Accuracy to detect neonatal hyperglycaemia using real-time continuous glucose monitoring during postoperative period

被引:0
|
作者
Wasiq, Mohammed Abdul [1 ]
Behura, Sushree Smita [1 ]
Sahoo, Swaranjika [1 ]
Panda, Santosh Kumar [1 ]
机构
[1] Kalinga Inst Med Sci, Dept Pediat, Bhubaneswar 751024, Orissa, India
关键词
Blood glucose; Continuous glucose monitoring; Hyperglycaemia; Neonates; CONSENSUS;
D O I
10.1007/s00431-022-04777-1
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Identifying hyperglycaemia during postoperative period is essential for neonates. The objective of the study was to analyse the accuracy and reliability of continuous glucose monitoring (CGM) device for detecting hyperglycaemia during postoperative period in neonates. In this prospective study, hourly glucose recordings by CGM device and six hourly by glucometer glucose (GG-reference test for patient management) were collected in ten surgical neonates during first three postoperative days. Mean absolute relative difference (MARD) and proportion of CGM values within +/- 15%/15mg/dL, +/- 20%/20 mg/dL, and +/- 30%/30 mg/dL of GG, were analysed from matched pair CGM and GG recordings. The diagnostic performance of CGM for neonatal hyperglycaemia (> 150 mg/dL) was expressed as sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV); and the clinical reliability was assessed in Clarke Error Grid Analysis. A total of 720 CGM glucose readings, 120 matched-paired glucosereadings by CGM --GG, and 37 episodes were hyperglycaemia by GG. The MARD was 10.76 %; the per-centages of glucose readings within 15%/15 mg/dL, 20%/20 mg/dL, and 30%/30 mg/dL were 94.6%, 97.3% and 100% during the hyperglycaemia period. The sensitivity, specificity, PPV and NPV to detect hyperglycaaemia by CGM device were 100%, 93.9%, 88% and 100 % respectively. In Clarke Error Grid Analysis, 97.3 % points were in zone A and B during the hyperglycaemia period. Conclusion: CGM device can be a clinically reliable tool for hyperglycaemia management during postoperative period in neonates.
引用
收藏
页码:1083 / 1087
页数:5
相关论文
共 50 条
  • [1] Accuracy to detect neonatal hyperglycaemia using real-time continuous glucose monitoring during postoperative period
    Mohammed Abdul Wasiq
    Sushree Smita Behura
    Swaranjika Sahoo
    Santosh Kumar Panda
    [J]. European Journal of Pediatrics, 2023, 182 : 1083 - 1087
  • [2] A Feasibility Study to Detect Neonatal Hypoglycemia Using Real-Time Continuous Glucose Monitoring
    Nally, Laura M.
    Bondy, Nicholas W.
    Doiev, Jasmine
    Buckingham, Bruce
    Wilson, Darrell
    [J]. DIABETES, 2018, 67
  • [3] Real-Time Improvement of Continuous Glucose Monitoring Accuracy
    Facchinetti, Andrea
    Sparacino, Giovanni
    Guerra, Stefania
    Luijf, Yoeri M.
    DeVries, J. Hans
    Mader, Julia K.
    Ellmerer, Martin
    Benesch, Carsten
    Heinemann, Lutz
    Bruttomesso, Daniela
    Avogaro, Angelo
    Cobelli, Claudio
    [J]. DIABETES CARE, 2013, 36 (04) : 793 - 800
  • [4] Accuracy of real-time glucose values using continuous glucose monitoring:: the Guardian™ II continuous glucose monitoring system
    Schwartz, SL
    Buchea, O
    Shin, J
    Hoss, U
    Shah, R
    [J]. DIABETOLOGIA, 2004, 47 : A315 - A315
  • [5] A Feasibility Study to Detect Neonatal Hypoglycemia in Infants of Diabetic Mothers Using Real-Time Continuous Glucose Monitoring
    Nally, Laura Marie
    Bondy, Nicholas
    Doiev, Jasmine
    Buckingham, Bruce A.
    Wilson, Darrell M.
    [J]. DIABETES TECHNOLOGY & THERAPEUTICS, 2019, 21 (04) : 170 - 176
  • [6] Accuracy of a new real-time continuous glucose monitoring algorithm
    Navas de Solis, M. S.
    [J]. AVANCES EN DIABETOLOGIA, 2010, 26 (02): : 129 - 130
  • [7] Accuracy of real-time continuous glucose monitoring in the MiniMed paradigm system
    Mastrototaro, John
    Soundararajan, Gopikrishnan
    Cooper, Ken
    Shah, Rajiv
    [J]. DIABETES, 2007, 56 : A112 - A112
  • [8] Assessment of glucose kinetics with real-time continuous glucose monitoring during labor
    Maki, Jota
    Eto, Eriko
    Tamada, Shoko
    Mitsui, Takashi
    Hayata, Kei
    Nakamura, Keiichiro
    Hiramatsu, Yuji
    Masuyama, Hisashi
    [J]. JOURNAL OF OBSTETRICS AND GYNAECOLOGY RESEARCH, 2019, 45 (09) : 1851 - 1859
  • [9] Accuracy Evaluation of a New Real-Time Continuous Glucose Monitoring Algorithm in Hypoglycemia
    Mahmoudi, Zeinab
    Jensen, Morten Hasselstrom
    Johansen, Mette Dencker
    Christensen, Toke Folke
    Tarnow, Lise
    Christiansen, Jens Sandahl
    Hejlesen, Ole
    [J]. DIABETES TECHNOLOGY & THERAPEUTICS, 2014, 16 (10) : 667 - 678
  • [10] Accuracy of a real-time continuous glucose monitoring system during cardiac surgery with hypothermic extracorporeal circulation
    Herzig, D.
    Vettoretti, M.
    Guensch, D. P.
    Melmer, A.
    Goerg, A. C. K.
    Kadner, A.
    Facchinetti, A.
    Vogt, A.
    Bally, L.
    [J]. DIABETOLOGIA, 2022, 65 (SUPPL 1) : S28 - S28