Calibration of 10 nF capacitance standard from DC quantum Hall resistance using a digital impedance bridge

被引:4
|
作者
Tran, Ngoc Thanh Mai [1 ]
Kucera, Jan [2 ]
Kim, Wan-Seop [3 ]
Kim, Dan Bee [3 ]
机构
[1] Ist Nazl Ric Metrolog, Str Cacce 91, I-10135 Turin, Italy
[2] Czech Metrol Inst, Okruzni 31, CZ-63800 Brno, Czech Republic
[3] Korea Res Inst Stand & Sci, 267 Gajeongno, Daejeon 34113, South Korea
关键词
quantum Hall resistance; capacitance standard; resistance standard; digital impedance bridge; direct comparison; FARAD; REALIZATION;
D O I
10.1088/1361-6501/acc6e2
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
This paper describes the calibration of a 10 nF capacitance standard traced to the DC quantum Hall resistance primary standard using a multi-purpose digital impedance bridge. The bridge can directly compare the unlike impedances in the fully digital (FD) mode, so the capacitance standard can be calibrated by a direct comparison against the AC resistance standard. The measurement ability of the bridge in the FD mode was evaluated in various angles, and the bridge voltage dependencies were improved by selecting appropriate impedances for the high current arms and also by using an additional injection circuit. Thereby, the measurement accuracy could reach up to 10(-7) in relative terms. Based on the bridge evaluation, a 10 nF capacitance standard was calibrated against a 10 k omega resistance standard, and the measurement value agreed with a reference value in the level of 10(-7).
引用
收藏
页数:6
相关论文
共 25 条
  • [21] A 1 pF-to-10 nF Generic Capacitance-to-Digital Converter Using Zero-Crossing ΔΣ Modulation
    Li, Bing
    Wang, Wei
    Liu, Jia
    Liu, Wen-Jun
    Yang, Qian
    Ye, Wen-Bin
    IEEE TRANSACTIONS ON CIRCUITS AND SYSTEMS I-REGULAR PAPERS, 2018, 65 (07) : 2169 - 2182
  • [22] Automated guarded bridge for calibration of multimegohm standard resistors from 10 M Omega to 1 T Omega
    Jarrett, DG
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1997, 46 (02) : 325 - 328
  • [23] RESISTANCE RATIO BRIDGE USING CRYOGENIC CURRENT COMPARATOR WITH DC-SUPERCONDUCTING QUANTUM INTERFERENCE DEVICE MAGNETOMETER
    NAKANISHI, M
    KINOSHITA, J
    ENDO, T
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 1995, 34 (6A): : 3284 - 3289
  • [24] Comparison of capacitance with resistance by IVD-based quadrature bridge at frequencies from 50Hz to 10kHz
    Hsu, JC
    Ku, YS
    2000 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2000, : 429 - 430
  • [25] A Capacitive Biosensor for Cancer Diagnosis Using a Functionalized Microneedle and a 13.7b-Resolution Capacitance-to-Digital Converter from 1 to 100nF
    Song, Seungwoo
    Na, Jukwan
    Jang, Moonhyung
    Lee, Hyeyeon
    Lee, Hyesoo
    Lim, Yongbeom
    Choi, Heonjin
    Chae, Youngcheol
    2019 IEEE INTERNATIONAL SOLID-STATE CIRCUITS CONFERENCE (ISSCC), 2019, 62 : 194 - +