Fabrication and characterization of inkjet-printed interdigitated electrodes for non-faradaic electrochemical detection of uromodulin in urine

被引:0
|
作者
Vora, Kunj [1 ]
Kordas, Norbert [1 ]
Seidl, Karsten [1 ,2 ]
机构
[1] Fraunhofer Inst Microelect Circuits & Syst IMS, Finkenstr 61, D-47057 Duisburg, Germany
[2] Univ Duisburg Essen, Dept Elect Components & Circuits, Forsthausweg 2, D-47057 Duisburg, Germany
关键词
Biosensor; Impedance spectroscopy; Inkjet-printing; Interdigitated electrodes; Kidney-disease; Non-faradaic; Single-use-sensor; SU-8; Uromodulin; IMPEDANCE SPECTROSCOPY; LABEL-FREE; BIOSENSOR; PAPER; PHYSIOLOGY; BIOMARKER; DESIGN; SENSOR; WATER;
D O I
10.1016/j.sbsr.2024.100715
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This work describes the simulation, design, and fabrication of an inkjet-printed impedance biosensor for uromodulin which is a urine biomarker for kidney disease. The interdigitated electrodes (IDE) were fabricated through inkjet printing of silver ink on polyethylene terephthalate (PET) substrate and subsequently passivated with SU-8 for non-faradaic measurements. We functionalized the passivated IDE with biotin-streptavidin complex and uromodulin antibody to detect uromodulin (6.75 mu g/ml-100 mu g/ml). The measured impedance due to uromodulin immobilized from artificial urine samples showed a 27.4 % change in impedance magnitude per mu g/ ml of uromodulin at 150 Hz (log scale). The limit of detection achieved is 25 ng/ml. The sensor demonstrated specificity to uromodulin when measured against albumin, which is another protein biomarker present in the urine. These results highlight the possibility of implementing non-faradaic impedance measurements using a passivated IDE as the sensor element. It also promotes the sensor fabrication method of inkjet printing. The sensor concept shows the implementation potential of IDE as a single-use-sensor for point-of-care applications.
引用
收藏
页数:7
相关论文
共 50 条
  • [41] Tuning the deposition parameters for optimizing the faradaic and non-faradaic electrochemical performance of nanowire array-shaped ITO electrodes prepared by electron beam evaporation
    Pruna, Raquel
    Lopez, Manel
    Teixidor, Francesc
    NANOSCALE, 2019, 11 (01) : 276 - 284
  • [42] Non-faradaic electrochemical modification of the catalytic activity for propane combustion of Pt/YSZ and Rh/YSZ catalyst-electrodes
    Bebelis, S.
    Kotsionopoulos, N.
    SOLID STATE IONICS, 2006, 177 (26-32) : 2205 - 2209
  • [43] Accurate ion type and concentration detection using two bare electrodes by machine learning of non-faradaic electrochemical impedance measurements of an automated fluidic system
    Sagiroglu, M. Zahid
    Demirel, Eda Deniz
    Mutlu, Senol
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2024, 961
  • [44] Fabrication and Characterization of Spiral Interdigitated Electrodes Based Biosensor for Salivary Glucose Detection
    Adelyn, P. Y. P.
    Hashim, U.
    Arshad, M. K. Md
    Voon, C. H.
    Liu, Wei-Wen
    Kahar, S. M.
    Huda, A. R. N.
    Lee, H. Cheun
    11TH ASIAN CONFERENCE ON CHEMICAL SENSORS (ACCS2015), 2017, 1808
  • [45] Three-Dimensional Inkjet-Printed Electrochemical Sensor on Shape Memory Polymer for Aqueous Lead Detection
    Arif, Annatoma
    Contreras, Angela Mendez
    Hossain, Sheikh Dobir
    Roberts, Robert C.
    IEEE SENSORS JOURNAL, 2023, 23 (13) : 13868 - 13875
  • [46] A Low-Cost Inkjet-Printed Aptamer-Based Electrochemical Biosensor for the Selective Detection of Lysozyme
    Khan, Niazul Islam
    Maddaus, Alec G.
    Song, Edward
    BIOSENSORS-BASEL, 2018, 8 (01):
  • [47] 3-D inkjet-printed solid oxide electrochemical reactors. II. LSM - YSZ electrodes
    Farandos, N. M.
    Li, T.
    Kelsall, G. H.
    ELECTROCHIMICA ACTA, 2018, 270 : 264 - 273
  • [48] Feasibility of a piezoelectric pressure sensor using metal-organic framework hafnium-based UiO-66 and inkjet-printed interdigitated electrodes
    Duong, Melinda
    Garnsey, Sean
    Flynn, Paul
    Bhalla, Amar S.
    Guo, Ruyan
    FERROELECTRICS, 2023, 611 (01) : 77 - 83
  • [49] Disposable inkjet-printed electrochemical platform for detection of clinically relevant HER-2 breast cancer biomarker
    Carvajal, Susanita
    Fera, Samantha N.
    Jones, Abby L.
    Baldo, Thaisa A.
    Mosa, Islam M.
    Rusling, James F.
    Krause, Colleen E.
    BIOSENSORS & BIOELECTRONICS, 2018, 104 : 158 - 162
  • [50] Non-Faradaic Electrochemical Detection of Exocytosis from Mast and Chromaffin Cells Using Floating-Gate MOS Transistors
    Krishna Jayant
    Amit Singhai
    Yingqiu Cao
    Joshua B. Phelps
    Manfred Lindau
    David A. Holowka
    Barbara A. Baird
    Edwin C. Kan
    Scientific Reports, 5