Determination of torasemide in human plasma and its bioequivalence study by high-performance liquid chromatography with electrospray ionization tandem mass spectrometry

被引:0
|
作者
Lin Zhang [1 ]
Rulin Wang [1 ]
Yuan Tian [1 ]
Zunjian Zhang [1 ,2 ]
机构
[1] Key Laboratory of Drug Quality Control and Pharmacovigilance(Ministry of Education),China Pharmaceutical University
[2] State Key Laboratory of Natural Medicine, China Pharmaceutical University
关键词
Torasemide; HPLC–ESI–MS; Human plasma; Bioequivalence; Pharmacokinetics;
D O I
暂无
中图分类号
R927.2 [药物含量测定]; O657.63 [质谱分析];
学科分类号
070302 ; 081704 ; 1007 ;
摘要
A sensitive and selective method using high-performance liquid chromatography coupled with electrospray ionization tandem mass spectrometry(HPLC–ESI–MS) to determine the concentration of torasemide in human plasma samples was developed and validated. Tolbutamide was chosen as the internal standard(IS). The chromatography was performed on a Gl Sciences Inertsil ODS-3 column(100 mm*2.1 mm i.d., 5.0μm) within 5 min, using methanol with 10 mM ammonium formate(60:40, v/v) as mobile phase at a fl ow rate of 0.2 mL/min. The targeted compound was detected in negative ionization at m/z 347.00 for torasemide and 269.00 for IS. The linearity range of this method was found to be within the concentration range of 1–2500 ng/mL(r=0.9984) for torasemide in human plasma. The accuracy of this measurement was between 94.05% and 103.86%. The extracted recovery ef fi ciency was from 84.20% to 86.47% at three concentration levels. This method was also successfully applied in pharmacokinetics and bioequivalence studies in Chinese volunteers.
引用
收藏
页码:95 / 102
页数:8
相关论文
共 50 条
  • [31] Quantitative determination of Ecteinascidin 743 in human plasma by miniaturized high-performance liquid chromatography coupled with electrospray ionization tandem mass spectrometry
    Rosing, H
    Hillebrand, MJX
    Jimeno, JM
    Gómez, A
    Floriano, P
    Faircloth, G
    Henrar, REC
    Vermorken, JB
    Cvitkovic, E
    Bult, A
    Beijnen, JH
    [J]. JOURNAL OF MASS SPECTROMETRY, 1998, 33 (11): : 1134 - 1140
  • [32] Determination of levocetirizine in human plasma by liquid chromatography-electrospray tandem mass spectrometry: Application to a bioequivalence study
    Morita, M. R.
    Berton, D.
    Boldin, R.
    Barros, F. A. P.
    Meurer, E. C.
    Amarante, A. R.
    Campos, D. R.
    Calafatti, S. A.
    Pereira, R.
    Abib, E., Jr.
    Pedrazolli, J., Jr.
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2008, 862 (1-2): : 132 - 139
  • [33] Determination of memantine in human plasma by liquid chromatography-electrospray tandem mass spectrometry: Application to a bioequivalence study
    Almeida, A. A.
    Campos, D. R.
    Bernasconi, G.
    Calafatti, S.
    Barros, F. A. P.
    Eberlin, M. N.
    Meurer, E. C.
    Paris, E. G.
    Pedrazzoli, J.
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2007, 848 (02): : 311 - 316
  • [34] High-performance liquid chromatography-tandem electrospray mass spectrometry for the determination of lidocaine and its metabolites in human plasma and urine
    Abdel-Rehim, M
    Bielenstein, M
    Askemark, Y
    Tyrefors, N
    Arvidsson, T
    [J]. JOURNAL OF CHROMATOGRAPHY B, 2000, 741 (02): : 175 - 188
  • [35] Determination of clindamycin in human plasma by liquid chromatography electrospray tandem mass spectrometry: application to the bioequivalence study of clindamycin
    Yu, LL
    Chao, CK
    Liao, WJ
    Twu, TY
    Liu, CM
    Yang, TH
    Lin, ET
    [J]. JOURNAL OF CHROMATOGRAPHY B, 1999, 724 (02): : 287 - 294
  • [36] Determination of nalmefene by high-performance liquid chromatography-electrospray ionization-tandem mass spectrometry
    Fang, WB
    Andrenyak, DM
    Moody, DE
    Nuwayser, ES
    [J]. JOURNAL OF ANALYTICAL TOXICOLOGY, 2005, 29 (03) : 169 - 174
  • [37] High-performance liquid chromatography-electrospray ionization mass spectrometry determination of sodium ferulate in human plasma
    Yang, Cheng
    Tian, Yuan
    Zhang, Zunjian
    Xu, Fengguo
    Chen, Yun
    [J]. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2007, 43 (03) : 945 - 950
  • [38] Rapid determination of coumatetralyl in human serum by high-performance liquid chromatography coupled with electrospray ionization tandem mass spectrometry
    Jin, Mi-cong
    OuYang, Xiao-kun
    Xu, Xiao-min
    Ren, Yi-ping
    Chen, Xiao-hong
    [J]. ANALYTICAL LETTERS, 2007, 40 (04) : 737 - 746
  • [39] Azithromycin quantitation in human plasma by high-performance liquid chromatography coupled to electrospray mass spectrometry: Application to bioequivalence study
    Xu, Fengguo
    Zhang, Zunjian
    Bian, Zuwei
    Tian, Yuan
    Jiao, Haoyan
    Liu, Ying
    [J]. JOURNAL OF CHROMATOGRAPHIC SCIENCE, 2008, 46 (06) : 479 - 484
  • [40] A bioequivalence study of gliclazide based on quantification by high-performance liquid chromatography coupled with electrospray tandem mass spectrometry
    Mendes, G. D.
    Moreira, L. D.
    Pereira, A. dos S.
    Borges, A.
    Yui, F.
    Mendes, F. D.
    de Nucci, G.
    [J]. INTERNATIONAL JOURNAL OF CLINICAL PHARMACOLOGY AND THERAPEUTICS, 2007, 45 (03) : 175 - 185