Optimized Camera-Based Patient Positioning in CT Impact on Radiation Exposure

被引:8
|
作者
Manava, Panagiota [1 ,2 ]
Galster, Marco [1 ]
Ammon, Josefin [3 ]
Singer, Julian [3 ]
Lell, Michael M. [1 ,2 ]
Rieger, Vera [1 ]
机构
[1] Paracelsus Med Univ, Dept Radiol & Nucl Med, Klinikum Nuemberg, Prof Ernst Nathan St 1, D-90419 Nurnberg, Germany
[2] Friedrich Alexander Univ Erlangen Nuremberg, Dept Radiol, Erlangen, Germany
[3] Paracelsus Med Univ, Inst Med Phys, Nurnberg, Germany
关键词
patient-centering; radiation dosage; diagnostic imaging; TUBE CURRENT MODULATION; COMPUTED-TOMOGRAPHY; IMAGE NOISE; VOLTAGE SELECTION; ORGAN; SYSTEMS; PHANTOM;
D O I
10.1097/RLI.0000000000000904
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveThe aim of this study was to evaluate whether a 3-dimensional (3D) camera can outperform highly trained technicians in precision of patient positioning and whether this transforms into a reduction in patient exposure.Materials and MethodsIn a single-center study, 3118 patients underwent computer tomography (CT) scans of the chest and/or abdomen on a latest generation single-source CT scanner supported with an automated patient positioning system by 3D camera. One thousand five hundred fifty-seven patients were positioned laser-guided by a highly trained radiographer (camera off) and 1561 patients with 3D camera (camera on) guidance. Radiation parameters such as effective dose, organ doses, CT dose index, and dose length product were analyzed and compared. Isocenter accuracy and table height were evaluated between the 2 groups.ResultsIsocenter positioning was significantly improved with the 3D camera (P < 0.001) as compared with visual laser-guided positioning. Absolute table height differed significantly (P < 0.001), being higher with camera positioning (165.6 +/- 16.2 mm) as compared with laser-guided positioning (170.0 +/- 20.4 mm). Radiation exposure decreased using the 3D camera as indicated by dose length product (321.1 +/- 266.6 mGy center dot cm; camera off: 342.0 +/- 280.7 mGy center dot cm; P = 0.033), effective dose (3.3 +/- 2.7 mSv; camera off: 3.5 +/- 2.9; P = 0.053), and CT dose index (6.4 +/- 4.3 mGy; camera off: 6.8 +/- 4.6 mGy; P = 0.011). Exposure of radiation-sensitive organs such as colon (P = 0.015) and red bone marrow (P = 0.049) were also lower using the camera.ConclusionsThe introduction of a 3D camera improves patient positioning in the isocenter of the scanner, which results in a lower and also better balanced dose reduction for the patients.
引用
收藏
页码:126 / 130
页数:5
相关论文
共 50 条
  • [41] Automatic registration of portal images and volumetric CT for patient positioning in radiation therapy
    Khamene, A
    Bloch, P
    Wein, W
    Svatos, M
    Sauer, F
    MEDICAL IMAGE ANALYSIS, 2006, 10 (01) : 96 - 112
  • [42] The impact of patient positioning and preparation with advanced radiation therapy treatment techniques
    Wallis, A.
    RADIOTHERAPY AND ONCOLOGY, 2019, 141 : S28 - S28
  • [43] The clinical impact of camera-based positron emission tomography imaging in patients with recurrent colorectal cancer
    Miller, E
    Lerman, H
    Gutman, M
    Figer, A
    Livshitz, G
    Even-Sapir, E
    INVESTIGATIVE RADIOLOGY, 2004, 39 (01) : 8 - 12
  • [44] Framework for Estimating the Impact of Camera-Based Intelligent Transportation Systems (ITS) Technology on Incident Duration
    Moylan, Emily
    Chand, Sai
    Waller, S. Travis
    TRANSPORTATION RESEARCH RECORD, 2018, 2672 (19) : 25 - 33
  • [45] Use of a 3D camera for automated patient positioning for chest-abdomen-pelvis CT scans: Effect on positioning accuracy and patient dose
    Nelson, R.
    Harkin, S.
    Iball, G.
    RADIOGRAPHY, 2024, 30 (04) : 1060 - 1067
  • [46] Contribution of CT scan to patient's radiation exposure in parathyroid SPECT/CT scintigraphy
    Rep, S.
    Jensterle, L.
    Zdesar, U.
    Zaletel, K.
    Tomse, P.
    Lezaic, L.
    RADIOGRAPHY, 2024, 30 (03) : 995 - 1000
  • [47] Accuracy of automated patient positioning in CT using a 3D camera for body contour detection
    Booij, Ronald
    Budde, Ricardo P. J.
    Dijkshoorn, Marcel L.
    van Straten, Marcel
    EUROPEAN RADIOLOGY, 2019, 29 (04) : 2079 - 2088
  • [48] Accuracy of automated patient positioning in CT using a 3D camera for body contour detection
    Ronald Booij
    Ricardo P.J. Budde
    Marcel L. Dijkshoorn
    Marcel van Straten
    European Radiology, 2019, 29 : 2079 - 2088
  • [49] The impact of clinical audits in patient radiation exposure: a preliminary study
    Samara, Eleni
    Sazgary, Lorraine
    Stussi, Anja
    Guckenberger, Matthias
    Saguner, Ardan
    SWISS MEDICAL WEEKLY, 2023, 153 : 89S - 89S
  • [50] THE IMPACT OF PATIENT POSITIONING ON RADIATION DOSE TO SMALL BOWEL IN PATIENTS WITH RECTAL CANCER
    Froseth, T. C.
    Strickert, T.
    Frykholm, G.
    Reidunsdatter, R. J.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S428 - S428