The Influence of Design on Stress Concentration Reduction in Dental Implant Systems Using the Finite Element Method

被引:2
|
作者
Pala, Eser [1 ]
Ozdemir, Ismail [2 ]
Grund, Thomas [2 ]
Lampke, Thomas [2 ]
机构
[1] MarlaDT Engn, Adalet Mahallesi Folkart Towers A Kule 47-B, TR-35530 Izmir, Turkiye
[2] Tech Univ Chemnitz, Mat & Surface Engn Grp, Erfenschlager Str 73, D-09125 Chemnitz, Germany
关键词
FEM analysis; stress distribution; Ti-6Al-4V alloy; dental implant; design;
D O I
10.3390/cryst14010020
中图分类号
O7 [晶体学];
学科分类号
0702 ; 070205 ; 0703 ; 080501 ;
摘要
Dental implant fracture is closely connected to the stress buildup surrounding the implant system during static loading. In areas where the cross-section of the implant rapidly changes or where the geometry of the implant system has discontinuities, stress concentrations arise. Therefore, the implant's design is crucial in preventing early failure of the implant system, including fracture, screw loosening, and increased leakage, in addition to reducing stresses at the implant-abutment interface. In the current work, three-dimensional (3D) models of mechanically connected Ti6-Al-4V implant systems in various dimensions were constructed. Finite element analysis (FEA) was used to conduct a stress study of the created implants under actual acting force static loading conditions in accordance with ISO 14801. In the created models, design elements including implant screw type, thickness, and taper angle of abutment were modified in order to increase the longevity of the implants. The results show that the equivalent stress level was dramatically reduced from 596.22 MPa to 212.72 MPa in the implant model, which exhibits a more homogeneous stress pattern under static loading conditions. By increasing the implant wall thickness from 0.15 mm to 0.40 mm in the region adjacent to the abutment, the stress levels, especially at the internal screw, were significantly reduced. Also, the design modification in Model B, establishing contact between the abutment and the upper part of the conical surface of the implant, resulted in a decrease in stress in the internal screw. Thus, enhanced homogeneity in stress distribution not only improves the harmony between the implant and surrounding tissues, thus increasing patient comfort and reducing the risk of complications, but also holds promise for the development of new implants capable of withstanding the forces encountered in the oral environment due to the relatively smoother stress transmission observed in this model.
引用
收藏
页数:12
相关论文
共 50 条
  • [21] Design of patient specific basal dental implant using Finite Element method and Artificial Neural Network technique
    Choudhury, Sandeep
    Rana, Masud
    Chakraborty, Arindam
    Majumder, Santanu
    Roy, Sandipan
    RoyChowdhury, Amit
    Datta, Shubhabrata
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2022, 236 (09) : 1375 - 1387
  • [22] Design and Analysis of Customized Hip Implant Using Finite Element Method
    Priya, M. Anu
    Snekhalatha, U.
    Mahalakshmi, R.
    Dhivya, T.
    Gupta, Nilkantha
    ARTIFICIAL INTELLIGENCE AND EVOLUTIONARY COMPUTATIONS IN ENGINEERING SYSTEMS, 2020, 1056 : 515 - 525
  • [23] Investigating a Novel Design Dental Implant by Using Finite Element Analysis and Experimental Setup
    Sen, Yasar
    JOURNAL OF BIOMATERIALS AND TISSUE ENGINEERING, 2020, 10 (07) : 915 - 921
  • [24] Finite Element Analysis of Stress Distribution in Immediately Loaded Dental Implant
    Bahrami, Babak
    Ghalichi, Farzan
    Mirzakouchaki, Behnam
    Ashtiani, Mohammed N.
    Marghoub, Arsalan
    2012 19TH IRANIAN CONFERENCE OF BIOMEDICAL ENGINEERING (ICBME), 2012, : 154 - 157
  • [25] A finite element analysis of the stress distribution of an osseointegrated dental implant.
    O'Mahony, A
    Bowles, O
    Spencer, P
    Woolsey, G
    JOURNAL OF DENTAL RESEARCH, 1998, 77 : 121 - 121
  • [26] Application of the finite element method in orthopedic implant design
    Saha, Subrata
    Roychowdhury, Amit
    Journal of Long-Term Effects of Medical Implants, 2009, 19 (01) : 55 - 82
  • [27] Stress Reduction of Pickup Truck Chassis Using Finite Element Method
    Kurdi, O.
    Yob, M. S.
    Dasson, S. R.
    Barrathi, S.
    Altayeb, A. A.
    Yulianti, I.
    3RD INTERNATIONAL CONFERENCE ON MATHEMATICS, SCIENCE AND EDUCATION 2016, 2017, 824
  • [28] Preliminary evaluation of a new dental implant design with finite element analysis
    Wang, XJ
    Wang, R
    Luo, JM
    Chen, JY
    Zhang, XD
    ASBM6: ADVANCED BIOMATERIALS VI, 2005, 288-289 : 657 - 660
  • [29] Stress distribution of two commercial dental implant systems: A three-dimensional finite element analysis
    Chang, Hao-Sheng
    Chen, Yi-Chin
    Hsieh, Yao-Dung
    Hsu, Ming-Lun
    JOURNAL OF DENTAL SCIENCES, 2013, 8 (03) : 261 - 271
  • [30] Assessment of Stress Shielding around a Dental Implant for Variation of Implant Stiffness and Parafunctional Loading using Finite Element Analysis
    Ishak, Muhammad Ikman
    Daud, Ruslizam
    Noor, Siti Noor Fazliah Mohd
    Khor, C. Y.
    Roslan, Husniyati
    ACTA OF BIOENGINEERING AND BIOMECHANICS, 2022, 24 (03) : 147 - 159