D-dimer assay interference detected by the discrepancy in D-dimer concentrations at different dilutions: a case report

被引:0
|
作者
Milic, Marija [1 ,2 ]
Barbaric, Dejana Brkic [3 ]
Lukic, Iva [1 ,2 ]
Kirin, Mirna [4 ]
Buljanovic, Vikica [5 ]
Seric, Vatroslav [1 ,2 ]
机构
[1] Univ Hosp Ctr Osijek, Clin Inst Lab Diagnost, Osijek, Croatia
[2] Univ Osijek, Fac Med, Dept Med Chem Biochem & Clin Chem, Osijek, Croatia
[3] Univ Hosp Ctr Osijek, Clin Inst Transfus Med, Osijek, Croatia
[4] Siemens Healthineers, Siemens Healthcare doo, Zagreb, Croatia
[5] Gen Hosp Nasice, Med Biochem Lab, Nashice, Croatia
关键词
D-dimer; heterophile antibodies; blood coagulation tests; interference;
D O I
10.11613/BM.2024.031001
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
This case report describes interference from heterophilic antibodies in D-dimer assay. The interference was suspected due to discrepancies between D-dimer concentrations in the original sample and diluted samples, as well as inconsistent clinical findings. The patient's medical history, laboratory results, and imaging studies were considered in the investigation. Heterophilic antibodies, likely developed during the SARS-CoV-2 infection, were identified as the probable cause of interference. The interference was confirmed through various methods, including dilution studies, blocking heterophilic antibodies, and comparing results with an alternative D-dimer method. This case highlights the importance of recognizing and addressing interference in D-dimer testing, emphasizing the need for collaboration between clinicians and laboratory specialists.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] The D-dimer assay
    Johnson, Eric D.
    Schell, John C.
    Rodgers, George M.
    AMERICAN JOURNAL OF HEMATOLOGY, 2019, 94 (07) : 833 - 839
  • [2] D-dimer assay
    Smally, AJ
    ANNALS OF EMERGENCY MEDICINE, 1998, 31 (03) : 416 - 416
  • [3] VIDAS D-DIMER - A NEW SYSTEM FOR THE ASSAY OF D-DIMER IN EMERGENCY
    BRONNER, J
    DROUIN, B
    SARTRE, J
    DUPUY, G
    PITTET, JL
    LELOUARN, J
    THROMBOSIS AND HAEMOSTASIS, 1995, 73 (06) : 1101 - 1101
  • [4] Comparison of the PATHFAST D-dimer assay with two POC D-dimer assays
    E Spanuth
    B Ivandic
    R Thomae
    E Giannitsis
    Critical Care, 19 (Suppl 1):
  • [5] Standardizing the D-dimer Assay: Proposing the D-dimer International Managed Ratio
    Kahler, Zachary P.
    Kline, Jeffrey A.
    CLINICAL CHEMISTRY, 2015, 61 (05) : 776 - 778
  • [6] Performance Evaluation of the VIDAS® D-Dimer ExclusionTM II Assay and the Beckman Coulter D-Dimer Assay
    Yen, Kuo-Tun
    Jiang, Jin-Wei Fan
    Chen, Chih-Chieh
    Er, Tze-Kiong
    CLINICAL LABORATORY, 2023, 69 (09) : 2023 - 2025
  • [8] Plasma D-dimer concentrations in different clinical conditions
    Ver Elst, K
    Jochmans, K
    De Pauw, A
    De Waele, M
    ACTA CLINICA BELGICA, 2002, 57 (06): : 325 - 330
  • [9] Perioperative lung cancer: To D-dimer or not to D-dimer that is the question?
    Staunton, Laura
    Henderson, Brian
    Gately, Kathy
    IRISH JOURNAL OF MEDICAL SCIENCE, 2024, 193 : S253 - S254