Single isocenter HyperArc treatment of multiple intracranial metastases: Targeting accuracy

被引:1
|
作者
Li, Fang [1 ,5 ]
Mail, Noor [1 ]
di Mayorca, Maria Stefania [2 ]
Mccaw, Travis J. [1 ]
Ozhasoglu, Cihat [1 ]
Lalonde, Ronald [1 ]
Chang, Jina [1 ]
Huq, Mohammed Saiful [3 ,4 ]
机构
[1] UPMC Hillman Canc Ctr, Radiat Oncol, Pittsburgh, PA USA
[2] Montefiore Med Ctr, Radiat Oncol, Bronx, NY USA
[3] Univ Pittsburgh, Sch Med, Pittsburgh, PA USA
[4] UPMC Hillman Canc Ctr, Pittsburgh, PA USA
[5] UPMC Hillman Canc Ctr, Radiat Oncol, 5230 Ctr Ave,Ground Floor,MG21, Pittsburgh, PA 15232 USA
来源
关键词
HyperArc; intracranial SRS/SBRT; LINEAR-ACCELERATOR; RADIOSURGERY; SRS; SYSTEMS; LESIONS;
D O I
10.1002/acm2.14234
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose/objectives(A) To examine the alignment accuracy of CBCT guidance for brain metastases with off centered isocenters, (B) to test dose delivery and targeting accuracy for single isocenter treatments with multiple brain metastases. We report the results of the end-to-end test for Truebeam stereotactic radiosurgery (SRS).Materials/methodsAn anthropomorphic CT head phantom was drilled with five MOSFET inserts and two PTW Pinpoint chamber inserts. The phantom was simulated, planned, and delivered. For the purpose of comparing the accuracy of alignment, CBCTs were acquired with the isocenter centered and offset superiorly 8 cm, inferiorly 8 cm, anteriorly 7 cm, posteriorly 7 cm, and right 5 cm. There were six degrees of freedom corrections applied to the plans, as well as intentional rotational and translational errors for dose comparisons. Dose accuracy checks were performed with MOSFET and PTW Pinpoint chamber, and targeting accuracy was assessed with GafChromic films.Result(A) Compared to centered CBCT, off-centered CBCT scan showed some alignment errors, with a maximum difference of 0.6-degree pitch and 0.9 mm translation when the phantom was placed 8 cm inferior off center. (B) For the single isocenter plan, measured doses of the five MOSFET were 95%-100% of the planned dose, whereas the multiple isocenter plans were 96%-100%. With intentional setup errors of 1-degree pitch, doses were 97.1%-100.4% compared to the perfect setup. The same was found for the two pinpoint chamber readings with 1-degree rotation and 1 mm translation. (C) Targeting accuracy for targets at the isocenter is 0.67 mm, within the machine specification of 0.75 mm. Targeting accuracy for isocenters 6-12 cm away from the target is in the range 0.67-1.18 mm.Conclusion(A) Single isocenter HyperArc treatments for multiple brain metastases are feasible and targeting accuracy is clinically acceptable. (B) The vertex in a cranial scan is very important for proper alignment.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Implementation of HyperArc for Single Isocenter Frameless Stereotactic Radiosurgery (SRS) of Multiple Brain Metastases
    Mutanga, Theodore
    Wang, Yizhen
    Giovinazzo, Julia
    Fineberg, Heather
    Sankreacha, Raxa
    Brade, Anthony
    Khan, Luluel
    [J]. MEDICAL PHYSICS, 2022, 49 (08) : 5676 - 5676
  • [2] Targeting Accuracy in the Stereotactic Treatment of Multiple Brain Lesions Using the Single Isocenter Technique
    Han, Z.
    Hacker, F.
    Friesen, S.
    Hu, Y.
    Bredfeldt, J.
    Beaudette, K.
    Cail, D.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E659 - E659
  • [3] Frameless single-isocenter intensity modulated stereotactic radiosurgery for simultaneous treatment of multiple intracranial metastases
    Lau, Steven K. M.
    Zhao, Xiao
    Carmona, Ruben
    Knipprath, Erik
    Simpson, Daniel R.
    Nath, Sameer K.
    Kim, Gwe-Ya
    Hattangadi, Jona A.
    Chen, Clark C.
    Murphy, Kevin T.
    [J]. TRANSLATIONAL CANCER RESEARCH, 2014, 3 (04) : 383 - 390
  • [4] Single Isocenter Frameless Volumetric Modulated Arc Radiosurgery for Multiple Intracranial Metastases
    Zhao, X.
    Lau, S. K. M.
    Zakeri, K.
    Carmona, R., Jr.
    Knipprath, E.
    Simpson, D. R.
    Nath, S. K.
    Kim, G. Y.
    Sanghvi, P.
    Hattangadi, J. A.
    Chen, C.
    Murphy, K. T.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 93 (03): : E92 - E93
  • [5] A Novel 3 Step Patient Specific QA Process for Single Isocenter Multiple Brain Metastases SRS (HyperArc)
    Desai, P.
    Godwin, G.
    Shakeshaft, J.
    Aland, T.
    Kamst, O.
    Anderson, L.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E645 - E645
  • [6] Dosimetric Evaluation On Single-Isocenter Versus Multiple-Isocenter Frameless Volumetric Modulated Arc Radiosurgery for Simultaneous Treatment of Multiple Intracranial Metastases
    Huang, Z.
    Feng, Y.
    Luo, C.
    Liu, T.
    Zhu, D.
    Lo, S.
    Mayr, N.
    Yuh, W.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E651 - E651
  • [7] Evaluation of Dosimetric Accuracy in the Treatment of Multiple Brain Metastases With Volumetric Modulated Arc Therapy and Single Isocenter
    Vieira, A.
    Sakuraba, R.
    Goncalves, V.
    Cruz, J.
    Weltman, E.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 84 (03): : S872 - S872
  • [8] On the Spatial Accuracy of Single Isocenter Multiple Brain Metastases SRS with BrainLab Elements
    Saenz, D.
    McCulloch, J.
    Rasmussen, K.
    Kirby, N.
    Myers, P.
    Stathakis, S.
    Papanikolaou, N.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E144 - E145
  • [9] Single-Isocenter Frameless Volumetric Modulated Arc Radiosurgery for Multiple Intracranial Metastases
    Lau, Steven K. M.
    Zakeri, Kaveh
    Zhao, Xiao
    Carmona, Ruben
    Knipprath, Erik
    Simpson, Daniel R.
    Nath, Sameer K.
    Kim, Gwe-Ya
    Sanghvi, Parag
    Hattangadi-Gluth, Jona A.
    Chen, Clark C.
    Murphy, Kevin T.
    [J]. NEUROSURGERY, 2015, 77 (02) : 233 - 240
  • [10] Dosimetric evaluation of Eclipse SRS HyperArc plans for single isocenter multiple targets
    Calvo-Ortega, J.
    Laosa-Bello, C.
    Moragues-Femenia, S.
    Pozo-Masso, M.
    Casals-Farran, J.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S1314 - S1315