Song, HM
    Thomas, R
    Fong, KD
    Aalami, OO
    Longaker, MT
    CRANIOFACIAL SURGERY 10, 2003, : 397 - 398
  • [42] Mandibular overgrowth after distraction osteogenesis
    Rahman, M. M.
    Stewart, D. A.
    Gianoutsos, M. P.
    CRANIOFACIAL SURGERY: PROCEEDINGS OF THE TWELFTH INTERNATIONAL CONGRESS OF THE INTERNATIONAL SOCIETY OF CRANIOFACIAL SURGERY, 2007, : 335 - 337
  • [43] Mandibular widening by intraoral distraction osteogenesis
    Guerrero, CA
    Bell, WH
    Contasti, GI
    Rodriguez, AM
    BRITISH JOURNAL OF ORAL & MAXILLOFACIAL SURGERY, 1997, 35 (06): : 383 - 392
  • [44] Mandibular distraction osteogenesis in the pediatric patient
    Tibesar, Robert J.
    Sidman, James D.
    CURRENT OPINION IN OTOLARYNGOLOGY & HEAD AND NECK SURGERY, 2008, 16 (06): : 548 - 554
  • [45] Analysis of the performance of different orthodontic devices for mandibular symphyseal distraction osteogenesis
    Boccaccio, A.
    Cozzani, M.
    Pappalettere, C.
    EUROPEAN JOURNAL OF ORTHODONTICS, 2011, 33 (02) : 113 - 120
  • [46] Mandibular distraction osteogenesis: a comparison of distraction rates in the rabbit model
    Stewart, KJ
    Lvoff, GO
    White, SA
    Bonar, SF
    Walsh, WR
    Smart, RC
    Poole, MD
    JOURNAL OF CRANIO-MAXILLOFACIAL SURGERY, 1998, 26 (01) : 43 - 49
  • [47] Denervation During Mandibular Distraction Osteogenesis Results in Impaired Osteogenesis
    Tevlin, Ruth
    Januszyk, Michael
    Griffin, Michelle
    Salhotra, Ankit
    Wan, Derrick C.
    Chan, Charles K. F.
    Longaker, Michael T.
    JOURNAL OF THE AMERICAN COLLEGE OF SURGEONS, 2021, 233 (05) : S196 - S197
  • [48] Reconstruction of segmental mandibular defects by distraction osteogenesis for mandibular reconstruction
    Kuriakose, MA
    Shnayder, Y
    DeLacure, MD
    HEAD AND NECK-JOURNAL FOR THE SCIENCES AND SPECIALTIES OF THE HEAD AND NECK, 2003, 25 (10): : 816 - 824
  • [49] Distraction Osteogenesis Plus Permanent Rigid Fixation for Mandibular Deformities in Nongrowing Patients: A Comparative Study
    Mahmood, Barakat Abdelreheem
    JOURNAL OF CRANIOFACIAL SURGERY, 2021, 32 (08) : E693 - E695
  • [50] DISTRACTION OSTEOGENESIS FOR RECONSTRUCTION OF MANDIBULAR SYMPHYSEAL DEFECTS
    ANNINO, DJ
    GOGUEN, LA
    KARMODY, CS
    ARCHIVES OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 1994, 120 (09) : 911 - 916