Development of a Reference Device for the Calibration of Optical One-Shot Time-Interval Measurements

被引:0
|
作者
Kuhinek, Dalibor [1 ]
Bratko, Tomislav [1 ]
Vukosic, Hrvoje [1 ]
Gavranic, Ivica [2 ]
机构
[1] Univ Zagreb, Fac Min Geol & Petr Engn, Pierottijeva 6, Zagreb 10002, Croatia
[2] INEL MONTAZA Doo, Dane Duica 3, Zagreb 10000, Croatia
关键词
velocity of detonation measurement; electronic detonation speedometer; precision time delay; interval generation;
D O I
10.3390/electronics12020439
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
This paper presents the development of a device that creates two or more light pulses with a precision time delay in a range from microseconds to several seconds and for the purpose of the calibration of velocity of detonation (VOD) measurement instruments. The device was assembled, programmed, and tested for functionality. Measurements were conducted using a reference counter. First, a precision OCXO (oven-controlled crystal oscillator) was used as a clock source. For verification of programmed subroutines, a counter oscillator output port was used as a source of clock signal for the microcontroller. This enabled the cancellation of possible oscillator errors. The signal had to be converted from an AC sine signal to a signal with positive values only using a clamper circuit. By the proposed solution, a calculable standard of time delay between two light pulses was achieved. According to the obtained results, this device can be used to calibrate field measurement devices for VOD measurements in explosives or some other use where the measurement device records the time interval between multiple light pulses. This enables more confidence in measurement results, faster recognition of instrument drift and increases the quality of measurement.
引用
收藏
页数:14
相关论文
共 50 条
  • [31] Magnesium optical one-shot sensor based on a coumarin chromoionophore
    Capitán-Vallvey, LE
    Fernández-Ramos, MD
    Lapresta-Fernández, A
    Brunet, E
    Rodríguez-Ubis, JC
    Juanes, O
    TALANTA, 2006, 68 (05) : 1663 - 1670
  • [32] The construction of probabilistic reliability model with one-shot device testing data
    Chimitova, E. V.
    2014 12TH INTERNATIONAL CONFERENCE ON ACTUAL PROBLEMS OF ELECTRONICS INSTRUMENT ENGINEERING (APEIE), 2014,
  • [33] One-shot three-dimensional measurements with a fiber bundle using a chirped optical frequency comb
    Uchida, Megumi
    Kato, Takashi
    Tanaka, Yurina
    Minoshima, Kaoru
    2017 CONFERENCE ON LASERS AND ELECTRO-OPTICS PACIFIC RIM (CLEO-PR), 2017,
  • [34] EM algorithm for one-shot device testing under the exponential distribution
    Balakrishnan, N.
    Ling, M. H.
    COMPUTATIONAL STATISTICS & DATA ANALYSIS, 2012, 56 (03) : 502 - 509
  • [35] Development of ultraviolet- and visible-light one-shot spectral domain optical coherence tomography and in situ measurements of human skin
    Hirayama, Heijiro
    Nakamura, Sohichiro
    JOURNAL OF BIOMEDICAL OPTICS, 2015, 20 (07)
  • [36] One-Shot Calibration of RF Circuits Based on Non-Intrusive Sensors
    Andraud, Martin
    Stratigopoulos, Haralampos-G
    Simeu, Emmanuel
    2014 51ST ACM/EDAC/IEEE DESIGN AUTOMATION CONFERENCE (DAC), 2014,
  • [37] A Comparative Study of One-Shot Statistical Calibration Methods for Analog/RF ICs
    Lu, Yichuan
    Suhramani, Kiruha S.
    Huang, He
    Kupp, Nathan
    Huang, Ke
    Makris, Yiorgos
    2015 IEEE INTERNATIONAL TEST CONFERENCE (ITC), 2015,
  • [38] One-Shot Learning for Real-Time Action Recognition
    Fanello, Sean Ryan
    Gori, Ilaria
    Metta, Giorgio
    Odone, Francesca
    PATTERN RECOGNITION AND IMAGE ANALYSIS, IBPRIA 2013, 2013, 7887 : 31 - 40
  • [39] CLOCKED CMOS ONE-SHOT HAS NO RC TIME CONSTANT
    GRIESMYER, G
    EDN MAGAZINE-ELECTRICAL DESIGN NEWS, 1978, 23 (10): : 164 - &