Endoscopic ultrasound-guided choledochoduodenostomies with fully covered self-expandable metallic stents

被引:1
|
作者
Tae Jun Song [1 ]
Yil Sik Hyun [2 ]
Sang Soo Lee [2 ]
Do Hyun Park [2 ]
Dong Wan Seo [2 ]
Sung Koo Lee [2 ]
Myung-Hwan Kim [2 ]
机构
[1] Department of Internal Medicine,Inje University Ilsan Paik Hospital,Koyang 411-706,South Korea
[2] Department of Internal Medicine,University of Ulsan College of Medicine,Asan Medical Center,Seoul 138-736,South Korea
关键词
Bile duct obstruction; Drainage; Endosonography; Self-expandable metallic stent; Neoplasms;
D O I
暂无
中图分类号
R445.1 [超声波诊断];
学科分类号
100207 ;
摘要
AIM:To investigate the long-term outcomes of endoscopic ultrasound-guided choledochoduodenostomy(EUS-CDS) with a fully covered self-expandable metallic stent(FCSEMS).METHODS:From April 2009 to August 2010,15 patients with distal malignant biliary obstructions who were candidates for alternative techniques for biliary decompression due to a failed endoscopic retrograde cholangiopancreatography(ERCP) were included.These 15 patients consisted of 8 men and 7 women and had a median age of 61 years(range:30-91 years).The underlying causes of the distal malignant biliary obstruction were pancreatic cancer(n = 9),ampulla of Vater cancer(n = 2),renal cell carcinoma(n = 1),advanced gastric cancer(n = 1),lymphoma(n = 1),and duodenal cancer(n = 1).RESULTS:The technical success rate of EUS-CDS with an FCSEMS was 86.7%(13/15),and functional success was achieved in 100%(13/13) of those cases.In two patients,the EUS-CDS failed because an FCSEMS with a delivery device could not be passed into the common bile duct.The mean duration of stent patency was 264 d.Early adverse events developed in three patients(3/13,23.1%),including self-limited pneumoperitoneum in two patients and cholangitis requiring stent reposition in one patient.During the follow-up period(median:186 d,range:52-388 d),distal stent migration occurred in four patients(4/13,30.8%).In 3 patients,the FCSEMS could be reinserted through the existing choledochoduodenal fistula tract.CONCLUSION:EUS-CDS with an FCSEMS is technically feasible and can lead to effective palliation of distal malignant biliary obstructions after failed ERCP.
引用
收藏
页码:4435 / 4440
页数:6
相关论文
共 50 条
  • [11] Chronic pancreatic pain successfully treated by endoscopic ultrasound-guided pancreaticogastrostomy using fully covered self-expandable metallic stent
    Chang, Arunchai
    Aswakul, Pitulak
    Prachayakul, Varayu
    WORLD JOURNAL OF CLINICAL CASES, 2016, 4 (04) : 112 - 117
  • [12] FULLY COVERED SELF-EXPANDABLE METALLIC STENTS FOR MALIGNANT AIRWAY DISORDERS
    Oki, Masahide
    Saka, Hideo
    Kogure, Yoshihito
    Oka, Saori
    Nakahata, Masashi
    Ise, Yuko
    Niwa, Hideyuki
    Ishida, Akane
    Shigematsu, Fumie
    Yamada, Arisa
    Torii, Atsushi
    Kitagawa, Chiyoe
    RESPIROLOGY, 2017, 22 : 242 - 243
  • [13] Fully covered self-expandable metallic stents for malignant airway disorders
    Ishida, Akane
    Oki, Masahide
    Saka, Hideo
    RESPIRATORY INVESTIGATION, 2019, 57 (01) : 49 - 53
  • [14] Transrectal endoscopic ultrasound-guided drainage of pelvic abscess with placement of a fully covered self-expandable metal stent
    Luigiano, C.
    Togliani, T.
    Cennamo, V.
    Maimone, A.
    Polifemo, A. M.
    Pilati, S.
    Fabbri, C.
    ENDOSCOPY, 2013, 45 : E245 - E246
  • [15] Endoscopic ultrasound-guided gastrojejunostomy with a direct technique without previous intestinal filling using a tubular fully covered self-expandable metallic stent
    Senturk, Hakan
    Koker, Ibrahim Hakki
    Kochan, Koray
    Kiremitci, Sercan
    Seven, Gulseren
    Ince, Ali Tuzun
    CLINICAL ENDOSCOPY, 2024, 57 (02) : 209 - 216
  • [16] Troubleshooting bile leakage after endoscopic ultrasound-guided hepaticogastrostomy with fully covered self-expandable metal stent deployment
    Kuraoka, Naosuke
    Ujihara, Tetsuro
    Sakai, Shun
    JOURNAL OF HEPATO-BILIARY-PANCREATIC SCIENCES, 2024, 31 (09) : e51 - e52
  • [17] Clinical Outcome of Endoscopic Ultrasound-Guided Liver Abscess Drainage Using Self-Expandable Covered Metallic Stent (with Video)
    Takeshi Ogura
    Daisuke Masuda
    Onda Saori
    Takagi Wataru
    Tatsushi Sano
    Atsushi Okuda
    Akira Miyano
    Masayuki Kitano
    Usama M. Abdel-aal
    Toshihisa Takeuchi
    Shinya Fukunishi
    Kazuhide Higuchi
    Digestive Diseases and Sciences, 2016, 61 : 303 - 308
  • [18] Clinical Outcome of Endoscopic Ultrasound-Guided Liver Abscess Drainage Using Self-Expandable Covered Metallic Stent (with Video)
    Ogura, Takeshi
    Masuda, Daisuke
    Saori, Onda
    Wataru, Takagi
    Sano, Tatsushi
    Okuda, Atsushi
    Miyano, Akira
    Kitano, Masayuki
    Abdel-aal, Usama M.
    Takeuchi, Toshihisa
    Fukunishi, Shinya
    Higuchi, Kazuhide
    DIGESTIVE DISEASES AND SCIENCES, 2016, 61 (01) : 303 - 308
  • [19] Endoscopic ultrasound-guided drainage of a right liver abscess with a self-expandable metallic stent
    Ogura, Takeshi
    Takagi, Wataru
    Onda, Saori
    Masuda, Daisuke
    Kitano, Masayuki
    Imoto, Akira
    Higuchi, Kazuhide
    ENDOSCOPY, 2015, 47 : E397 - E398
  • [20] Fully Covered Self-Expandable Metal Stents Performance
    Cubero, Noelia
    Carnevali, Leopoldo
    Lisbona, Rosa Lopez
    Diez, Marta
    Morales, Arturo
    Dorca, Jordi
    Rosell, Antoni