Energy response of CaSO4:Dy thermoluminescence detector to medical linear accelerator photon beams

被引:0
|
作者
Herwiningsih, Sri [1 ]
Izzati, Lia Wilda [1 ]
Bunawas, Bunawas [2 ]
机构
[1] Univ Brawijaya, Fac Math & Nat Sci, Phys Dept, 1 Vet St, Malang 65145, East Java, Indonesia
[2] Nuklindo Lab, Res & Dev Div, Dept Math, 5A Ir H Juanda St,Block B Kav P-Q, South Tangerang 15412, Indonesia
关键词
D O I
10.1093/rpd/ncae211
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The study aims to evaluate the energy response of the thermoluminescent dosemeter (TLD) CaSO4:Dy from high energy photon beams produced from medical linear accelerator. The test was performed on the polymethyl methacrylate phantom surface and at the depth of dose maximum with a source-to-surface distance of 100 cm and the radiation field size of 12 x 12 cm. The results were compared with the TLD response exposed to Co-60 standard source. The results show that the TLD response to the linac's photon differs >30% compared to the TLD response to Co-60 exposure when measured at the phantom surface, while at the depth of maximum dose the response is similar (< +/- 10%), indicating that the TLD response is no longer dependent on the energy. This implies that the current dose reading evaluation based on Co-60 calibration curve should be corrected to obtain more accurate dose report of the radiation workers.
引用
收藏
页码:2014 / 2020
页数:7
相关论文
共 50 条
  • [21] Studies of UV induced phototransferred thermoluminescence in CaSO4:Dy pellets
    Grossi, FH
    Campos, LL
    RADIATION PROTECTION DOSIMETRY, 2002, 100 (1-4) : 421 - 424
  • [23] Preliminary Studies of Thermoluminescence Dosimeter (TLD) CaSO4:Dy Synthesis
    Nuraeni, N.
    Iskandar, F.
    Waris, A.
    Haryanto, F.
    INTERNATIONAL CONFERENCE ON ENERGY SCIENCES (ICES 2016), 2017, 877
  • [24] Estimation of bremsstrahlung photon energy in the environment of high-energy electron accelerator using CaSO4:Dy based TL dosemeter
    Bakshi, A. K.
    Nayak, M. K.
    Haridas, G.
    Chatterjee, S.
    Kher, R. K.
    RADIATION PROTECTION DOSIMETRY, 2008, 128 (01) : 43 - 48
  • [25] EPR response of anhydrite crystal (CaSO4) for dosimetry of gamma photon beams
    Cano, Nilo F.
    Rao, T. K. Gundu
    Ayala Arenas, Jorge S.
    Javier-Ccallata, Henry S.
    Watanabe, S.
    RADIATION PHYSICS AND CHEMISTRY, 2021, 180 (180)
  • [26] THERMOLUMINESCENCE OF CaSO4:Dy - SEMIEMPIRICAL APPROACH FOR A QUALITATIVE ANALYSIS OF NONLINEAR THERMOLUMINESCENCE EFFECTS.
    Srivastava, J.K.
    Supe, S.J.
    1600, (92):
  • [27] PHOTON ENERGY-DEPENDENCE OF CASO4 - DY EMBEDDED TEFLON TLD DISKS
    PRADHAN, AS
    KHER, RK
    DERE, A
    BHATT, RC
    INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1978, 29 (4-5): : 243 - 245
  • [28] Evaluation of thermoluminescence of 200 keV carbon ion irradiated CaSO4:Dy nanophosphors for medical dosimetry
    Bhadane, Mahesh S.
    Akhtar, Shahzad
    Dahiwale, S. S.
    Hareesh, K.
    Asokan, K.
    Kanjilal, D.
    Bhoraskar, V. N.
    Dhole, S. D.
    JOURNAL OF LUMINESCENCE, 2017, 192 : 695 - 700
  • [29] Thermoluminescence studies of CaSO4:Dy nanophosphor for application in high dose measurements
    Mandlik, Nandkumar
    Dhole, S. D.
    Sahare, P. D.
    Bakare, J. S.
    Balraj, A.
    Bhatt, B. C.
    APPLIED RADIATION AND ISOTOPES, 2019, 148 : 253 - 261
  • [30] Study of Residual TL of CaSO4:Dy-based Thermoluminescence Dosemeter
    Sahai, Manish K.
    Pradhan, S. M.
    Bakshi, A. K.
    Datta, D.
    RADIATION PROTECTION DOSIMETRY, 2017, 177 (03) : 342 - 347