Sensitive LC-MS/MS Assay for Total Testosterone Quantification on Unit Resolution and High-Resolution Instruments

被引:0
|
作者
Wolken, Jill K. [1 ]
Peterson, Meghan M. [1 ]
Cao, Wenjing [1 ,2 ]
Challoner, Keith [1 ]
Jin, Zhicheng [1 ,2 ]
机构
[1] Univ Wisconsin Hosp & Clin, Madison, WI 53792 USA
[2] Univ Wisconsin Madison, Dept Pathol & Lab Med, 600 Highland Ave, Madison, WI 53792 USA
关键词
testosterone; multiple reaction monitoring; high-resolution mass spectrometry; assay harmonization; parallel reaction monitoring; TANDEM MASS-SPECTROMETRY; SERUM TESTOSTERONE; VALIDATION;
D O I
10.3390/jcm13237056
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Testosterone is an androgenic hormone that plays important roles in both males and females. The circulating levels of total testosterone vary from 1 to 1480 ng/dL. High-throughput immunoassays often lack accuracy in lower concentration ranges (below 100 ng/dL), particularly when used for females or children. To address this limitation, we developed a total testosterone LC-MS/MS assay on three instruments. Methods: Sample preparation began with the dilution and conditioning of 200 mu L of serum. A supported liquid extraction cartridge was used to extract the analyte from biological matrices. Chromatographic separation was achieved using a C18 column with a runtime of 5 min per sample. This assay was validated on a Triple Quad 6500 and an API 4500 instrument. Results: Method validation was carried out according to the CLSI C62-ED2 guideline and our hospital protocol. The within-day coefficient of variation (CV) was less than 10% and the between-day CV was less than 15%. The assay had a limit of quantitation of 0.5 ng/dL with an analyte measure range of 2-1200 ng/dL. A comparison using Deming regression and Bland-Altman plots showed that this assay correlated well with a reference method. The results from the API 4500 and an Orbitrap were consistent with those from the TQ 6500. Both serum-separator tubes (BD) and serum-activator tubes were found to be suitable. Conclusions: We successfully developed and validated a robust total testosterone LC-MS/MS assay for routine clinical testing. This assay was harmonized across two triple quadrupole instruments and one high-resolution mass spectrometer.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] A Sensitive LC-MS/MS Assay for Total Testosterone Quantification
    Cao, W.
    Wolken, J. K.
    Peterson, M. M.
    Challoner, K. E.
    Jin, Z.
    CLINICAL CHEMISTRY, 2024, 70
  • [2] Structure Annotation and Quantification of Wheat Seed Oxidized Lipids by High-Resolution LC-MS/MS
    Riewe, David
    Wiebach, Janine
    Altmann, Thomas
    PLANT PHYSIOLOGY, 2017, 175 (02) : 600 - 618
  • [3] Simultaneous quantification of total and free testosterone in human serum by LC-MS/MS
    Huang, Rongmei
    Hong, Yi
    Wu, Yike
    Li, Weifeng
    Liu, Wenlan
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2023, 415 (27) : 6851 - 6861
  • [4] A sensitive assay for the measurement of testosterone in human serum by LC-MS/MS.
    Nelson, RE
    Singh, RJ
    Grebe, SK
    CLINICAL CHEMISTRY, 2004, 50 (06) : A106 - A106
  • [5] The bioavailable testosterone assay with LC-MS/MS in aging male
    Horiuchi, T
    Chiba, Y
    Kurosaki, T
    Murayama, T
    Onouchi, T
    PROCEEDINGS OF THE 12TH INTERNATIONAL CONGRESS OF ENDOCRINOLOGY, 2004, : 1283 - 1286
  • [6] Determination of Mycotoxins in Food Matrices Using LC-MS/MS Compared With High-resolution Orbitrap™ MS Technology
    Gottfried, Sebastian
    Herebian, Diran
    CURRENT ANALYTICAL CHEMISTRY, 2013, 9 (01) : 99 - 107
  • [7] Characterization and quantification of oxyntomodulin in human and rat plasma using high-resolution accurate mass LC-MS
    Cox, Jennifer M.
    Berna, Michael J.
    Jin, Zhaoyan
    Cox, Amy L.
    Sloop, Kyle W.
    Gutierrez, Jesus A.
    Ackermann, Bradley L.
    BIOANALYSIS, 2016, 8 (15) : 1579 - 1595
  • [8] Total testosterone quantitative measurement in serum by LC-MS/MS
    Wang, Yuesong
    Gay, Gabrielle D.
    Botelho, Julianne Cook
    Caudill, Samuel P.
    Vesper, Hubert W.
    CLINICA CHIMICA ACTA, 2014, 436 : 263 - 267
  • [9] Sensitive glucagon quantification by immunochemical and LC-MS/MS methods
    Lapko, Veniamin N.
    Miller, Patrick S.
    Brown, G. Paul
    Islam, Rafiqul
    Peters, Sarah K.
    Sukovaty, Richard L.
    Ruhn, Peggy F.
    Kafonek, Chris J.
    BIOANALYSIS, 2013, 5 (23) : 2957 - 2972
  • [10] Exploration of the Nurse Shark (Ginglymostoma cirratum) Plasma Immunoproteome Using High-Resolution LC-MS/MS
    Bakke, Fiona K.
    Gundappa, Manu Kumar
    Matz, Hanover
    Stead, David A.
    Macqueen, Daniel J.
    Dooley, Helen
    FRONTIERS IN IMMUNOLOGY, 2022, 13