Fabrication and evaluation of breast tissue equivalent phantoms for image quality assessment in ultrasound imaging

被引:0
|
作者
Hariyanto, A. P. [1 ]
Mugni, F. F. [1 ]
Khumaira, L. [1 ]
Sensusiati, A. D. [2 ]
Nursela, A. L. [3 ]
Suprijanto [4 ]
Ng, K. H. [5 ]
Haryanto, F. [6 ]
Endarko [1 ]
机构
[1] Inst Teknol Sepuluh Nopember, Dept Phys, Kampus ITS, Sukolilo Surabaya 60111, East Java, Indonesia
[2] Univ Airlangga Hosp, Dept Radiol, Surabaya 60115, East Java, Indonesia
[3] Gambiran Gen Hosp, Radiol Installat, Kediri 64133, East Java, Indonesia
[4] Inst Teknol Bandung, Fac Ind Technol, Instrumentat & Control Res Grp, Jl Ganesha 10, Labtek 40132, Bandung, Indonesia
[5] Univ Malaya, Fac Med, Dept Biomed Imaging, Kuala Lumpur 50603, Malaysia
[6] Inst Teknol Bandung, Fac Math & Nat Sci, Dept Phys, Jl Ganesa 10, Bandung 40132, Indonesia
关键词
Phantom; Tissue equivalent; Elasticity; Acoustic properties; Ultrasound; LESION DETECTABILITY; MIMICKING;
D O I
10.1016/j.radi.2024.11.023
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: Phantom materials with tissue-equivalent physical properties that require regular evaluation using patented phantoms are essential for medical device quality assurance programs. This study evaluated phantom materials for tissue equivalence and their use in image quality assessment for breast ultrasound scanner performance testing using two custom-made phantoms. Methods: Two types of phantoms were developed: phantoms A and B. Phantom A was made from a base material consisting of polyvinyl chloride-plastisol with the addition of glycerol, whereas phantom B consisted of polyvinyl chloride-plastisol with the addition of graphite. Each phantom had a stiff and soft lesion shaped like a sphere, with a diameter of 1.4 cm. The phantoms were cuboids with dimensions of 10 x 10 cm(2) and a thickness of 5 cm. A series of phantom evaluations was performed, consisting of density, elasticity, acoustic properties, B-mode ultrasound images, and strain ratio. Results: The characterisation results show that background A closely resembles fibroglandular tissue in terms of density and acoustic properties (<5% variation); background B only resembles fibroglandular tissue in terms of density (-1.8% variation). In terms of elasticity, both backgrounds were close to the minimum value of fibroglandular tissue elasticity. The soft lesion on the phantom had a slightly lower density and elasticity than the carcinoma, whereas its acoustic properties (speed of sound and attenuation coefficient) were slightly higher than those of the reference carcinoma. Both phantoms were consistent with the literature in terms of strain ratio, geometric accuracy, lesion detection, and mean pixel value and showed good potential stability over one year. Conclusion: This study successfully described the fabrication and evaluation sequence of a phantom equivalent to breast fibroglandular tissue and its evaluation via ultrasound imaging. (c) 2024 The College of Radiographers. Published by Elsevier Ltd. All rights are reserved, including those for text and data mining, AI training, and similar technologies.
引用
收藏
页码:254 / 263
页数:10
相关论文
共 50 条
  • [41] Performance Evaluation of an Ultrasonic Imaging System Using Tissue-Mimicking Phantoms for Quality Assurance
    Oglat, Ammar A.
    BIOMIMETICS, 2022, 7 (03)
  • [42] Ultrasound image quality evaluation
    Loizou, C
    Pattichis, CS
    Istepanian, R
    Pantziaris, M
    Kyriakou, E
    Tyllis, T
    Nicolaides, A
    ITAB 2003: 4TH INTERNATIONAL IEEE EMBS SPECIAL TOPIC CONFERENCE ON INFORMATION TECHNOLOGY APPLICATIONS IN BIOMEDICINE, CONFERENCE PROCEEDINGS: NEW SOLUTIONS FOR NEW CHALLENGES, 2003, : 138 - 141
  • [43] Evaluation of Virtual Touch Tissue Imaging Quantification, a New Shear Wave Velocity Imaging Method, for Breast Lesion Assessment by Ultrasound
    Golatta, Michael
    Schweitzer-Martin, Mirjam
    Harcos, Aba
    Schott, Sarah
    Gomez, Christina
    Stieber, Anne
    Rauch, Geraldine
    Domschke, Christoph
    Rom, Joachim
    Schuetz, Florian
    Sohn, Christof
    Heil, Joerg
    BIOMED RESEARCH INTERNATIONAL, 2014, 2014
  • [44] ULTRASOUND PHASE CONTRAST THERMAL IMAGING WITH REFLEX TRANSMISSION IMAGING METHODS IN TISSUE PHANTOMS
    Farny, Caleb H.
    Clement, Gregory T.
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2009, 35 (12): : 1995 - 2006
  • [45] ANTHROPOMORPHIC PHANTOMS FOR ASSESSING SYSTEMS USED IN ULTRASOUND IMAGING OF THE COMPRESSED BREAST
    MADSEN, EL
    KELLYFRY, E
    FRANK, GR
    ULTRASOUND IN MEDICINE AND BIOLOGY, 1988, 14 : 183 - 201
  • [46] Tissue harmonic imaging sonography. Is the image quality in abdominal ultrasound improved?
    Stiskal, M
    Steinbach, R
    Obholzer, G
    Frank, W
    Fischer, H
    Czembirek, H
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2000, 172 (12): : 1006 - 1010
  • [47] Tissue harmonic imaging sonography of breast lesions improved margin analysis, conspicuity, and image quality compared to conventional ultrasound
    Rosen, EL
    Soo, MS
    CLINICAL IMAGING, 2001, 25 (06) : 379 - 384
  • [48] Tissue harmonic imaging sonography of breast lesions: Improved margin analysis, conspicuity, and image quality compared to standard ultrasound
    Rosen, EL
    Soo, MS
    RADIOLOGY, 1999, 213P : 198 - 199
  • [49] Tissue-inspired phantoms: a new range of equivalent tissue simulating breast, cortical bone and lung tissue
    Coulaud, J.
    Brumas, V.
    Fiallo, M.
    Sharrock, P.
    Gonneau, E.
    Courbon, F.
    Caselles, O.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2018, 4 (03):
  • [50] Image quality control in breast ultrasound
    Lagalla, R
    Midiri, M
    EUROPEAN JOURNAL OF RADIOLOGY, 1998, 27 : S229 - S233