Prospective Comparison Between Manual and Computer-Assisted (FOX) Cochlear Implant Fitting in Newly Implanted Patients

被引:4
|
作者
Wathour, Justine [1 ]
Govaerts, Paul J. [2 ,3 ]
Derue, Lucie [1 ]
Vanderbemden, Suzanne [1 ]
Huaux, Helene [1 ]
Lacroix, Emilie [4 ,5 ]
Deggouj, Naima [1 ]
机构
[1] Clin Univ St Luc, Brussels, Belgium
[2] Eargroup, Antwerp, Belgium
[3] Univ Antwerp, Fac Med & Hlth Sci, Translat Neurosci Otorhinolaryngol & Head & Neck S, Antwerp, Belgium
[4] Royal Mil Acad, LIFE Dept, VIPER Res Unit, Brussels, Belgium
[5] Catholic Univ Louvain, Inst Res Psychol Sci IPSY, Louvain La Neuve, Belgium
来源
EAR AND HEARING | 2023年 / 44卷 / 03期
关键词
Artificial intelligence; Cochlear implant; Newly users; Programming; NEURAL RESPONSE TELEMETRY; PROCESSOR;
D O I
10.1097/AUD.0000000000001314
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Objective: A prospective, longitudinal, randomized controlled trial with an original crossover design for 1 year was conducted to compare manual fitting to artificial intelligence-based fitting in newly implanted patients. Design: Twenty-four patients who received their first cochlear implant (CI) were randomly assigned to the manual or Fitting to Outcome eXpert (FOX) arm; they followed the corresponding fitting procedures for 1 year. After 1 year, each patient was switched to another arm. The number of fittings, auditory outcomes (pure-tone thresholds, loudness scaling curves, spectral discrimination scores, bisyllabic word recognition in quiet and noise, and speech tracking), fitting session questionnaire, and CI parameters (T level, C level, Threshold Sound Pressure Level (T-SPL), Comfortable Sound Pressure Level (C-SPL), and loudness growth value) were compared between the two groups. Differences between the two groups were analyzed using the Mann-Whitney test, and Holm corrections were applied for multiple statistical tests. At the end of the crossover session, patients were offered the choice to continue with their old or new map. Results: As early as 3 mo postactivation, the FOX group showed less variability and significantly better speech intelligibility in quiet conditions at 40 and 55 dB SPL and noise (p < 0.05) with median phoneme scores of 50%, 70%, and 50% at 55, 70, and 85 dB SPL compared with 45%, 50%, and 40%, respectively. This group showed better results at 12 mo postactivation (p < 0.05). In the manual group, 100% of the patients decided to keep the new FOX map, and 82% performed better with the FOX map. In the FOX group, 63% of the patients decided to keep the manual map, although the measurable outcome had not improved. In this group, participants reported to prefer the manual map because it felt more comfortable, even if the FOX map gave better measured outcome. Conclusion: Although the study size remains relatively small, the AI-FOX approach was equivalent to or even outperformed the manual approach in hearing performance, comfort, and resources. Furthermore, FOX is a tool capable of continuous improvement by comparing its predictions with observed results and is continuously learning from clinicians' practice, which is why this technology promises major advances in the future.
引用
收藏
页码:494 / 505
页数:12
相关论文
共 50 条
  • [1] Setting and Reaching Targets with Computer-Assisted Cochlear Implant Fitting
    Vaerenberg, Bart
    De Ceulaer, Geert
    Szlavik, Zoltan
    Mancini, Patrizia
    Buechner, Andreas
    Govaerts, Paul J.
    SCIENTIFIC WORLD JOURNAL, 2014,
  • [2] Investigation of an Outcomes-Driven, Computer-Assisted Approach to CI Fitting in Newly Implanted Patients
    Zwolan, Teresa A.
    Presley, Regina
    Chenier, Laura
    Buck, Barbara
    EAR AND HEARING, 2021, 42 (03): : 558 - 564
  • [3] Comparison between manual and computer-assisted measurements of hallux valgas parameters
    Panchbhavi, VK
    Trevino, S
    FOOT & ANKLE INTERNATIONAL, 2004, 25 (10) : 708 - 711
  • [4] Research on Computer-assisted Minimally Invasive Cochlear Implant System
    Zhu, Yunfeng
    Xu, Ying
    Ke, Jia
    Ma, Furong
    Hu, Lei
    Li, Changsheng
    APPLIED DECISIONS IN AREA OF MECHANICAL ENGINEERING AND INDUSTRIAL MANUFACTURING, 2014, 577 : 1241 - +
  • [5] Radiographic Angles in Hallux Valgus: Comparison between Manual and Computer-Assisted Measurements
    Srivastava, Subodh
    Chockalingam, Nachiappan
    El Fakhri, Tarek
    JOURNAL OF FOOT & ANKLE SURGERY, 2010, 49 (06): : 523 - 528
  • [6] Comparison of computer-assisted and manual screening of cervical cytology
    Lozano, Richard
    GYNECOLOGIC ONCOLOGY, 2007, 104 (01) : 134 - 138
  • [7] Preliminary evaluation of computer-assisted home training for French cochlear implant recipients
    Kerneis, Sandrine
    Galvin III, John J.
    Borel, Stephanie
    Baque, Jean
    Fu, Qian-Jie
    Bakhos, David
    PLOS ONE, 2023, 18 (04):
  • [8] Comparison between visual and computer-assisted analysis
    Bhargava, V
    Bhalla, MA
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2003, 19 (04): : 267 - 269
  • [9] Comparison between visual and computer-assisted analysis
    Valmik Bhargava
    Meenakshi Awasthi Bhalla
    The International Journal of Cardiovascular Imaging, 2003, 19 (4) : 267 - 269
  • [10] Open-wedge high tibial osteotomy: comparison between manual and computer-assisted techniques
    Iorio, R.
    Pagnottelli, M.
    Vadala, A.
    Giannetti, S.
    Di Sette, P.
    Papandrea, P.
    Conteduca, F.
    Ferretti, A.
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2013, 21 (01) : 113 - 119