Micro-CT Structure Analysis on Dental Implants: Preliminary In Vitro Trial

被引:0
|
作者
Galletti, Fulvia [1 ]
D'Angelo, Tommaso [1 ]
Fiorillo, Luca [1 ,2 ,3 ,4 ]
Lo Giudice, Paola [5 ]
Irrera, Natasha [4 ]
Rizzo, Giuseppina [4 ]
Cervino, Gabriele [1 ]
机构
[1] Univ Messina, Dept Biomed & Dent Sci & Morphofunct Imaging, Via Consolare Valeria 1, I-98125 Messina, Italy
[2] Univ Campania Luigi Vanvitelli, Multidisciplinary Dept Med Surg & Odontostomatol S, I-80121 Naples, Italy
[3] Dr DY Patil Vidyapeeth, Dr DY Patil Dent Coll & Hosp, Dept Dent Cell Res, Pune 411018, India
[4] Univ Messina, Dept Clin & Expt Med, Via Consolare Valeria 1, I-98125 Messina, Italy
[5] Univ Catania, Dept Med & Surg Sci, Sect Dent, I-95123 Catania, Italy
来源
PROSTHESIS | 2024年 / 6卷 / 06期
关键词
dental implants; microcomputed tomography (micro-CT); osseointegration; implant stability; OSSEOINTEGRATION;
D O I
10.3390/prosthesis6060104
中图分类号
TB3 [工程材料学]; R318.08 [生物材料学];
学科分类号
0805 ; 080501 ; 080502 ;
摘要
Introduction: This preliminary in vitro study aims to evaluate the application of micro-CT in analyzing the microstructural coupling between dental implant fixtures and prosthetic abutments, with an emphasis on understanding the effectiveness and limitations of this technique in dental implantology. Materials and Methods: A search of PubMed, MEDLINE, and the Cochrane Library up to May 2024 identified eight relevant studies that examined different facets of dental implantology, such as osseointegration, implant stability, and the comparative accuracy of micro-CT versus other imaging techniques. A comparative micro-CT radiographic analysis was performed on five different implant fixtures with respective prosthetic and healing abutments, by using SkyScan1174 micro-CT. Results: The reviewed studies demonstrated that micro-CT is reliable for assessing bone quality, implant stability, and the microstructural integrity of dental implants. Micro-computed tomography (micro-CT) studies reveal bone-implant contact (BIC) ratios of 40-80%, bone volume per total volume (BV/TV) values of 20-60%, and detect microgaps as small as 0.3 mu m, highlighting its high-resolution capability (5-10 mu m) for detailed implant analysis. The comparative analysis of the implant fixtures analyzed the implant-abutment connection, highlighting the relevance of implant design for ensuring stability. Conclusions: Micro-CT analysis has proven to be a valuable tool for evaluating the intricate microstructural properties of dental implants, offering insights into implant stability, bone quality, and osseointegration. The literature reviewed highlights consistent findings that underscore micro-CT's accuracy and reliability in capturing high-resolution data, suggesting its potential as a standard imaging modality in implant research and clinical assessment.
引用
收藏
页码:1437 / 1447
页数:11
相关论文
共 50 条
  • [21] Tooth fracture risk analysis based on a new finite element dental structure models using micro-CT data
    Chen, G.
    Fan, W.
    Mishra, S.
    El-Atem, A.
    Schuetz, M. A.
    Xiao, Y.
    COMPUTERS IN BIOLOGY AND MEDICINE, 2012, 42 (10) : 957 - 963
  • [22] Medial Axis Detection from Dental Micro-CT Records
    Benyo, Balazs
    Szilagyi, Laszlo
    Dobo-Nagy, Csaba
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 4: IMAGE PROCESSING, BIOSIGNAL PROCESSING, MODELLING AND SIMULATION, BIOMECHANICS, 2010, 25 : 1688 - 1691
  • [23] Surface extraction can provide a reference for micro-CT analysis of retrieved total knee implants
    Teeter, Matthew G.
    Milner, Jaques S.
    Naudie, Douglas D. R.
    MacDonald, Steven J.
    KNEE, 2014, 21 (04): : 801 - 805
  • [24] Can implants move in bone? A longitudinal in vivo micro-CT analysis of implants under constant forces in rat vertebrae
    Becker, Kathrin
    Schwarz, Frank
    Rauch, Nicole Jasmin
    Khalaph, Silava
    Mihatovic, Ilja
    Drescher, Dieter
    CLINICAL ORAL IMPLANTS RESEARCH, 2019, 30 (12) : 1179 - 1189
  • [25] Preliminary experimental results from a MARS Micro-CT system
    He, Peng
    Yu, Hengyong
    Thayer, Patrick
    Jin, Xin
    Xu, Qiong
    Bennett, James
    Tappenden, Rachael
    Wei, Biao
    Goldstein, Aaron
    Renaud, Peter
    Butler, Anthony
    Butler, Phillip
    Wang, Ge
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2012, 20 (02) : 199 - 211
  • [26] Preliminary studies of a micro-CT for a combined small animal PET/CT scanner
    Khodaverdi, M
    Pauly, F
    Weber, S
    Schröder, G
    Ziemons, K
    Sievering, R
    Halling, H
    2001 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORDS, VOLS 1-4, 2002, : 1605 - 1606
  • [27] Micro finite element analysis of continuously loaded mini-implants - A micro-CT study in the rat tail model
    Kerberger, Robert
    Brunello, Giulia
    Drescher, Dieter
    van Rietbergen, Bert
    Becker, Kathrin
    BONE, 2023, 177
  • [28] Multislice-CT for structure - Analysis of trabecular bone - a comparison with Micro-CT and biomechanical strength
    Bauer, JS
    Issever, AS
    Fischbeck, M
    Burghardt, A
    Eckstein, F
    Rummeny, EJ
    Majumdar, S
    Link, TM
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2004, 176 (05): : 709 - 718
  • [29] Use of rheometry and micro-CT analysis to understand pore structure development in coke
    Steel, Karen M.
    Dawson, Robin E.
    Jenkins, David R.
    Pearce, Robin
    Mahoney, Merrick R.
    FUEL PROCESSING TECHNOLOGY, 2017, 155 : 106 - 113
  • [30] Validity of micro-CT for in vitro caries detection: a systematic review and meta-analysis
    Oliveira, Luciana Butini
    Massignan, Carla
    Oenning, Anne Caroline
    Rovaris, Karla
    Bolan, Michele
    Porporatti, Andre Luis
    Canto, Graziela De Luca
    DENTOMAXILLOFACIAL RADIOLOGY, 2020, 49 (07)