Comparison of Cerebrospinal Fluid Cellular Analysis by Optical Microscopy and Sysmex XN 9000

被引:1
|
作者
Ren, Xiu-Qin [1 ]
Jiang, Xi [1 ]
Pi, Ying-Qi [1 ]
Wu, Zhi-Kun [1 ]
Jiang, Hong-bo [1 ]
Wang, Hua-Cheng [1 ,2 ]
机构
[1] Sun Yat Sen Univ, Affiliated Hosp 7, Shenzhen, Peoples R China
[2] Sun Yat Sen Univ, Affiliated Hosp 7, 628 Zhenyuan Rd, Shenzhen 518107, Peoples R China
关键词
Sysmex XN9000; cellular analysis; body fluid module; cell slide centrifuge; cerebrospinal fluid; PERFORMANCE; GUIDELINES; XE-5000; MODE;
D O I
10.7754/Clin.Lab.2023.230403
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: This study was designed to compare the body fluid module of Sysmex XN9000 (XN-BF) with optical microscopy (OM) for cerebrospinal fluid (CSF) analysis after two-step cell slide centrifuge (TSCSC), defining the best procedure for CSF optical microscopy analysis. Methods: Items of RBC, WBC enumeration and differentiation were observed. The cell count and morphologic evaluation of the cellular composition by OM was carried out both with and without two-step cell slide centrifuge (TSCSC) and were compared the data with XN-BF. Results: There were 69.98 +/- 4.94 RBC and 36.98 +/- 3.39 WBC in one OSCSC microscopic field whereas there were 96.35 +/- 5.41 RBC and 66.15 +/- 4.85 WBC in one TSCSC microscopic field in the same sample (*200). There was a statistical difference between those two methods (p = 0.000). Excellent correlation was found between total cell count with both OM and XN-BF. The R2 value for RBC and WBC counts were 0.99 and 0.96, respectively. For WBC differential, the R2 values were 0.98 for PMN and 0.70 for MN. Correlation of MN was poorer than PMN. As far as the tumor cell, phagocyte, and plasma cell with high fluorescence were concerned, OM were not consistent with XN-BF. Conclusions: The TSCSC procedure contributes to the separation of cells and other ingredients. XN-BF displays excellent performance at RBC and WBC cell count except for mononuclear cells, tumor cells, phagocytes, and leukemia cells. which makes it just a practical alternative to total cell (WBC, RBC) count for CSF samples. Detailed morphologic workup of CSF samples is mandated in all cases with meningoencephalitis, elevated cell count, sub-arachnoid hemorrhage and meningeal carcinomatosis, the TSCSC procedure is recommended. (Clin. Lab. 2023;69:2274-2279. DOI: 10.7754/Clin.Lab.2023.230403)
引用
收藏
页码:2274 / 2279
页数:6
相关论文
共 50 条
  • [31] METHOD COMPARISON AND CLINICAL ASSESSMENT OF BIOLOGICAL FLUID MODE (BF) ON THE XN-1000 VS XE-5000 SYSMEX ANALYZERS.
    Martinez-Villanueva, Miriam
    Lozoya-Angulo, Elena
    Martinez-Lopez de Ca, Ana
    Antonio Noguera-Velasco, Jose
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2014, 36 : 35 - 35
  • [32] Comparison of basophil count by Beckman Coulter UniCel DxH 800, Sysmex XN-1000, and manual microscopy in cases of suspected chronic myeloid leukemia
    Chopra, Parul
    Bhardwaj, Sunanda
    Arora, Anil
    IRAQI JOURNAL OF HEMATOLOGY, 2021, 10 (02) : 91 - 96
  • [33] Optical analysis of suspended particles in the cerebrospinal fluid obtained by puncture from patients diagnosed with the disorders of cerebrospinal fluid (CSF) circulation
    Staron, Waldemar
    Herbowski, Leszek
    Gurgul, Henryk
    15TH CZECH-POLISH-SLOVAK CONFERENCE ON WAVE AND QUANTUM ASPECTS OF CONTEMPORARY OPTICS, 2007, 6609
  • [34] Analysis of cerebrospinal fluid cells by flow cytometry: Comparison to conventional cytology
    Zeman, David
    Revendova, Kamila
    Bunganic, Radovan
    Ryzi, Marketa
    Masarovicova, Petra
    Kusnierova, Pavlina
    Kotrlova, Vera
    Hradilek, Pavel
    Stejskal, David
    Thon, Vojtech
    BIOMEDICAL PAPERS-OLOMOUC, 2023, 167 (02): : 121 - 130
  • [35] Proteomic analysis of astrocytic secretion in the mouse -: Comparison with the cerebrospinal fluid proteome
    Lafon-Cazal, M
    Adjali, O
    Galélotti, N
    Poncet, J
    Jouin, P
    Homburger, V
    Bockaert, J
    Marin, P
    JOURNAL OF BIOLOGICAL CHEMISTRY, 2003, 278 (27) : 24438 - 24448
  • [36] Comparison of cerebellomedullary and lumbar cerebrospinal fluid analysis in dogs with neurological disease
    Lampe, Rachel
    Foss, Kari D.
    Vitale, Samantha
    Hague, Devon W.
    Barger, Anne M.
    JOURNAL OF VETERINARY INTERNAL MEDICINE, 2020, 34 (02) : 838 - 843
  • [37] BODY FLUID MODE ON THE SYSMEX XN-10 AUTOMATED COUNTER: PERFORMANCE EVALUATION, CELL COUNTS AND NEW ANALYSIS OF CELL POPULATION DATA
    Gerard, Delphine
    Henry, Sylvain
    Chatelain, Bernard
    Lesesve, Jean-Francois
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2018, 40 : 46 - 46
  • [38] Comparison of cerebrospinal fluid results analysis against meningitis pathogens in meningitis patients
    Mukti, Magendi Indra
    Aryati, Paulus
    Sugianto, Paulus
    BALI MEDICAL JOURNAL, 2023, 12 (03) : 2752 - 2757
  • [39] Validation of reflex testing rules and establishment of a new workflow for body fluid cell analysis using a Sysmex XN-550 automatic hematology analyzer
    Cho, Y-U
    You, E.
    Jang, S.
    Park, C-J
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2018, 40 (03) : 258 - 267
  • [40] Comparison of two quantitative and one semiquantitative method for the analysis of protein in cerebrospinal fluid of dogs
    Rebel, K
    Fischer, A
    TIERARZTLICHE PRAXIS AUSGABE KLEINTIERE HEIMTIERE, 2002, 30 (01): : 70 - 76