Validation of Noninvasive Derivation of the Central Aortic Pressure Waveform from Fingertip Photoplethysmography Using a Novel Selective Transfer Function Method

被引:0
|
作者
Cox, James R. [1 ]
Akeila, Ehad [2 ]
Avolio, Alberto P. [1 ,2 ]
Butlin, Mark [1 ]
Liao, Catherine [2 ]
Bentley, Gisele J. [1 ]
Qasem, Ahmad [1 ,2 ]
机构
[1] Macquarie Univ, Macquarie Med Sch, Fac Med Hlth & Human Sci, Sydney, NSW, Australia
[2] Cardiex Ltd, Sydney, NSW, Australia
关键词
Photoplethysmography; Transfer function; Waveform analysis; Aortic pressure; Blood pressure; BLOOD-PRESSURE; ARTERIAL STIFFNESS; AUGMENTATION INDEX; RISK; REFLECTIONS;
D O I
10.1159/000540666
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Introduction: Central aortic pressure waveform analyses can provide clinically relevant information beyond conventional brachial blood pressure (BP) assessment. This waveform can be reproduced noninvasively through application of a generalized transfer function (GTF) on a peripheral waveform, as conventionally performed by applanation tonometry. Photoplethysmography (PPG) is an alternate approach; however, differences in measurement site and modality demand the use of a transfer function (TF) specific for those differences. This study aimed to compare central aortic waveform features generated from radial tonometry (reference) using a proprietary GTF with a central aortic waveform and its features generated from a simultaneous fi ngertip PPG measurement using a selective method where one of three different TFs is chosen based on the input signal harmonic profile. Methods: Brachial BP was measured in triplicate under resting conditions and was used for subsequent calibration. Multiple simultaneous radial tonometry (SphygmoCor CVMS) and fi ngertip PPG measurements were then performed in individual participants (n = 21, 10 females, age: 39 +/- 16 years). Measurements were converted into central aortic waveforms with their respective TFs. Twenty central aortic pressure waveform parameters were compared through correlation analysis, Bland-Altman plots, and a repeated measure mixed-effects ANOVA model. Central aortic waveform shape was compared using the root-mean-squared error (RMSE). Results: Correlation (r) of PPG-derived parameters with radially tonometry-derived central aortic parameters was high ranging from 0.79 to 0.99. Mean differences of pressure-related parameters were within 1.3 mm Hg, and differences of time-related parameters ranged from - 2.2 to 3.4%. While some parameters were statistically different, these differences are not physiologically meaningful. Central aortic waveform shape had an average RMSE of 1.8 +/- 0.9%. Conclusion: Fingertip PPG-derived central aortic waveform parameters using a novel selective TF were comparable to central aortic waveform features derived from radial tonometry using a previously validated GTF. (c) 2024 The Author(s). Published by S. Karger AG, Basel
引用
收藏
页码:95 / 105
页数:11
相关论文
共 50 条
  • [1] Validation of an Adaptive Transfer Function Method to Estimate the Aortic Pressure Waveform
    Yao, Yang
    Xu, Lisheng
    Sun, Yingxian
    Fu, Qiang
    Zhou, Shuran
    He, Dianning
    Zhang, Yahui
    Guo, Liang
    Zheng, Dingchang
    IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2017, 21 (06) : 1599 - 1606
  • [2] Noninvasive Estimation of Intracranial Pressure Waveform from Central Aortic Pressure Waveform
    Evensen, Karen Brastad
    Eide, Per Kristian
    WORLD NEUROSURGERY, 2019, 121 : 257 - 258
  • [3] 'Generalizability' of a radial-aortic transfer function for the derivation of central aortic waveform parameters
    Hope, Sarah A.
    Meredith, Ian T.
    Tay, David
    Cameron, James D.
    JOURNAL OF HYPERTENSION, 2007, 25 (09) : 1812 - 1820
  • [4] Estimation of central aortic pressure waveform by mathematical transformation of radial tonometry pressure - Validation of generalized transfer function
    Chen, CH
    Nevo, E
    Fetics, B
    Pak, PH
    Yin, FCP
    Maughan, WL
    Kass, DA
    CIRCULATION, 1997, 95 (07) : 1827 - 1836
  • [5] Personalized aortic pressure waveform estimation from brachial pressure waveform using an adaptive transfer function
    Du, Shuo
    Yao, Yang
    Sun, Guozhe
    Wang, Lu
    Alastruey, Jordi
    Avolio, Alberto P.
    Xu, Lisheng
    COMPUTERS IN BIOLOGY AND MEDICINE, 2023, 155
  • [6] Development and Validation of a Novel Method to Derive Central Aortic Systolic Pressure From the Radial Pressure Waveform Using an N-Point Moving Average Method
    Williams, Bryan
    Lacy, Peter S.
    Yan, Peter
    Hwee, Chua-Ngak
    Liang, Chen
    Ting, Choon-Meng
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2011, 57 (08) : 951 - 961
  • [7] A novel method to derive central aortic systolic pressure (CASP) from the radial artery pressure waveform: validation with data from the CAFE study
    Williams, B.
    Lacy, P. S.
    Ting, C. M.
    Chua, N. H.
    Chen, L.
    JOURNAL OF HUMAN HYPERTENSION, 2008, 22 (10) : 733 - 734
  • [8] Parametric model derivation of transfer function for noninvasive estimation of aortic pressure by radial tonometry
    Fetics, B
    Nevo, E
    Chen, CH
    Kass, DA
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1999, 46 (06) : 698 - 706
  • [9] An adaptive transfer function for deriving the aortic pressure waveform from a peripheral artery pressure waveform
    Swamy, Gokul
    Xu, Da
    Olivier, N. Bari
    Mukkamala, Ramakrishna
    AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 2009, 297 (05): : H1956 - H1963
  • [10] Validation of the transfer function technique for generating central from peripheral upper limb pressure waveform
    Gallagher, D
    Adji, A
    O'Rourke, MF
    AMERICAN JOURNAL OF HYPERTENSION, 2004, 17 (11) : 1059 - 1067