Comparison of Cerebrospinal Fluid Cellular Analysis by Optical Microscopy and Sysmex XN 9000

被引:1
|
作者
Ren, Xiu-Qin [1 ]
Jiang, Xi [1 ]
Pi, Ying-Qi [1 ]
Wu, Zhi-Kun [1 ]
Jiang, Hong-bo [1 ]
Wang, Hua-Cheng [1 ,2 ]
机构
[1] Sun Yat Sen Univ, Affiliated Hosp 7, Shenzhen, Peoples R China
[2] Sun Yat Sen Univ, Affiliated Hosp 7, 628 Zhenyuan Rd, Shenzhen 518107, Peoples R China
关键词
Sysmex XN9000; cellular analysis; body fluid module; cell slide centrifuge; cerebrospinal fluid; PERFORMANCE; GUIDELINES; XE-5000; MODE;
D O I
10.7754/Clin.Lab.2023.230403
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: This study was designed to compare the body fluid module of Sysmex XN9000 (XN-BF) with optical microscopy (OM) for cerebrospinal fluid (CSF) analysis after two-step cell slide centrifuge (TSCSC), defining the best procedure for CSF optical microscopy analysis. Methods: Items of RBC, WBC enumeration and differentiation were observed. The cell count and morphologic evaluation of the cellular composition by OM was carried out both with and without two-step cell slide centrifuge (TSCSC) and were compared the data with XN-BF. Results: There were 69.98 +/- 4.94 RBC and 36.98 +/- 3.39 WBC in one OSCSC microscopic field whereas there were 96.35 +/- 5.41 RBC and 66.15 +/- 4.85 WBC in one TSCSC microscopic field in the same sample (*200). There was a statistical difference between those two methods (p = 0.000). Excellent correlation was found between total cell count with both OM and XN-BF. The R2 value for RBC and WBC counts were 0.99 and 0.96, respectively. For WBC differential, the R2 values were 0.98 for PMN and 0.70 for MN. Correlation of MN was poorer than PMN. As far as the tumor cell, phagocyte, and plasma cell with high fluorescence were concerned, OM were not consistent with XN-BF. Conclusions: The TSCSC procedure contributes to the separation of cells and other ingredients. XN-BF displays excellent performance at RBC and WBC cell count except for mononuclear cells, tumor cells, phagocytes, and leukemia cells. which makes it just a practical alternative to total cell (WBC, RBC) count for CSF samples. Detailed morphologic workup of CSF samples is mandated in all cases with meningoencephalitis, elevated cell count, sub-arachnoid hemorrhage and meningeal carcinomatosis, the TSCSC procedure is recommended. (Clin. Lab. 2023;69:2274-2279. DOI: 10.7754/Clin.Lab.2023.230403)
引用
收藏
页码:2274 / 2279
页数:6
相关论文
共 50 条
  • [1] Optimization of Cellular analysis of Synovial Fluids by optical microscopy and automated count using the Sysmex XN Body Fluid Mode
    Seghezzi, Michela
    Buoro, Sabrina
    Manenti, Barbara
    Mecca, Tommaso
    Ferrari, Roberto
    Zappala, Giorgio
    Castelli, Claudio Carlo
    Balboni, Fiamma
    Pezzati, Paola
    Ottomano, Cosimo
    Lippi, Giuseppe
    CLINICA CHIMICA ACTA, 2016, 462 : 41 - 48
  • [2] AUTOMATED CEREBROSPINAL FLUID CELL COUNTS IN THE SYSMEX XN-SERIES
    Tejerina, Patricia
    Serrando, Maite
    Manuel Ramirez, Jose
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2014, 36 : 89 - 89
  • [3] BLOOD CELLS ANALYSIS IN CEREBROSPINAL FLUID USING SYSMEX® XN-1000 HEMATOLOGY ANALYSER
    Alic, Elizabeta Boznar
    Kase, Simona
    Skitek, Milan
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2014, 36 : 81 - 81
  • [4] Two cases with discrepancy in the quantitative cytological assessment of cerebrospinal fluid in neonatal samples using light microscopy in comparison with Sysmex XN-1000
    Broz, Pavel
    Kukralova, Simona
    Palatova, Jana
    Kolduskova, Klara
    Zenkova, Jana
    Rajdl, Daniel
    Racek, Jaroslav
    BIOCHEMIA MEDICA, 2024, 34 (02)
  • [5] Cytological examination of cerebrospinal fluid: Sysmex OF-1000i versus optical microscopy
    Maleb, A.
    Bouayadi, O.
    El Malki, J.
    Rifai, S.
    Lamrabat, S.
    Benaissa, E.
    Ben Lahlou, Y.
    Frikh, M.
    Elouennass, M.
    ANALYTICAL BIOCHEMISTRY, 2020, 589
  • [6] Performance of the Sysmex XN-V body fluid module for canine cerebrospinal fluid cell count
    Ortiz-Nisa, Sonia
    Sanz, Alba
    Pastor, Josep
    De la Fuente, Cristian
    Anor, Sonia
    VETERINARY CLINICAL PATHOLOGY, 2021, 50 (03) : 359 - 368
  • [7] Comparison between "body fluid" mode of two hematology analyzers and the microscopy method for cellular analysis of the cerebrospinal fluid
    Seghezzi, Michela
    Mecca, Tommaso
    Dominoni, Paola
    Lochis, Eleonora
    Saleri, Cheren
    Manenti, Barbara
    Crippa, Alberto
    Buoro, Sabrina
    BIOCHIMICA CLINICA, 2016, 40 (04) : 322 - 327
  • [8] Comparison between optical microscopy and the Sysmex XN-3000 for schistocyte determination in patients suspected of having schistocytosis
    Hantaweepant, Chattree
    Sasijareonrat, Natthaporn
    Chutvanichkul, Boonyanuch
    Karaketklang, Khemajira
    Chinthammitr, Yingyong
    HEALTH SCIENCE REPORTS, 2020, 3 (01)
  • [9] A New Automated Technology for Cerebrospinal Fluid Cell Counts Comparison of Accuracy and Clinical Impact of GloCyte, Sysmex XN, and Manual Methods
    Sandhaus, Linda M.
    Dillman, Christine A.
    Hinkle, Warren P.
    MacKenzie, Julie M.
    Hong, George
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2017, 147 (05) : 507 - 514
  • [10] Designing and Validating Autoverification Rules for Hematology Analysis in Sysmex XN-9000 Hematology System
    Fu, Qiang
    Ye, Congxiu
    Han, Bo
    Zhan, Xiaoxia
    Chen, Kang
    Huang, Fuda
    Miao, Lisao
    Yang, Shanhong
    Wang, Weijia
    CLINICAL LABORATORY, 2020, 66 (04) : 549 - 556