Assessment of diagnostic accuracy of Gazelle: A point-of-care testing device for screening β-thalassemia trait

被引:0
|
作者
Singh, Neha [1 ]
Seth, Tulika [2 ]
Dass, Jasmita [2 ]
机构
[1] All India Inst Med Sci, Ctr Community Med, New Delhi, India
[2] All India Inst Med Sci, Dept Haematol, New Delhi 110029, India
关键词
Point of care testing; device validation; beta - thalassemia trait; screening; diagnostic accuracy; Hb A2; TUBE OSMOTIC FRAGILITY; IRON-DEFICIENCY; POPULATION;
D O I
10.25259/ijmr_1986_23
中图分类号
R392 [医学免疫学]; Q939.91 [免疫学];
学科分类号
100102 ;
摘要
Background & objectives: Timely detection of population with beta-thalassemia trait (BTT) followed by genetic counselling is an advocated method of preventing the birth of a child with beta-thalassemia major. In this study we aim to assess the diagnostic accuracy of Gazelle, a point-of-care (POC) testing device, in screening for BTT in hospital laboratory setting. Methods: Standards for Reporting Diagnostic Accuracy (STARD) guidelines were followed in developing study design, recruiting study participants and sample size calculation for the current research. A consecutive sample of 446 participants was recruited for this study and was tested for the reference test Gazelle as well as the gold standard cation exchange high performance liquid chromatography (CE-HPLC). Low serum ferritin levels are known to interfere with the production of HbA2 and in turn lead to false negative results. Hence, the study population was divided into two categories with respect to a cut off value of 15ng/dl of serum ferritin and the results were analyzed. Results: Overall diagnostic accuracy of Gazelle for detecting BTT was found to be 95.3 per cent with a confidence interval (CI) of (92.9 - 97.1) and the sensitivity was 94.5 per cent with a 95% CI of (84.8 - 98.8). When analyzed by the serum ferritin level the diagnostic accuracy was found to be 94.7 per cent (91.1% - 97.1%) and 95.7 per cent (91.8% - 98.1%) for participants with serum ferritin level as > 15 ng/ml and < 15 ng/ml, respectively. Interpretation & conclusions: This study found Gazelle to be a good screening tool for beta-thalassemia trait with high sensitivity, specificity and accuracy. However, it is recommended that the final confirmation of the diagnosis done by a diagnostic test like HPLC or Capillary Zone Electrophoresis (CZE).
引用
收藏
页码:194 / 200
页数:7
相关论文
共 50 条
  • [31] Point-of-care testing for procalcitonin in identifying bacterial infections in young infants: a diagnostic accuracy study
    Waterfield, Thomas
    Maney, Julie-Ann
    Hanna, Martin
    Fairley, Derek
    Shields, Michael D.
    BMC PEDIATRICS, 2018, 18
  • [32] Operator Influence on Blinded Diagnostic Accuracy of Point-of-Care Antigen Testing for Group A Streptococcal Pharyngitis
    Penney, Carla
    Porter, Robert
    O'Brien, Mary
    Daley, Peter
    CANADIAN JOURNAL OF INFECTIOUS DISEASES & MEDICAL MICROBIOLOGY, 2016, 2016 : 1 - 4
  • [33] Diagnostic Accuracy of Point-of-Care Ultrasound Performed by Pulmonary Critical Care Physicians for the Assessment of the Right Ventricle
    Filopei, Jason
    Acquah, Samuel
    Bondarsky, Eric
    Ramesh, Navitha
    Patrawalla, Paru
    CHEST, 2016, 150 (04) : 657A - 657A
  • [34] Point-of-care testing improves accuracy of STI care in an emergency department
    Huppert, Jill S.
    Taylor, Regina G.
    St Cyr, Sarah
    Hesse, Elizabeth A.
    Reed, Jennifer L.
    SEXUALLY TRANSMITTED INFECTIONS, 2013, 89 (06) : 489 - 494
  • [35] Accuracy of Bilistick (a Point-of-Care Device) to Detect Neonatal Hyperbilirubinemia
    Kamineni, Bhargavi
    Tanniru, Anusha
    Vardhelli, Venkateshwarlu
    Sharma, Deepak
    Pawale, Dinesh
    Kulkarni, Dattatray
    Muppidi, Pranitha Reddy
    Deshabhotla, Saikiran
    Murki, Srinivas
    JOURNAL OF TROPICAL PEDIATRICS, 2020, 66 (06) : 630 - 636
  • [36] Evaluation of a point-of-care haemozoin assay (Gazelle device) for rapid detection of Plasmodium knowlesi malaria
    Angelica F. Tan
    Priyaleela Thota
    Sitti Saimah Binti Sakam
    Yao Long Lew
    Giri S. Rajahram
    Timothy William
    Bridget E. Barber
    Steven Kho
    Nicholas M. Anstey
    David Bell
    Matthew J. Grigg
    Scientific Reports, 13
  • [37] Diagnostic accuracy of point-of-care ultrasound in detecting clavicle fractures
    Haak, Svenja L.
    Rietveld, Vincent
    Vos, Hendrika
    Kroon, Tineke
    Borgstede, Marion G.
    Boendermaker, Annemieke E.
    Rosendaal, Arthur
    Lameijer, Heleen
    Maaten, Jan C. Ter
    Stolmeijer, Renate
    ter Avest, Ewoud
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2025, 88 : 156 - 161
  • [38] DIAGNOSTIC ACCURACY OF POINT-OF-CARE GASTRIC ULTRASOUND: A BAYESIAN ANALYSIS
    Chaparro, Luis E.
    Kruisselbrink, Richelle
    Gharapetian, Angineh
    Ami, Noam
    Richler, Dustin
    Chan, Vincent W.
    Perlas, Anahi
    ANESTHESIA AND ANALGESIA, 2018, 126 (04): : 736 - 737
  • [39] The accuracy of point-of-care ultrasound as a diagnostic tool for patella fractures
    Kilic, Turgay Yilmaz
    Yesilaras, Murat
    Atilla, Ozge Duman
    Turgut, Ali
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2016, 34 (08): : 1576 - 1578
  • [40] The point of point-of-care testing
    Kassaye, Seble G.
    Katzenstein, David
    LANCET, 2011, 378 (9802): : 1532 - 1533