Gantry-Free High-Resolution Cone- Beam CT: Efficacy for Distal Radius and Scaphoid Fracture Detection and Characterization

被引:5
|
作者
Grunz, Jan-Peter [1 ]
Jordan, Martin C. [2 ]
Schmitt, Rainer [1 ,3 ]
Luetkens, Karsten S. [1 ]
Huflage, Henner [1 ]
Meffert, Rainer H. [2 ]
Bley, Thorsten A. [1 ]
Kunz, Andreas S. [1 ]
机构
[1] Univ Hosp Wurzburg, Dept Diagnost & Intervent Radiol, Wurzburg, Bavaria, Germany
[2] Univ Hosp Wurzburg, Dept Trauma Hand Plast & Reconstruct Surg, Wurzburg, Bavaria, Germany
[3] Ludwig Maximilians Univ Munchen, Univ Hosp, Dept Radiol, Munich, Bavaria, Germany
关键词
Cone-beam computed tomography; Wrist; Trauma; Fracture; Radiography; COMPUTED-TOMOGRAPHY; DISPLACEMENT; RADIOGRAPHS; DIAGNOSIS;
D O I
10.1016/j.acra.2022.08.030
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: Gantry-free cone-beam CT (CBCT) allows for comfortable patient positioning due to an open scanner architecture. Since CBCT without gantry is not yet established for clinical wrist trauma imaging, this study's aim was to investigate its diagnostic value in the preoperative workup of patients with distal radius and scaphoid fractures.Methods: Within a 12-month period, 113 patients with severe wrist trauma underwent both radiography and CBCT with the same gantry-free multi-use scanner before surgery. Two radiologists retrospectively analyzed all datasets for the morphology of distal radius (n = 95) and scaphoid fractures (n = 20). In all 115 wrists (two bilateral injuries), surgical reports served as the standard of reference. Results: While accuracy for distal radius fractures was comparable among CBCT and radiographs, the former was superior with regard to scaphoid fractures (Reader 1: 100.0% vs. 75.0%; Reader 2: 100.0% vs. 65.0%). Accuracy for multi-fragmentary radius injuries (100.0% vs. 90.5%; 100.0% vs. 93.7%), and articular affliction (99.0% vs. 84.2%; 100.0% vs. 83.2%) was also higher in CBCT. Regarding scaph-oid fractures, CBCT proved superior for diagnosis of proximal pole or waist involvement (100.0% vs. 70.0%; 100.0% vs. 65.0%) and com-minuted patterns (100.0% vs. 70.0%; 100.0% vs. 75.0%). Median effective dose of CBCT was as low as 3.65 mSv compared with 0.16 mSv for standard radiography.Conclusion: Gantry-free CBCT allows for excellent diagnostic accuracy in the assessment of distal radius and scaphoid fracture morphol-ogy. Even in patients with limited mobility, very low radiation dose is sufficient to maintain high image quality.
引用
收藏
页码:1358 / 1366
页数:9
相关论文
共 50 条
  • [21] Three-rooted premolar analyzed by high-resolution and cone beam CT
    Caroline Marca
    Paul M. H. Dummer
    Susan Bryant
    Fabiana Vieira Vier-Pelisser
    Marcus Vinicius Reis Só
    Vania Fontanella
    Vinicius D’avila Dutra
    José Antonio Poli de Figueiredo
    Clinical Oral Investigations, 2013, 17 : 1535 - 1540
  • [22] Three-rooted premolar analyzed by high-resolution and cone beam CT
    Marca, Caroline
    Dummer, Paul M. H.
    Bryant, Susan
    Vier-Pelisser, Fabiana Vieira
    Reis So, Marcus Vinicius
    Fontanella, Vania
    Dutra, Vinicius D'avila
    Poli de Figueiredo, Jose Antonio
    CLINICAL ORAL INVESTIGATIONS, 2013, 17 (06) : 1535 - 1540
  • [23] Multiresolution iterative reconstruction in high-resolution extremity cone-beam CT
    Cao, Qian
    Zbijewski, Wojciech
    Sisniega, Alejandro
    Yorkston, John
    Siewerdsen, Jeffrey H.
    Stayman, J. Webster
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (20): : 7263 - 7281
  • [24] Ultrasound simulation in the distal radius using clinical high-resolution peripheral-CT images
    Le Floch, Vincent
    Mcmahon, Donald J.
    Luo, Gangming
    Cohen, Adi
    Kaufman, Jonathan J.
    Shane, Elizabeth
    Siffert, Robert S.
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2008, 34 (08): : 1317 - 1326
  • [25] Modeling and evaluation of a high-resolution CMOS detector for cone-beam CT of the extremities
    Cao, Qian
    Sisniega, Alejandro
    Brehler, Michael
    Stayman, J. Webster
    Yorkston, John
    Siewerdsen, Jeffrey H.
    Zbijewski, Wojciech
    MEDICAL PHYSICS, 2018, 45 (01) : 114 - 130
  • [26] Load transfer analysis of the distal radius from in-vivo high-resolution CT-imaging
    Ulrich, D
    van Rietbergen, B
    Laib, A
    Rüegsegger, P
    JOURNAL OF BIOMECHANICS, 1999, 32 (08) : 821 - 828
  • [27] Image-Based Motion Compensation for High-Resolution Extremities Cone-Beam CT
    Sisniega, A.
    Stayman, J. W.
    Cao, Q.
    Yorkston, J.
    Siewerdsen, J. H.
    Zbijewski, W.
    MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING, 2016, 9783
  • [28] High-resolution cone beam CT for evaluation of vascular channel in intracranial partial thrombosed aneurysm
    Shintai, Kazunori
    Matsubara, Noriaki
    Izumi, Takashi
    NAGOYA JOURNAL OF MEDICAL SCIENCE, 2018, 80 (02): : 279 - 284
  • [29] Quantitative analysis of high-resolution, contrast-enhanced, cone-beam CT for the detection of intracranial in-stent hyperplasia
    Flood, Thomas F.
    van der Bom, Imramsjah M. J.
    Strittmatter, Lara
    Puri, Ajit S.
    Hendricks, Gregory M.
    Wakhloo, Ajay K.
    Gounis, Matthew J.
    JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2015, 7 (02) : 118 - 125
  • [30] Comparison of high-resolution and standard zoom imaging modes in cone beam computed tomography for detection of longitudinal root fracture: An in vitro study
    Taramsari, Mehran
    Kajan, Zahra Dalili
    Bashirzadeh, Parinaz
    Salamat, Fatemeh
    IMAGING SCIENCE IN DENTISTRY, 2013, 43 (03) : 171 - 177