Comparison of Home and Office Blood Pressure Devices in the Clinical Setting

被引:2
|
作者
Juraschek, Stephen P. [1 ,2 ,3 ]
Vyavahare, Medha [2 ]
Cluett, Jennifer L. [1 ,2 ,3 ]
Turkson-Ocran, Ruth-Alma [1 ,2 ,3 ]
Mukamal, Kenneth J. [1 ,2 ]
Ishak, Anthony M. [2 ,3 ]
机构
[1] Harvard Med Sch, Beth Israel Deaconess Med Ctr, Dept Med, Boston, MA 02115 USA
[2] Harvard Med Sch, Dept Med, Boston, MA 02115 USA
[3] Beth Israel Deaconess Med Ctr, Healthcare Associates, Dept Med, Boston, MA 02215 USA
关键词
blood pressure; home blood pressure monitoring; hypertension; pharmacist; self-measured blood pressure; validation; VALIDATION; STANDARD;
D O I
10.1093/ajh/hpad120
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
BACKGROUND: Self-measured blood pressure (SMBP) monitoring is increasingly used for remote hypertension management, but the real-world performance of home blood pressure (BP) devices is unknown. We examined BP measurements from patients' home devices using the American Medical Association's (AMA) SMBP Device Accuracy Test tool.METHODS: Patients at a single internal medicine clinic underwent up to five seated, same-arm BP readings using a home device and an automated BP device (Omron HEM-907XL). Following the AMA's three-step protocol, we used the patient's home device for the first, second, and fourth measurements and the office device for the third and fifth (if needed) measurements. Device agreement failure was defined as an absolute difference in systolic BP >10 mm Hg between the home and office devices in either of two confirmatory steps. Performance was examined by brand (Omron vs. non-Omron). Moreover, we examined patient factors associated with agreement failure via logistic regression models adjusted for demographic characteristics.RESULTS: We evaluated 152 patients (mean age 60 +/- 15 years, 58% women, 31% Black) seen between October 2020 and November 2021. Device agreement failure occurred in 22.4% (95% CI: 16.4%, 29.7%) of devices tested, including 19.1% among Omron devices and 27.6% among non-Omron devices (P = 0.23). No patient characteristics were associated with agreement failure.CONCLUSIONS: Over one-fifth of home devices did not agree based on the AMA SMBP device accuracy protocol. These findings confirm the importance of office-based device comparisons to ensure the accuracy of home BP monitoring.
引用
收藏
页码:342 / 348
页数:7
相关论文
共 50 条
  • [1] Office blood pressure measurements with oscillometric devices in adolescents: a comparison with home blood pressure
    Jardim, Thiago Veiga
    Gaziano, Thomas A.
    Nascente, Flavia Miquetichuc
    Carneiro, Carolina de Souza
    Morais, Polyana
    Roriz, Vanessa
    Mendonca, Karla Lorena
    Rolim Povoa, Thais Inacio
    Sebba Barroso, Weimar Kunz
    Lima Sousa, Ana Luiza
    Veiga Jardim, Paulo Cesar
    BLOOD PRESSURE, 2017, 26 (05) : 272 - 278
  • [2] Comparison of home blood pressure and office blood pressure measurement in dogs and cats
    Koo, Siu To
    Carr, Anthony P.
    CANADIAN JOURNAL OF VETERINARY RESEARCH-REVUE CANADIENNE DE RECHERCHE VETERINAIRE, 2022, 86 (03): : 203 - 208
  • [3] A comparison of office blood pressure, telemedical home blood pressure and ambulatory blood pressure monitoring
    Hoffmann-Petersen, N.
    Tambo, T.
    Pedersen, E. Bjerregaard
    EUROPEAN HEART JOURNAL, 2014, 35 : 573 - 573
  • [4] Home blood pressure in children and adolescents: a comparison with office and ambulatory blood pressure measurements
    Salgado, Claudia Maria
    Brandao Veiga Jardim, Paulo Cesar
    Brito Viana, Jackeline Karoline
    Veiga Jardim, Thiago de Souza
    Castillo Velasquez, Paola Patricia
    ACTA PAEDIATRICA, 2011, 100 (10) : E163 - E168
  • [5] Performance Of Home Blood Pressure Monitors In The Clinical Setting
    Vyavahare, Medha
    Cluett, Jennifer
    Wood, Julia
    Turkson-Ocran, Ruth-Alma
    Mukamal, Ken J.
    Juraschek, Stephen P.
    HYPERTENSION, 2022, 79
  • [6] Comparison of Office Blood Pressure, Automated Unattended Office Blood Pressure, Home Blood Pressure, and 24-Hour Ambulatory Blood Pressure Measurements
    Kim, Byung Jin
    Lee, Sung Ho
    Lee, Mi Yeon
    Lee, Seung Jae
    Choi, Hyo In
    JOURNAL OF KOREAN MEDICAL SCIENCE, 2023, 38 (48)
  • [7] The clinical significance of home and office blood pressure in diabetic nephropathy
    Suzuki, Daisuke
    Hoshide, Satoshi
    Kario, Kazuomi
    HYPERTENSION RESEARCH, 2023, 46 (05) : 1347 - 1349
  • [8] The clinical significance of home and office blood pressure in diabetic nephropathy
    Daisuke Suzuki
    Satoshi Hoshide
    Kazuomi Kario
    Hypertension Research, 2023, 46 : 1347 - 1349
  • [9] Clinical evaluation of wrist devices for home blood pressure monitoring in comparison to arm devices and ambulatory monitoring
    Stergiou, G.
    Christodoulakis, G.
    Kalogeropoulos, P.
    Giovas, P.
    Achimastos, A.
    JOURNAL OF HYPERTENSION, 2007, 25 : S165 - S165
  • [10] Comparison of Different Automated Office Blood Pressure Measurement Devices: Evidence of Nonequivalence and Clinical Implications
    Rinfret, Felix
    Cloutier, Lyne
    Wistaff, Robert
    Birnbaum, Leora M.
    Cheong, Nathalie Ng
    Laskine, Mikhael
    Roederer, Ghislaine
    Nguyen, Paul Van
    Bertrand, Michel
    Rabasa-Lhoret, Remi
    Dufour, Robert
    Lamarre-Cliche, Maxime
    CANADIAN JOURNAL OF CARDIOLOGY, 2017, 33 (12) : 1639 - 1644