Augmented reality-based surgical navigation of pelvic screw placement: an ex-vivo experimental feasibility study

被引:4
|
作者
Heining, Sandro-Michael [1 ]
Raykov, Vladislav [2 ]
Wolff, Oliver [3 ]
Alkadhi, Hatem [4 ]
Pape, Hans-Christoph [1 ]
Wanner, Guido A. [5 ]
机构
[1] Univ Hosp Zurich, Dept Traumatol, Zurich, Switzerland
[2] Landeskrankenhaus Bludenz, Dept Orthoped & Traumatol, Bludenz, Austria
[3] Hsch Luzern Tech & Architektur, Luzern, Switzerland
[4] Univ Hosp Zurich, Dept Radiol, Zurich, Switzerland
[5] Private Hosp Bethanien, Spine Clin & Traumatol, Swiss Med Network, Zurich, Switzerland
关键词
Augmented reality; Surgical navigation; Pelvic trauma; Head-mounted display; INSERTION; FIXATION; CORRIDOR; SURGERY;
D O I
10.1186/s13037-023-00385-6
中图分类号
R61 [外科手术学];
学科分类号
摘要
BackgroundMinimally invasive surgical treatment of pelvic trauma requires a significant level of surgical training and technical expertise. Novel imaging and navigation technologies have always driven surgical technique, and with head-mounted displays being commercially available nowadays, the assessment of such Augmented Reality (AR) devices in a specific surgical setting is appropriate.MethodsIn this ex-vivo feasibility study, an AR-based surgical navigation system was assessed in a specific clinical scenario with standard pelvic and acetabular screw pathways. The system has the following components: an optical-see-through Head Mounted Display, a specifically designed modular AR software, and surgical tool tracking using pose estimation with synthetic square markers.ResultsThe success rate for entry point navigation was 93.8%, the overall translational deviation of drill pathways was 3.99 +/- 1.77 mm, and the overall rotational deviation of drill pathways was 4.3 +/- 1.8 degrees. There was no relevant theoretic screw perforation, as shown by 88.7% Grade 0-1 and 100% Grade 0-2 rating in our pelvic screw perforation score. Regarding screw length, 103 +/- 8% of the planned pathway length could be realized successfully.ConclusionThe novel innovative system assessed in this experimental study provided proof-of-concept for the feasibility of percutaneous screw placement in the pelvis and, thus, could easily be adapted to a specific clinical scenario. The system showed comparable performance with other computer-aided solutions while providing specific advantages such as true 3D vision without intraoperative radiation; however, it needs further improvement and must still undergo regulatory body approval. Future endeavors include intraoperative registration and optimized tool tracking.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] A Novel Pedicle Screw Placement Surgery Based on Integration of Surgical Guides and Augmented Reality
    Kong, Huiyang
    Wang, Shuyi
    Zhang, Can
    Chen, Zan
    Wu, Zhanglei
    Wang, Jiayu
    JOURNAL OF NEUROLOGICAL SURGERY PART A-CENTRAL EUROPEAN NEUROSURGERY, 2025,
  • [22] Development of Augmented Reality-Based Neuro-Navigation System for use in External Ventrictilar Drain Placement
    Cutler, Andrew Benjamin
    Rahimpour, Shervin
    Liu, Yameng
    Lad, Nandan
    Kopper, Regis
    Codd, Patrick
    JOURNAL OF NEUROSURGERY, 2017, 126 (04) : A1438 - A1438
  • [23] The Accuracy of Cup Placement in Total Hip Arthroplasty (THA) Using an Augmented Reality-Based Navigation System
    Shimizu, Junya
    Nagoya, Satoshi
    Kosukegawa, Ima
    Kanaizumi, Arata
    Nakahashi, Naoya
    Teramoto, Atsushi
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (04)
  • [24] Feasibility of mixed reality-based intraoperative three-dimensional image-guided navigation for atlanto-axial pedicle screw placement
    Wu, Xinghuo
    Liu, Rong
    Xu, Song
    Yang, Cao
    Yang, Shuhua
    Shao, Zengwu
    Li, Suyun
    Ye, Zhewei
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2019, 233 (12) : 1310 - 1317
  • [25] Development and evaluation of robot-assisted ultrasound navigation system for pedicle screw placement: An ex-vivo animal validation
    Li, Ruixuan
    Davoodi, Ayoob
    Cai, Yuyu
    Borghesan, Gianni
    Cavalcanti, Nicola
    Laux, Christoph J.
    Farshad, Mazda
    Carrillo, Fabio
    Fuernstahl, Philipp
    Vander Poorten, Emmanuel
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2024, 20 (01):
  • [26] Mixed Reality-Based Navigation for Pedicle Screw Placement: A Preliminary Study Using a 3D-Printed Spine Model
    Ohashi, Masayuki
    Sato, Masayuki
    Tashi, Hideki
    Minato, Keitaro
    Makino, Tatsuo
    Kawashima, Hiroyuki
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (04)
  • [27] Minimally Invasive Transforaminal Lumbar Interbody Fusion Using Augmented Reality Surgical Navigation for Percutaneous Pedicle Screw Placement
    Charles, Yann P.
    Cazzato, Roberto L.
    Nachabe, Rami
    Chatterjea, Anindita
    Steib, Jean-Paul
    Gangi, Afshin
    CLINICAL SPINE SURGERY, 2021, 34 (07): : E415 - E424
  • [28] Augmented Reality-Based Navigation System for Wrist Arthroscopy: Feasibility (vol 2, page4 , 2013)
    Zemirline, Ahmed
    Agnus, Vincent
    Soler, Luc
    Mathoulin, Christophe L.
    Obdeijn, Miryam
    Liverneaux, Philippe A.
    JOURNAL OF WRIST SURGERY, 2014, 3 (04) : E2 - E2
  • [29] Augmented reality and artificial intelligence-assisted surgical navigation: Technique and cadaveric feasibility study
    Siemionow, Kris B.
    Katchko, Karina M.
    Lewicki, Paul
    Luciano, Cristian J.
    JOURNAL OF CRANIOVERTEBRAL JUNCTION AND SPINE, 2020, 11 (02): : 81 - 85
  • [30] Simulated augmented reality-based calibration of optical see-through head mound display for surgical navigation
    Ha, Ho-Gun
    Gu, Kyeongmo
    Jeung, Deokgi
    Hong, Jaesung
    Lee, Hyunki
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2024, 19 (08) : 1647 - 1657